Screening for lupus anticoagulant: improving the performance of the lupus-sensitive PTT-LA

被引:2
作者
Luginbuehl, R.
Barizzi, G.
Sulzer, I.
Laemmle, B.
Alberio, L. [1 ]
机构
[1] Univ Hosp Bern, Inselspital, Deparment Haematol, CH-3010 Bern, Switzerland
基金
瑞士国家科学基金会;
关键词
Lupus; anticoagulant; screening; sensitivity; aPTT; ratio; PARTIAL THROMBOPLASTIN TIME; ANTIPHOSPHOLIPID SYNDROME; MONOCLONAL-ANTIBODIES; IDENTIFICATION; THROMBOSIS; DIAGNOSIS; REAGENTS; SAMPLES; UPDATE; RISK;
D O I
10.1111/j.1751-553X.2010.01262.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
P>Introduction: The aim of the present work was to verify whether calculating a ratio between clotting times obtained with the sensitive PTT-LA and a less sensitive activated partial thromboplastin time (aPTT)-reagent may represent a valuable aPTT-based screening strategy for lupus anticoagulants (LA). Methods: For the pilot study, plasma samples from normal subjects (n = 15) and from patients with LA (n = 10), therapeutic anticoagulation with vitamin K-antagonists (VKA) (n = 15) or unfractionated heparin (n = 15), coagulation factors deficiency (n = 16), and inhibitory antibodies against factor VIII or IX (n = 11) were studied. For the evaluation study, 1553 consecutive plasma samples from nonanticoagulated patients investigated for LA between January 2005 and December 2007 at our institution were studied. Following screening strategies were employed: Pathromtin-SL (aPTT-SL), PTT-LA (aPTT-LA), ratio aPTT-LA/aPTT-SL (aPTT-ratio), and Russell's viper venom (RVV) based LA-Check. LA positive samples were identified by mixing studies and diluted RVV confirmation test (LA-Check/LA-Sure). Results: Pilot study: All screening strategies had a 100% sensitivity, and the aPTT-ratio reached the highest specificity (82%; 95%CI: 74-90%). Within the evaluation study, following sensitivities for LA screening were observed: aPTT-SL 59.0% (95%CI: 57-61%), aPTT-LA 82.1% (95%CI: 80-84%), aPTT-ratio 92.3% (95%CI: 91-94), and LA-Check 83.3% (95%CI: 82-85%). Conclusion: Calculating a ratio between the LA-sensitive PTT-LA and the less sensitive Pathromtin-SL improves the performance of the PTT-LA itself and represents a simple and sensitive aPTT-based integrated strategy for LA screening.
引用
收藏
页码:168 / 175
页数:8
相关论文
共 21 条
[1]  
46Lewis SM., 2001, Dacie and Lewis Practical Hematology
[2]  
ADCOCK DM, 1992, ARCH PATHOL LAB MED, V116, P837
[3]  
Arnout J, 1999, THROMB HAEMOSTASIS, V81, P929
[4]   A rapid screen for lupus anticoagulant with good discrimination from oral anticoagulants, congenital factor deficiency and heparin, is provided by comparing a sensitive and an insensitive APTT reagent [J].
Brancaccio, V ;
Ames, PRJ ;
Glynn, J ;
Iannaccone, L ;
Mackie, IJ .
BLOOD COAGULATION & FIBRINOLYSIS, 1997, 8 (03) :155-160
[5]  
Brandt JT, 1995, THROMB HAEMOSTASIS, V74, P1597
[6]  
DENISMAGDELAINE A, 1995, HAEMOSTASIS, V25, P98
[7]   KAOLIN CLOTTING TIME AND DILUTE RUSSELLS VIPER VENOM TIME DISTINGUISH BETWEEN PROTHROMBIN-DEPENDENT AND BETA-2-GLYCOPROTEIN I-DEPENDENT ANTIPHOSPHOLIPID ANTIBODIES [J].
GALLI, M ;
FINAZZI, G ;
BEVERS, EM ;
BARBUI, T .
BLOOD, 1995, 86 (02) :617-623
[8]   Lupus anticoagulants are stronger risk factors for thrombosis than anticardiolipin antibodies in the antiphospholipid syndrome: a systematic review of the literature [J].
Galli, M ;
Luciani, D ;
Bertolini, G ;
Barbui, T .
BLOOD, 2003, 101 (05) :1827-1832
[9]  
Galli M, 1999, THROMB HAEMOSTASIS, V81, P695
[10]   The importance of locally derived reference ranges and standardized calculation of dilute Russell's viper venom time results in screening for lupus anticoagulant [J].
Gardiner, C ;
Mackie, IJ ;
Malia, RG ;
Jones, DW ;
Winter, M ;
Leeming, D ;
Taberner, DA ;
Machin, SJ ;
Greaves, M .
BRITISH JOURNAL OF HAEMATOLOGY, 2000, 111 (04) :1230-1235