Pre-clinical laboratory evaluation of the new 'AF' arterial line filter range

被引:1
作者
Yarham, Gemma [1 ,2 ]
Mulholland, John [1 ,2 ]
机构
[1] Essex Cardiothorac Ctr, Dept Clin Perfus Sci, London, England
[2] Imperial Coll Hlth Sci Ctr, Dept Clin Perfus Res, London, England
来源
PERFUSION-UK | 2010年 / 25卷 / 04期
关键词
cardiopulmonary bypass; arterial line filter; non-physiological surface area; prime volume; pressure drop; micro air handling; gross air handling; CARDIOPULMONARY BYPASS; AIR; MICROEMBOLI; MICROBUBBLES; EMBOLISM; SURGERY;
D O I
10.1177/0267659110377679
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: The presence of emboli was recognised relatively early in the history of open heart surgery. The emboli produced during cardiopulmonary bypass have the predisposition to distribute into, and ultimately obstruct, microvessels of all tissues. The Sorin Group has recently developed a new range of arterial line filters. Before the Sorin AF range of filters was released for pre-launch clinical trials, our group performed in vitro laboratory testing of the AF range against a selection of commercially available filters on the global market. Results: The Sorin AF620 and AF640 demonstrate both the smallest prime volume and smallest surface contact area (92ml and 290cm(2), respectively). The results of the GME Handling Efficiency experiments ranged by 39.6%, from 95.9% to 56.3%. In terms of an air bolus handling, the results of the Limit Bolus experiment ranged by 97ml, from 147.5ml down to 50ml. The pressure drop across all the filters was measured under steady state experimental conditions. All of the above investigations were considered against surface area and prime volume. Conclusion: It is clear from the results that some commercially available arterial line filters perform better than others, not only in overall performance, but also with regard to individual characteristics. Evaluating arterial line filters for hospital-specific use has to balance pressure drop, surface area, micro air handling, prime volume and gross air handling; all points need to be considered. In the AF620 and AF640, Sorin boast that they are the two smallest prime and smallest surface area filters commercially available on the global market. The Sorin AF filter range performs well in all of the areas we investigated and will be a competitive option for centres, irrespective of which characteristics they use to evaluate and select their arterial line filter.
引用
收藏
页码:267 / 276
页数:10
相关论文
共 15 条
  • [1] Clinical evaluation of the air removal characteristics of an oxygenator with integrated arterial filter in a minimized extracorporeal circuit
    Stehouwer, Marco C.
    Boers, Chris
    de Vroege, Roel
    Kelder, Johannes C.
    Yilmaz, Alaaddin
    Bruins, Peter
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (04) : 374 - 382
  • [2] Clinical Evaluation of New Generation Oxygenators With Integrated Arterial Line Filters for Cardiopulmonary Bypass
    Onorati, Francesco
    Santini, Francesco
    Raffin, Federica
    Menon, Tiziano
    Graziani, Maria S.
    Chiominto, Bartolomeo
    Milano, Aldo
    Faggian, Giuseppe
    Mazzucco, Alessandro
    ARTIFICIAL ORGANS, 2012, 36 (10) : 875 - 885
  • [3] Pre-clinical evaluation of new dibucaine formulations for preventive analgesia
    Furlan, Beatriz
    de Melo, Beatriz T.
    Papini, Juliana Z. B.
    Sperandio, Marcelo
    Oliveira, Juliana D.
    de Paula, Eneida
    Cereda, Cintia M. S.
    Tofoli, Giovana R.
    JOURNAL OF LIPOSOME RESEARCH, 2021, 31 (03) : 230 - 236
  • [4] Evaluation of air handling in a new generation neonatal oxygenator with integral arterial filter
    Gomez, D.
    Preston, T. J.
    Olshove, V. F.
    Phillips, A. B.
    Galantowicz, M. E.
    PERFUSION-UK, 2009, 24 (02): : 107 - 112
  • [5] Evaluation of the Maquet Neonatal and Pediatric Quadrox I with an integrated arterial line filter during cardiopulmonary bypass
    Melchior, R. W.
    Schiavo, K.
    Frey, T.
    Rogers, D.
    Patel, J.
    Chelnik, K.
    Rosenthal, T.
    PERFUSION-UK, 2012, 27 (05): : 399 - 406
  • [6] Clinical evaluation of emboli removal by integrated versus non-integrated arterial filters in new generation oxygenators
    Jabur, Ghazwan N. S.
    Sidhu, Karishma
    Willcox, Timothy W.
    Mitchell, Simon J.
    PERFUSION-UK, 2016, 31 (05): : 409 - 417
  • [7] A clinical evaluation of the Maquet Quadrox-i Neonatal oxygenator with integrated arterial filter
    Ginther, R. M., Jr.
    Gorney, R.
    Cruz, R.
    PERFUSION-UK, 2013, 28 (03): : 194 - 199
  • [8] Clinical evaluation of the air-handling properties of contemporary oxygenators with integrated arterial filter
    Stehouwer, Marco C.
    Legg, Kristina R.
    de Vroege, Roel
    Kelder, Johannes C.
    Hofman, Erik
    de Mol, Bastian A.
    Bruins, Peter
    PERFUSION-UK, 2017, 32 (02): : 118 - 125
  • [9] Evaluation of leukocyte-reducing arterial line filter (LG6) for postoperative lung function, using cardiopulmonary bypass
    Ohto T.
    Yamamoto F.
    Nakajima N.
    The Japanese Journal of Thoracic and Cardiovascular Surgery, 2000, 48 (5): : 295 - 300
  • [10] Clinical evaluation of the new BMU 40 in-line blood analysis monitor
    Schaarschmidt, Jan
    Seeburger, Joerg
    Borger, Michael A.
    Grosse, F. Oliver
    Kraemer, Klaus
    Mohr, Friedrich W.
    PERFUSION-UK, 2009, 24 (04): : 277 - 286