Individual bone cement spacers (IBCS) for septic hip revision - preliminary report

被引:27
作者
Schoellner, C
Fuerderer, S
Rompe, JD
Eckardt, A
机构
[1] Johannes Gutenberg Univ Mainz, Sch Med, Dept Orthoped Surg, D-55101 Mainz, Germany
[2] Univ Kliniken Koln, Orthopad Univ & Poliklin, D-50931 Cologne, Germany
关键词
arthroplasty; septic revision; spacer; two-stage procedure; biomechanics;
D O I
10.1007/s00402-003-0502-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Two-stage septic hip revision with intermediate resection arthroplasty leads to temporary poor function and difficult reimplantation. Antibiotic-loaded cement spacers improve patient mobility and maintain stability of the joint as well as length of the limb. Prefabricated spacers are readily available but lack adaptability. We developed modular moulds separately for stem and head to create more individual spacers with optimised femoral press-fit and smooth head surface. Methods: Moulds are fabricated preoperatively by forming 4 mm polyethylene over trial stems and cups under vacuum (MS 30, Center-pulse, Switzerland). After gas sterilization, Refobacin Palacos (Merck, Darmstadt, Germany) is poured into stem and head moulds of appropriate size. A double K-wire is inserted as a rod. Fluoroscopy is used to check for air inclusions, which might impair resistance. For mechanical resistance testing trials, 5 spacers were resin-fixed with 80% stem insertion and loaded in a craniocaudal direction at 20 N/s (universal testing machine, Zwick, Germany) until failure. Results: The static testing showed an average failure load of 1550 N (range 1350-2000 N). The K-wires prevented complete dislocation of the fragments. Conclusion: Partial weight-bearing is possible with this system, but a fall might lead to an acute fracture. New materials for the core can be tested with our biomechanical model. Following the presented craniocaudal static testing, torsion testing and dynamic testing in Ringer's solution are planned.
引用
收藏
页码:254 / 259
页数:6
相关论文
共 31 条
  • [21] Migaud H, 1997, REV CHIR ORTHOP, V83, P466
  • [22] NELSON CL, 1992, CLIN ORTHOP RELAT R, V284, P303
  • [23] New uses for gentamicin-impregnated polymethyl methacrylate spacers in two-stage revision hip arthroplasty
    Oxborrow, NJ
    Stamer, J
    Andrews, M
    Stone, MH
    [J]. JOURNAL OF ARTHROPLASTY, 1997, 12 (06) : 709 - 710
  • [24] Ozaki T, 1998, J Orthop Sci, V3, P163
  • [25] RAUT VV, 1995, CLIN ORTHOP RELAT R, V321, P202
  • [26] An inexpensive molding method for antibiotic-impregnated cement spacers in infected total hip arthroplasty
    Ries, MD
    Jergesen, H
    [J]. JOURNAL OF ARTHROPLASTY, 1999, 14 (06) : 764 - 765
  • [27] Necessary steps for the implementation of the DRG Systems
    Schmid, M
    Hopf, C
    [J]. ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2001, 139 (04): : 279 - 286
  • [28] In vitro evaluation of antibiotic elution from polymethylmethacrylate (PMMA) and mechanical assessment of antibiotic-PMMA composites
    Weisman, DL
    Olmstead, ML
    Kowalski, JJ
    [J]. VETERINARY SURGERY, 2000, 29 (03) : 245 - 251
  • [29] Treatment of infection associated with segmental bone loss in the proximal part of the femur in two stages with use of an antibiotic-loaded interval prosthesis
    Younger, ASE
    Duncan, CP
    Masri, BA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (01) : 60 - 69
  • [30] The outcome of two-stage arthroplasty using a custom-made interval spacer to treat the infected hip
    Younger, ASE
    Duncan, CP
    Masri, BA
    McGraw, RW
    [J]. JOURNAL OF ARTHROPLASTY, 1997, 12 (06) : 615 - 623