Optimal Fat Suppression in Head and Neck MRI: Comparison of Multipoint Dixon with 2 Different Fat-Suppression Techniques, Spectral Presaturation and Inversion Recovery, and STIR

被引:33
作者
Gaddikeri, S. [1 ,2 ]
Mossa-Basha, M. [3 ]
Andre, J. B. [5 ]
Hippe, D. S. [4 ]
Anzai, Y. [6 ]
机构
[1] Rush Univ, Med Ctr, Dept Neuroradiol, Chicago, IL 60612 USA
[2] Rush Univ, Med Ctr, Dept Radiol, 1653 W Congress Pkwy, Chicago, IL 60612 USA
[3] Univ Washington, Dept Neuroradiol, Seattle, WA 98195 USA
[4] Univ Washington, Dept Radiol, Seattle, WA 98195 USA
[5] Univ Washington, Med Ctr, Dept Neuroradiol, Seattle, WA 98195 USA
[6] Univ Utah, Hlth Ctr, Dept Radiol, Salt Lake City, UT USA
关键词
WATER; SEPARATION; SEQUENCE; IMAGES; PROTON; LIVER;
D O I
10.3174/ajnr.A5483
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: Uniform complete fat suppression is essential for identification and characterization of most head and pathology. Our aim was to compare the multipoint Dixon turbo spin-echo fat-suppression technique with 2 different fat-suppression techniques, including a hybrid spectral presaturation with inversion recovery technique and an inversion recovery STIR technique, in head and neck fat-suppression MR imaging. MATERIALS AND METHODS: Head and neck MR imaging datasets of 72 consecutive patients were retrospectively reviewed. All patients were divided into 2 groups based on the type of fat-suppression techniques used (group A: STIR and spectral presaturation with inversion recovery gadolinium-T1WI; group B: multipoint Dixon T2 TSE and multipoint Dixon gadolinium-T1WI TSE). Objective and subjective image quality and scan acquisition times were assessed and compared between multipoint Dixon T2 TSE versus STIR and multipoint Dixon gadolinium-T1WI TSE versus spectral presaturation with inversion recovery gadolinium-T1WI using the Mann-Whitney U test. RESULTS: A total of 64 patients were enrolled in the study (group A, n = 33 and group B, n = 31). Signal intensity ratios were significantly higher for multipoint Dixon T2 and gadolinium-T1WI techniques compared with STIR (P < .001) and spectral presaturation with inversion recovery gadolinium-T1WI (P < .001), respectively. Two independent blinded readers revealed that multipoint Dixon T2 and gadolinium-T1WI techniques had significantly higher overall image quality (P =.022 and P < .001) and fat-suppression grades (P < .013 and P < .001 across 3 different regions) than STIR and spectral presaturation with inversion recovery gadolinium-T1WI, respectively. The scan acquisition time was relatively short for the multipoint Dixon technique (2 minutes versus 4 minutes 56 seconds for the T2-weighted sequence and 2 minutes versus 3 minutes for the gadolinium-T1WI sequence). CONCLUSIONS: The multipoint Dixon technique offers better image quality and uniform fat suppression at a shorter scan time compared with STIR and spectral presaturation with inversion recovery gadolinium-T1WI techniques.
引用
收藏
页码:362 / 368
页数:7
相关论文
共 17 条
  • [1] MR IMAGING OF THE LIVER USING SHORT TI INVERSION RECOVERY SEQUENCES
    BYDDER, GM
    STEINER, RE
    BLUMGART, LH
    KHENIA, S
    YOUNG, IR
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (06) : 1084 - 1089
  • [2] MR IMAGING - CLINICAL USE OF THE INVERSION RECOVERY SEQUENCE
    BYDDER, GM
    YOUNG, IR
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (04) : 659 - 675
  • [3] Fat-Suppression Techniques for 3-T MR Imaging of the Musculoskeletal System
    Del Grande, Filippo
    Santini, Francesco
    Herzka, Daniel A.
    Aro, Michael R.
    Dean, Cooper W.
    Gold, Garry E.
    Carrino, John A.
    [J]. RADIOGRAPHICS, 2014, 34 (01) : 217 - 233
  • [4] SIMPLE PROTON SPECTROSCOPIC IMAGING
    DIXON, WT
    [J]. RADIOLOGY, 1984, 153 (01) : 189 - 194
  • [5] MULTIPOINT DIXON TECHNIQUE FOR WATER AND FAT PROTON AND SUSCEPTIBILITY IMAGING
    GLOVER, GH
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (05): : 521 - 530
  • [6] 3-POINT DIXON TECHNIQUE FOR TRUE WATER FAT DECOMPOSITION WITH BO INHOMOGENEITY CORRECTION
    GLOVER, GH
    SCHNEIDER, E
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (02) : 371 - 383
  • [7] SEPARATION OF FAT AND WATER IN FAST SPIN-ECHO MR-IMAGING WITH THE 3-POINT DIXON TECHNIQUE
    HARDY, PA
    HINKS, RS
    TKACH, JA
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (02): : 181 - 185
  • [8] Fast Spin-Echo Triple Echo Dixon: Initial Clinical Experience With a Novel Pulse Sequence for Simultaneous Fat-Suppressed and Nonfat-Suppressed T2-Weighted Spine Magnetic Resonance Imaging
    Low, Russell N.
    Austin, Matthew J.
    Ma, Jingfei
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 33 (02) : 390 - 400
  • [9] Improving Fat-Suppressed T2-Weighted Imaging of the Head and Neck with 2 Fast Spin-Echo Dixon Techniques: Initial Experiences
    Ma, J.
    Jackson, E. F.
    Kumar, A. J.
    Ginsberg, L. E.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (01) : 42 - 45
  • [10] T2-weighted spine Imaging with a fast three-point Dixon technique: Comparison with chemical shift selective fat suppression
    Ma, JF
    Singh, SK
    Kumar, AJ
    Leeds, NE
    Zhan, JP
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (06) : 1025 - 1029