Organ dose in cardiac dual-energy computed tomography: a Monte Carlo study

被引:0
|
作者
Pakravan, Delaram [1 ]
Mofrad, Farshid Babapour [1 ]
Deevband, Mohammad Reza [2 ]
Ghorbani, Mahdi [2 ]
Pouraliakbar, Hamidreza [3 ]
机构
[1] Islamic Azad Univ, Dept Med Radiat Engn, Sci & Res Branch, Tehran, Iran
[2] Shahid Beheshti Univ Med Sci, Sch Med, Biomed Engn & Med Phys Dept, Tehran, Iran
[3] Iran Univ Med Sci, Rajaie Cardiovasc Med & Res Ctr, Tehran, Iran
关键词
Monte Carlo simulation; Dual-energy computed tomography; Cardiac; Organ dose; Dose index; MULTIDETECTOR CT MDCT; IMAGE QUALITY; ANGIOGRAPHY; SCANNER; PATIENT; MODEL;
D O I
10.1007/s13246-021-01098-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dual-energy computed tomography (DECT) has appeared as a novel approach with the aim of evaluating artery-related diseases. With the advent of DECT, concerns have been raised about the induction of diseases such as cancer due to high radiation exposure of patients. Therefore, the dose received by patients in DECT should be considered. The parameter most commonly used for patient dosimetry is the effective dose (ED). The purpose of this study is to model and validate a DECT scanner by a developed MCNP Monte Carlo code and to calculate the organ doses, the ED, and the conversion factor (k-factor) used in determining ED in the cardiac imaging protocol. To validate the DECT scanner simulation, a standard dosimetry body phantom was modeled in two radiation modes of single energy CT and DECT. The results of simulated CT dose index (CTDI) were compared with those of ImPACT or measurement data. Then dosimetry phantom was replaced by the male and female ORNL phantoms and the organ doses were calculated. The organ doses were also calculated by ImPACT dose software. In the initial validation stage, the minimum and maximum observed relative differences between results of MNCP simulation and measured were 2.77% and 5.79% for the central CTDI and 1.91% and 5.83% for the averaged peripheral CTDI, respectively. The mean ED of simulation and the ImPACT were 3.23 and 5.55 mSv/100 mAs, and the mean k-factor was 0.016 and 0.032 mSv mGy(-1) cm(-1) in the male and female phantoms, respectively. The k-factor obtained for males is close to the currently used k-factor, but the k-factor for females is almost twice.
引用
收藏
页码:157 / 166
页数:10
相关论文
共 50 条
  • [1] Organ dose in cardiac dual-energy computed tomography: a Monte Carlo study
    Delaram Pakravan
    Farshid Babapour Mofrad
    Mohammad Reza Deevband
    Mahdi Ghorbani
    Hamidreza Pouraliakbar
    Physical and Engineering Sciences in Medicine, 2022, 45 : 157 - 166
  • [2] A new Monte Carlo tool for organ dose estimation in computed tomography
    Adrien, C.
    Le Loirec, C.
    Dreuil, S.
    Bordy, J. -M.
    RADIOPROTECTION, 2020, 55 (02) : 123 - 134
  • [3] Cardiac Applications of Dual-Energy Computed Tomography
    Lempel, Matthew
    Frishman, William H.
    CARDIOLOGY IN REVIEW, 2019, 27 (04) : 208 - 210
  • [4] Evaluation of organ dose, effective dose and cancer risk of head and neck dual-energy computed tomography
    Safari, Arash
    Falahati, Farzaneh
    Mahdavi, Maziyar
    Akbarnejad, Mohamad
    Mohammadzadeh, Ali
    RADIATION PHYSICS AND CHEMISTRY, 2024, 218
  • [5] Dual-Energy Computed Tomography in Cardiac Imaging
    Boettcher, Benjamin
    Zsarnoczay, Emese
    Varga-Szemes, Akos
    Schoepf, Uwe Joseph
    Meinel, Felix G.
    van Assen, Marly
    De Cecco, Carlo N.
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 995 - 1009
  • [6] Optimal Contrast of Computed Tomography Portal Venography Using Dual-Energy Computed Tomography
    Wang, Qi
    Shi, Gaofeng
    Liu, Xiang
    Wu, Runze
    Wang, Shijie
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (02) : 142 - 148
  • [7] Dual-Energy Computed Tomography: Is There a Penalty in Image Quality and Radiation Dose Compared With Single-Energy Computed Tomography?
    Thomas, Christoph
    Ketelsen, Dominik
    Tsiflikas, Ilias
    Reimann, Anja
    Brodoefel, Harald
    Claussen, Claus D.
    Heuschmid, Martin
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2010, 34 (02) : 309 - 315
  • [8] Dose uncertainty due to energy dependence in dual-energy computed tomography
    Kawahara, Daisuke
    Toyoda, Takatoshi
    Yokomachi, Kazushi
    Fujioka, Chikako
    Nagata, Yasushi
    POLISH JOURNAL OF RADIOLOGY, 2023, 88 : e270 - e274
  • [9] Dual-energy Computed Tomography Imaging of the Aorta
    Vlahos, Ioannis
    Godoy, Myrna C. B.
    Naidich, David P.
    JOURNAL OF THORACIC IMAGING, 2010, 25 (04) : 289 - 300
  • [10] Estimating organ dose in computed tomography using tube current modulation: A Monte Carlo simulation
    Hosseinzadeh, V
    Ghaffari, H.
    Rezaeyan, A.
    Deilami, S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (03): : 575 - 581