Impedance measurements for early detection of breast cancer in younger women: A preliminary assessment

被引:6
作者
Sumkin, JH [1 ]
Stojadinovic, A [1 ]
Huerbin, M [1 ]
Klym, A [1 ]
McHugh, L [1 ]
Sobran, CM [1 ]
Leader, JK [1 ]
Zheng, B [1 ]
Gur, D [1 ]
机构
[1] Univ Pittsburgh, Dept Radiol, Pittsburgh, PA 15213 USA
来源
MEDICAL IMAGING 2003: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT | 2003年 / 5034卷
关键词
breast; cancer; early detection; electrical impedance measurements;
D O I
10.1117/12.480073
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The purpose of this preliminary investigation is to explore the possibility that electrical impedance measurements of the breast can ultimately be used to screen younger women for early detection of breast cancer. As a part of a comprehensive protocol to compare different modalities, participating women undergo a series of diagnostic examinations, including impedance measurements under IRB-approved protocols. The results of the frequency-dependent algorithm are compared with the results of other. imaging modalities as well as diagnostic outcome when available. In a preliminary series of 83 patients (divided into two groups) with varying risk levels, a significant correlation between impedance measurements and results from other diagnostic modalities was observed. The specific algorithm developed for high specificity resulted in an overall performance level of 90 percent specificity.. The procedure was found to be "simple," "fast," and "easy to use" by the technologists. The interpretation of the results is straightforward. Our preliminary assessment is encouraging and indicates that the system may prove extremely useful for the purpose it was designed. Further technical improvements and clinical assessments are underway.
引用
收藏
页码:197 / 203
页数:7
相关论文
共 24 条
[1]   The T-SCAN™ technology:: electrical impedance as a diagnostic tool for breast cancer detection [J].
Assenheimer, M ;
Laver-Moskovitz, O ;
Malonek, D ;
Manor, D ;
Nahaliel, U ;
Nitzan, R ;
Saad, A .
PHYSIOLOGICAL MEASUREMENT, 2001, 22 (01) :1-8
[2]   Ex vivo discrimination between normal and pathological tissues in human breast surgical biopsies using bioimpedance spectroscopy [J].
Chauveau, N ;
Hamzaoui, L ;
Rochaix, P ;
Rigaud, B ;
Voigt, JJ ;
Morucci, JP .
ELECTRICAL BIOIMPEDANCE METHODS: APPLICATIONS TO MEDICINE AND BIOTECHNOLOGY, 1999, 873 :42-50
[3]  
FOSTER KR, 1989, CRIT REV BIOMED ENG, V17, P25
[4]  
Fuchsjaeger MH, 2001, RADIOLOGY, V221, P556
[5]   Novel EIS postprocessing algorithm for breast cancer diagnosis [J].
Glickman, YA ;
Filo, O ;
Nachaliel, U ;
Lenington, S ;
Amin-Spector, S ;
Ginor, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (06) :710-712
[6]   Characteristics of breast carcinomas missed by screening radiologists [J].
Goergen, SK ;
Evans, J ;
Cohen, GPB ;
MacMillan, JH .
RADIOLOGY, 1997, 204 (01) :131-135
[7]   The impedivity of freshly excised human breast tissue [J].
Jossinet, J .
PHYSIOLOGICAL MEASUREMENT, 1998, 19 (01) :61-75
[8]   Variability of impedivity in normal and pathological breast tissue [J].
Jossinet, J .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1996, 34 (05) :346-350
[9]  
Kolb TM, 2002, RADIOLOGY, V225, P554
[10]  
Kopans D., 1998, BREAST IMAGING