Dosimetric impact of tantalum markers used in the treatment of uveal melanoma with proton beam therapy

被引:60
作者
Newhauser, Wayne D.
Koch, Nicholas C.
Fontenot, Jonas D.
Rosenthal, Stanley J.
Gombos, Dan S.
Fitzek, Markus M.
Mohan, Radhe
机构
[1] Univ Texas, MD Anderson Canc Ctr, Unit 94, Dept Head & Neck Surg,Sect Ophthalmol, Houston, TX 77030 USA
[2] Univ Texas, Grad Sch Biomed Sci, Houston, TX 77030 USA
[3] Massachusetts Gen Hosp, Boston, MA 02114 USA
[4] Midwest Proton Radiotherapy Inst, Bloomington, IN 47408 USA
[5] Indiana Univ, Sch Med, Indianapolis, IN 46202 USA
关键词
D O I
10.1088/0031-9155/52/13/021
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Metallic fiducial markers are frequently implanted in patients prior to external-beam radiation therapy to facilitate tumor localization. There is little information in the literature, however, about the perturbations in proton absorbed-dose distribution these objects cause. The aim of this study was to assess the dosimetric impact of perturbations caused by 2.5 mm diameter by 0.2 mm thick tantalum fiducial markers when used in proton therapy for treating uveal melanoma. Absorbed dose perturbations were measured using radiochromic film and confirmed by Monte Carlo simulations of the experiment. Additional Monte Carlo simulations were performed to study the effects of range modulation and fiducial placement location on the magnitude of the dose shadow for a representative uveal melanoma treatment. The simulations revealed that the fiducials caused perturbations in the absorbed-dose distribution, including absorbed-dose shadows of 22% to 82% in a typical proton beam for treating uveal melanoma, depending on the marker depth and orientation. The clinical implication of this study is that implanted fiducials may, in certain circumstances, cause dose shadows that could lower the tumor dose and theoretically compromise local tumor control. To avoid this situation, fiducials should be positioned laterally or distally with respect to the target volume.
引用
收藏
页码:3979 / 3990
页数:12
相关论文
共 56 条
[51]   Radiochromic film dosimetry for verification of dose distributions delivered with proton-beam radiosurgery [J].
Vatnitsky, SM ;
Schulte, RWM ;
Galindo, R ;
Meinass, HJ ;
Miller, DW .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (10) :1887-1898
[52]   Radiochromic film dosimetry for clinical proton beams [J].
Vatnitsky, SM .
APPLIED RADIATION AND ISOTOPES, 1997, 48 (05) :643-651
[53]  
VAVILOV PV, 1957, ZH EKSP TEOR FIZ, V32, P920
[54]   Secondary electron fluence perturbation by high-Z interfaces in clinical proton beams: a Monte Carlo study [J].
Verhaegen, F ;
Palmans, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (01) :167-183
[55]  
Waters L, 2002, LAUR022607 LOS AL NA
[56]  
ZOGRAFOS L, 1990, STRAHLENTHER ONKOL, V166, P114