Design of high performance QRS complex detector for wearable healthcare devices using biorthogonal spline wavelet transform

被引:17
|
作者
Berwal, Deepak [1 ]
Kumar, Ashish [2 ]
Kumar, Yogendera [3 ]
机构
[1] Indian Inst Technol, Dept Elect Engn, Bombay 400076, Maharashtra, India
[2] Bennett Univ, Sch Engn & Appl Sci, Elect & Commun Div, Greater Noida 201310, UP, India
[3] Galgotias Univ, Sch Elect Elect & Commun Engn, VLSI Div, Plot 2,Sect 17-A, Greater Noida 201309, UP, India
关键词
Wearable healthcare devices; ECG; QRS complex detector; QRS complex; Biorthogonal wavelet transform; Heart rate variability; ECG-ON-CHIP; COMPRESSION;
D O I
10.1016/j.isatra.2018.08.002
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A high performance QRS complex detector applicable for wearable healthcare devices is proposed in the present work. Since, higher SNR results in better detection accuracy and lesser number of coefficients reduces the hardware resources as well as power dissipation during on chip implementation. Biorthogonal spline wavelet transform is chosen for the proposed detector as it has high signal to noise ratio (SNR) and uses only four coefficients for decomposition. In the proposed approach, a Biorthogonal wavelet filter bank with fourth level decomposition is first used to separate the different frequency components and then a fourth level wavelet filter bank is used to get the denoised electrocardiogram (ECG) signals. Wavelet filter bank outputs are multiplied and soft threshold method is applied to get the QRS complex peaks by the QRS complex peak detector block. Add and shift multiplier used in the earlier designs has been replaced by a Booth multiplier in our approach to achieve the higher performance. Booth multiplier and QRS complex peak detector blocks have been designed for low hardware complexity, high performance and accurate detection of the QRS complex peaks. Time interval between the consecutive QRS peaks is calculated using the R-R peak time calculator block and the heart rate (HR) by the HR calculator block. Heart Rate Variability (HRV) and arrhythmia are detected based on these heart rate calculations. Proposed design has been tested for its robustness on multiple datasets (namely, MIT-BIH arrhythmia, MIT-BIH noise stress test, and MIT-BIH atrial fibrillation databases). Sensitivity of 99.31%, positive predictivity of 99.19% and the Detection Error Rate (DER) of 1.49% shown by the proposed design makes it preferable for QRS complex detectors used in wearable healthcare devices.
引用
收藏
页码:222 / 230
页数:9
相关论文
共 50 条
  • [1] Design of High-Performance ECG Detector for Implantable Cardiac Pacemaker Systems using Biorthogonal Wavelet Transform
    Kumar, Ashish
    Berwal, Deepak
    Kumar, Yogendera
    CIRCUITS SYSTEMS AND SIGNAL PROCESSING, 2018, 37 (09) : 3995 - 4014
  • [2] Design of High-Performance ECG Detector for Implantable Cardiac Pacemaker Systems using Biorthogonal Wavelet Transform
    Ashish Kumar
    Deepak Berwal
    Yogendera Kumar
    Circuits, Systems, and Signal Processing, 2018, 37 : 3995 - 4014
  • [3] QRS detector pre-processing using the complex wavelet transform
    Schuck, A
    Wisbeck, JO
    PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 2590 - 2593
  • [4] Wavelet transform-based QRS complex detector
    Kadambe, S
    Murray, R
    Boudreaux-Bartels, GF
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (07) : 838 - 848
  • [5] A Real-Time QRS Detector Based on Discrete Wavelet Transform and Cubic Spline Interpolation
    Zheng, Huabin
    Wu, Jiankang
    TELEMEDICINE JOURNAL AND E-HEALTH, 2008, 14 (08): : 809 - 815
  • [6] A new complex wavelet transform by using RI-spline wavelet
    Zhang, Z
    Fujiwara, H
    Toda, H
    Kawabata, H
    2004 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH, AND SIGNAL PROCESSING, VOL II, PROCEEDINGS: SENSOR ARRAY AND MULTICHANNEL SIGNAL PROCESSING SIGNAL PROCESSING THEORY AND METHODS, 2004, : 937 - 940
  • [7] Multiplierless design of biorthogonal dual-tree complex wavelet transform using lifting scheme
    Abbas, Adeel
    Tran, Trac D.
    2006 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING, ICIP 2006, PROCEEDINGS, 2006, : 1605 - +
  • [8] A new mathematical based QRS detector using continuous wavelet transform
    Ghaffarl, A.
    Golbayani, H.
    Ghasemi, M.
    COMPUTERS & ELECTRICAL ENGINEERING, 2008, 34 (02) : 81 - 91
  • [9] Detection of QRS complex using mathematical morphology and wavelet transform
    Liu, Shaoying
    Lu, Jilai
    Hao, Li
    Hu, Guangshu
    Qinghua Daxue Xuebao/Journal of Tsinghua University, 2004, 44 (06): : 852 - 855
  • [10] A real-time adaptive wavelet transform-based QRS complex detector
    Rudnicki, Marek
    Strumillo, Pawel
    ADAPTIVE AND NATURAL COMPUTING ALGORITHMS, PT 2, 2007, 4432 : 281 - +