To compare central corneal thickness measurements obtained by Pentacam with those obtained by IOLMaster 700, Cirrus anterior segment optical coherence tomography and Tomey specular microscopy in normal healthy eyes

被引:11
|
作者
Kumar, Kiran K. [1 ]
Prakash, Akshata A. [1 ]
Neeraja, T. G. [1 ]
Adappa, Karishma T. [1 ]
Prabha, Chandra T. S. [1 ]
Gangasagara, Suresh Babu [1 ]
机构
[1] BMCRI, Minto Ophthalm Hosp, Reg Inst Ophthalmol, C96,5th Cross,8th Main Vasanthnagar, Bengaluru 560052, Karnataka, India
关键词
Agreement; AS-OCT; central corneal thickness; IOL master; pentacam; specular microscopy; ULTRASOUND PACHYMETRY; SCHEIMPFLUG; AGREEMENT; OCT;
D O I
10.4103/ijo.IJO_3364_20
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare central corneal thickness measurements obtained by Pentacam with those obtained by IOL Master 700, Cirrus Anterior segment optical coherence tomography and Tomey Specular microscopy in normal healthy eyes. Methods: Two hundred and six eyes of healthy subjects were included in the study. Each subject was assessed by four different methods of measuring central corneal thickness using Pentacam, IOL Master 700, Cirrus AS-OCT and Tomey Specular microscopy by a single examiner. Results: The mean CCT [+/- standard deviation (SD)] for Pentacam, IOL Master 700, Cirrus AS-OCT and Tomey Specular microscopy were Pentacam (Oculus), AS-OCT (Cirrus), IOL Master 700 and Specular microscopy (Tomey) were 523.75 (+/- 27.75), 525.29 (+/- 28.81),517.13 (+/- 28.43) and 512.82 (+/- 27.60) mu m, respectively. All the means were significantly different from one another (P < 0.000). The differences between pairs of mean central corneal thickness (CCT) for Pentacam and IOL Master, Pentacam and anterior segment-optical coherence tomography (AS-OCT), and Pentacam and Specular microscopy are statistically significant. Bland-Altman plots showed that pentacam and IOL Master 700 have the closest agreement, followed by AS-OCT. Specular microscopy was found to have the poorest agreement with Pentacam. Conclusion: We found that CCT measurements of Pentacam did not correlate with measurements of IOL Master, or AS-OCT or Specular microscopy. In clinical practice, the devices analyzed should not be used interchangeably due to low agreement regarding CCT values.
引用
收藏
页码:1713 / 1717
页数:5
相关论文
共 15 条
  • [1] Central corneal thickness measurements obtained with anterior segment spectral domain optical coherence tomography compared to ultrasound pachymetry in healthy subjects
    Vollmer, Lori
    Sowka, Joseph
    Pizzimenti, Joseph
    Yu, Xinha
    OPTOMETRY-JOURNAL OF THE AMERICAN OPTOMETRIC ASSOCIATION, 2012, 83 (05) : 167 - 172
  • [2] Comparison of central corneal thickness measurements in corneal edema using ultrasound pachymetry, Visante anterior-segment optical coherence tomography, Cirrus optical coherence tomography, and Pentacam Scheimpflug camera tomography
    Wongchaisuwat, Nida
    Metheetrairat, Ankana
    Chonpimai, Pratuangsri
    Nujoi, Waree
    Prabhasawat, Pinnita
    CLINICAL OPHTHALMOLOGY, 2018, 12 : 1865 - 1873
  • [3] Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography
    Artuc, Tuncay
    Batur, Muhammed
    SAUDI JOURNAL OF OPHTHALMOLOGY, 2023, 37 (01) : 1 - 5
  • [4] Comparison between Pentacam-HR and optical coherence tomography central corneal thickness measurements in healthy feline eyes
    Cleymaet, Allison M.
    Hess, Ann M.
    Freeman, Kate S.
    VETERINARY OPHTHALMOLOGY, 2016, 19 : 105 - 114
  • [5] Comparison of corneal measurements in normal and keratoconus eyes using Anterior Segment Optical Coherence Tomography (AS-OCT) and Pentacam HR topographer
    Said, Omar M.
    Kamal, Mahmoud
    Tawfik, Sara
    Saif, Ahmed Tamer Sayed
    BMC OPHTHALMOLOGY, 2023, 23 (01)
  • [6] Comparison of corneal measurements in normal and keratoconus eyes using Anterior Segment Optical Coherence Tomography (AS-OCT) and Pentacam HR topographer
    Omar M. Said
    Mahmoud Kamal
    Sara Tawfik
    Ahmed Tamer Sayed Saif
    BMC Ophthalmology, 23
  • [7] Corneal Thickness and Volume Measurements by Swept Source Anterior Segment Optical Coherence Tomography in Normal Subjects
    Fukuda, Reina
    Usui, Tomohiko
    Miyai, Takashi
    Mori, Yosai
    Miyata, Kazunori
    Amano, Shiro
    CURRENT EYE RESEARCH, 2013, 38 (05) : 531 - 536
  • [8] Comparison of central corneal thickness measurements by visante anterior segment optical coherence tomography with ultrasound pachymetry
    Zhao, Paul S.
    Wong, Tien Yin
    Wong, Wan-Ling
    Saw, Seang-Mei
    Aung, Tin
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (06) : 1047 - 1049
  • [9] Intra and inter-user reliability of central corneal thickness measurements obtained in healthy feline eyes using a portable spectral-domain optical coherence tomography device
    Alario, Anthony F.
    Pirie, Christopher G.
    VETERINARY OPHTHALMOLOGY, 2013, 16 (06) : 446 - 450
  • [10] Normal corneal thickness measurements in pigmented rabbits using spectral-domain anterior segment optical coherence tomography
    Wang, Xiaogang
    Wu, Qiang
    VETERINARY OPHTHALMOLOGY, 2013, 16 (02) : 130 - 134