Comparison of Electronic Portal Imaging and Cone Beam Computed Tomography for Position Verification in Children

被引:12
作者
Zaghloul, M. S. [1 ,2 ]
Mousa, A. G. [1 ,2 ]
Eldebawy, E. [1 ,2 ]
Attalla, E. [1 ,2 ]
Shafik, H. [1 ]
Ezzat, S. [3 ]
机构
[1] Childrens Canc Hosp, Dept Radiat Oncol, Cairo, Egypt
[2] Cairo Univ, Natl Canc Inst, Cairo, Egypt
[3] Childrens Canc Hosp, Res Dept, Cairo, Egypt
关键词
Cone beam CT; EPID; paediatric tumours; PTV margin; radiotherapy position verification; set-up error; CURRENT CLINICAL-PRACTICE; SETUP UNCERTAINTIES; RADIATION-THERAPY; PROSTATE-CANCER; QUALITY-CONTROL; RADIOTHERAPY; IRRADIATION; ACCURACY; ERRORS; DEVIATIONS;
D O I
10.1016/j.clon.2010.08.006
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim To compare the accuracy of radiotherapy set-up using an electronic portal imaging device (EPID) versus megavoltage cone beam computed tomography (MV-CBCT) in paediatric patients Materials and methods In total 204 pairs of EPID and MV-CBCT were carried out for 72 patients in the first 3 treatment days and weekly thereafter Results For the whole group the mean systematic EPID set-up errors were 1 8(+/- 1 7) 1 6(+/- 1 3) 1 4(v 1 5) mm and 2 3 (+/- 1 7) 1 6 (+/- 1 3) 2 4(+/- 1 6) mm for MV-CBCT in the longitudinal lateral and vertical directions respectively whereas the mean EPID random errors were 2 0 (+/- 1 7) 1 4 (+/- 15) 1 2 +/-( 16) and 19 (+/- 1 5) 1 5 (+/- 1 3) 2 1 (+/- 1 7) mm for MV-CBCT in the longitudinal lateral and vertical directions respectively For systematic errors of head and neck patients there was a statistically significant difference in the lateral and vertical directions (P = 0 027 0 003) whereas in the non-head and neck patients there was a statistically significant difference in the lateral direction only (P = 0 031) In head and neck patients the mean random errors were significantly different in the vertical and lateral directions whereas in non-head and neck patients they were significantly different in the vertical direction only The larger values alternate between the two modalities The systematic and random errors (detected by EPID and MV-CBCT) were significantly correlated in almost all direction in all tumour sites Conclusions The comparison between set-up error in EPID and MV-CBCT was not in favour of any of the two modalities However the two modalities were strongly correlated but fairly agreed and the differences between the shifts reported were small and hardly influenced the recommended planning target volume margin (C) 2010 The Royal College of Radiologists Published by Elsevier Ltd All rights reserved
引用
收藏
页码:850 / 861
页数:12
相关论文
共 29 条
[1]   Dosimetric consequences of the application of off-line setup error correction protocols and a hull-volume definition strategy for intensity modulated radiotherapy of prostate cancer [J].
Baum, C ;
Birkner, M ;
Alber, M ;
Paulsen, F ;
Nüsslin, F .
RADIOTHERAPY AND ONCOLOGY, 2005, 76 (01) :35-42
[2]   SETUP DEVIATIONS IN WEDGED PAIR IRRADIATION OF PAROTID-GLAND AND TONSILLAR TUMORS, MEASURED WITH AN ELECTRONIC PORTAL IMAGING DEVICE [J].
BEL, A ;
KEUS, R ;
VIJLBRIEF, RE ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (02) :153-159
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Kilo-voltage cone-beam computed tomography setup measurements for lung cancer patients; First clinical results and comparison with electronic portal-imaging device [J].
Borst, Gerben R. ;
Sonke, Jan-Jakob ;
Betgen, Anja ;
Remeijer, Peter ;
Van Herk, Marcel ;
Lebesque, Joos V. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (02) :555-561
[5]   A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS) [J].
BOYER, AL ;
ANTONUK, L ;
FENSTER, A ;
VANHERK, M ;
MEERTENS, H ;
MUNRO, P ;
REINSTEIN, LE ;
WONG, J .
MEDICAL PHYSICS, 1992, 19 (01) :1-16
[6]   FITTING OF NORMAL TISSUE TOLERANCE DATA TO AN ANALYTIC-FUNCTION [J].
BURMAN, C ;
KUTCHER, GJ ;
EMAMI, B ;
GOITEIN, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01) :123-135
[7]   A quality control study of the accuracy of patient positioning in irradiation of pelvic fields [J].
Creutzberg, CL ;
Althof, VGM ;
deHoog, M ;
Visser, AG ;
Huizenga, H ;
Wijnmaalen, A ;
Levendag, PC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (03) :697-708
[8]   TIME TREND OF PATIENT SETUP DEVIATIONS DURING PELVIC IRRADIATION USING ELECTRONIC PORTAL IMAGING [J].
ELGAYED, AAH ;
BEL, A ;
VIJLBRIEF, R ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) :162-171
[9]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829
[10]  
Gifford Kent A, 2002, J Appl Clin Med Phys, V3, P26, DOI 10.1120/1.1434315