Organ dose in cardiac dual-energy computed tomography: a Monte Carlo study

被引:0
|
作者
Pakravan, Delaram [1 ]
Mofrad, Farshid Babapour [1 ]
Deevband, Mohammad Reza [2 ]
Ghorbani, Mahdi [2 ]
Pouraliakbar, Hamidreza [3 ]
机构
[1] Islamic Azad Univ, Dept Med Radiat Engn, Sci & Res Branch, Tehran, Iran
[2] Shahid Beheshti Univ Med Sci, Sch Med, Biomed Engn & Med Phys Dept, Tehran, Iran
[3] Iran Univ Med Sci, Rajaie Cardiovasc Med & Res Ctr, Tehran, Iran
关键词
Monte Carlo simulation; Dual-energy computed tomography; Cardiac; Organ dose; Dose index; MULTIDETECTOR CT MDCT; IMAGE QUALITY; ANGIOGRAPHY; SCANNER; PATIENT; MODEL;
D O I
10.1007/s13246-021-01098-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dual-energy computed tomography (DECT) has appeared as a novel approach with the aim of evaluating artery-related diseases. With the advent of DECT, concerns have been raised about the induction of diseases such as cancer due to high radiation exposure of patients. Therefore, the dose received by patients in DECT should be considered. The parameter most commonly used for patient dosimetry is the effective dose (ED). The purpose of this study is to model and validate a DECT scanner by a developed MCNP Monte Carlo code and to calculate the organ doses, the ED, and the conversion factor (k-factor) used in determining ED in the cardiac imaging protocol. To validate the DECT scanner simulation, a standard dosimetry body phantom was modeled in two radiation modes of single energy CT and DECT. The results of simulated CT dose index (CTDI) were compared with those of ImPACT or measurement data. Then dosimetry phantom was replaced by the male and female ORNL phantoms and the organ doses were calculated. The organ doses were also calculated by ImPACT dose software. In the initial validation stage, the minimum and maximum observed relative differences between results of MNCP simulation and measured were 2.77% and 5.79% for the central CTDI and 1.91% and 5.83% for the averaged peripheral CTDI, respectively. The mean ED of simulation and the ImPACT were 3.23 and 5.55 mSv/100 mAs, and the mean k-factor was 0.016 and 0.032 mSv mGy(-1) cm(-1) in the male and female phantoms, respectively. The k-factor obtained for males is close to the currently used k-factor, but the k-factor for females is almost twice.
引用
收藏
页码:157 / 166
页数:10
相关论文
共 50 条
  • [41] Study of the Natural Crystalline Lens Characteristics Using Dual-Energy Computed Tomography
    Sachs, Jeffrey R.
    Nahmias, Javier A.
    Hiatt, Kevin D.
    Bomar, James G.
    West, Thomas G.
    Bunch, Paul M.
    Benayoun, Marc D.
    Lack, Chris
    Thompson, Atalie C.
    DIAGNOSTICS, 2022, 12 (11)
  • [42] Thorax organ dose estimation in computed tomography based on patient CT data using Monte Carlo simulation
    Mohammadi, Gh. R. Fallah
    Alam, N. Riyahi
    Geraily, Gh.
    Paydar, R.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2016, 14 (04): : 313 - 321
  • [43] Spectral Signal Density of Carotid Plaque Using Dual-Energy Computed Tomography
    Reynoso, Exequiel
    Rodriguez-Granillo, Gaston A.
    Capunay, Carlos
    Deviggiano, Alejandro
    Meli, Francisco
    Carrascosa, Patricia
    JOURNAL OF NEUROIMAGING, 2017, 27 (05) : 511 - 516
  • [44] Dual-Energy Computed Tomography in Urological Diseases: A Narrative Review
    Coppola, Andrea
    Tessitore, Luigi
    Fontana, Federico
    Piacentino, Filippo
    Recaldini, Chiara
    Minenna, Manuela
    Capogrosso, Paolo
    Minici, Roberto
    Lagana, Domenico
    Ierardi, Anna Maria
    Carrafiello, Gianpaolo
    D'Angelo, Fabio
    Carcano, Giulio
    Cacioppa, Laura Maria
    Deho, Federico
    Venturini, Massimo
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (14)
  • [45] Adaptive noise correction of dual-energy computed tomography images
    Rafael Simon Maia
    Christian Jacob
    Amy K. Hara
    Alvin C. Silva
    William Pavlicek
    J. Ross Mitchell
    International Journal of Computer Assisted Radiology and Surgery, 2016, 11 : 667 - 678
  • [46] Dual-Energy Computed Tomography Lung in patients of Pulmonary Tuberculosis
    Khan, Ahmad Umar
    Khanduri, Sachin
    Tarin, Zikra
    Abbas, Syed Zain
    Husain, Mushahid
    Singh, Anchal
    Yadav, Poonam
    Jain, Shreshtha
    JOURNAL OF CLINICAL IMAGING SCIENCE, 2020, 10
  • [47] Comparing image quality of single- and dual-energy computed tomography of the brain
    Dodig, Doris
    Kovacic, Slavica
    Matana Kastelan, Zrinka
    Zuza, Iva
    Benic, Filip
    Slaven, Jurkovic
    Miletic, Damir
    Rumboldt, Zoran
    NEURORADIOLOGY JOURNAL, 2020, 33 (03): : 259 - 266
  • [48] Efficiency of dual-energy computed tomography enterography in the diagnosis of Crohn's disease
    Chen, Jinghao
    Zhou, Jie
    Yang, Jushun
    Cong, Ruochen
    Sun, Jinjie
    Xiao, Jing
    Shi, Jianhua
    He, Bosheng
    BMC MEDICAL IMAGING, 2021, 21 (01)
  • [49] Traumatic bone marrow lesions in dual-energy computed tomography
    Ren, Qiuping
    Tang, Deqiu
    Xiong, Zhiyuan
    Zhao, Heng
    Zhang, Shuixing
    INSIGHTS INTO IMAGING, 2022, 13 (01)
  • [50] Dual-energy computed tomography in crystalline arthritis: knowns and unknowns
    Pascart, Tristan
    Budzik, Jean-Francois
    CURRENT OPINION IN RHEUMATOLOGY, 2022, 34 (02) : 103 - 110