A standard test phantom for the performance assessment of magnetic resonance guided high intensity focused ultrasound (MRgHIFU) thermal therapy devices

被引:8
作者
Ambrogio, S. [1 ,2 ]
Baesso, R. M. [2 ]
Bosio, F. [3 ]
Fedele, F. [1 ]
Ramnarine, K., V [1 ]
Zeqiri, B. [2 ]
Miloro, P. [2 ]
机构
[1] Guys & St Thomas NHS Fdn Trust, Med Phys Dept, London, England
[2] Natl Phys Lab, Ultrasound & Underwater Acoust Div, Teddington, Middx, England
[3] Kings Coll London, Sch Biomed Engn & Imaging Sci, London, England
关键词
HIFU; thermal ablation; quality control; phantom; thermal dosimetry; TISSUE-MIMICKING-MATERIAL; TEMPERATURE-MEASUREMENTS; EXPERIMENTAL VALIDATION; ABLATION; MODEL; ABSORPTION; MRGFUS;
D O I
10.1080/02656736.2021.2017023
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Test objects for High Intensity Focused Ultrasound (HIFU) are required for the standardization and definition of treatment, Quality Assurance (QA), comparison of results between centers and calibration of devices. This study describes a HIFU test object which provides temperature measurement as a function of time, in a reference material compatible with Magnetic Resonance (MR) and ultrasound. Materials and methods: T-Type fine wire thermocouples were used as sensors and 5 correction methods for viscous heating artifacts were assessed. The phantom was tested in a MR-HIFU Philips Sonalleve device over a period of 12 months, demonstrating stability and validity to evaluate the performance of the device. Results: The study furnished useful information regarding the MR-HIFU sessions and highlighted potential limitations of the existing QA and monitoring methods. The importance of temperature monitoring along the whole acoustic path was demonstrated as MR Thermometry readings differed in the three MR plane views (coronal, sagittal, transverse), in particular when the focus was near a soft-tissue/bone interface, where there can be an MR signal loss with significant temperature and thermal dose underestimation (138% variation between the three plane views). Conclusions: The test object was easy to use and has potential as a valid tool for training, QA, research and development for MR guided HIFU and potentially ultrasound guided devices.
引用
收藏
页码:57 / 68
页数:12
相关论文
共 50 条
[31]   Sonodynamic therapy and magnetic resonance-guided focused ultrasound: new therapeutic strategy in glioblastoma [J].
Bonosi, Lapo ;
Marino, Silvia ;
Benigno, Umberto Emanuele ;
Musso, Sofia ;
Buscemi, Felice ;
Giardina, Kevin ;
Gerardi, Rosamaria ;
Brunasso, Lara ;
Costanzo, Roberta ;
Iacopino, Domenico Gerardo ;
Maugeri, Rosario .
JOURNAL OF NEURO-ONCOLOGY, 2023, 163 (01) :219-238
[32]   Magnetic Resonance-Guided High-Intensity Focused Ultrasound (MR-HIFU): Overview of Emerging Applications (Part 2) [J].
Siedek, Florian ;
Yeo, Sin Yuin ;
Heijman, Edwin ;
Grinstein, Olga ;
Bratke, Grischa ;
Heneweer, Carola ;
Puesken, Michael ;
Persigehl, Thorsten ;
Maintz, David ;
Gruell, Holger .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2019, 191 (06) :531-539
[33]   Neurosurgical Applications of High-Intensity Focused Ultrasound with Magnetic Resonance Thermometry [J].
Colen, Rivka R. ;
Sahnoune, Iman ;
Weinberg, Jeffrey S. .
NEUROSURGERY CLINICS OF NORTH AMERICA, 2017, 28 (04) :559-+
[34]   Focused ultrasound using a novel targeting method four-tract tractography for magnetic resonance-guided high-intensity focused ultrasound targeting [J].
Feltrin, Fabricio S. ;
Chopra, Rajiv ;
Pouratian, Nader ;
Elkurd, Mazen ;
El-Nazer, Rasheda ;
Lanford, Lauren ;
Dauer, William ;
Shah, Bhavya R. .
BRAIN COMMUNICATIONS, 2022, 4 (06)
[35]   The applicability and efficacy of magnetic resonance-guided high intensity focused ultrasound system in the treatment of primary trigeminal neuralgia [J].
Ma, Yaping ;
Hsu, Grace ;
Zhang, Fugui .
MEDICAL HYPOTHESES, 2020, 139
[36]   Magnetic Resonance-Guided High-Intensity-Focused Ultrasound for Palliation of Painful Skeletal Metastases: A Pilot Study [J].
Chan, Michael ;
Dennis, Kristopher ;
Huang, Yuexi ;
Mougenot, Charles ;
Chow, Edward ;
DeAngelis, Carlo ;
Coccagna, Jennifer ;
Sahgal, Arjun ;
Hynynen, Kullervo ;
Czarnota, Gregory ;
Chu, William .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2017, 16 (05) :570-576
[37]   A Clinically Feasible Treatment Protocol for Magnetic Resonance-Guided High-Intensity Focused Ultrasound Ablation in the Liver [J].
Wijlemans, Joost W. ;
de Greef, Martijn ;
Schubert, Gerald ;
Bartels, Lambertus W. ;
Moonen, Chrit T. W. ;
van den Bosch, Maurice A. A. J. ;
Ries, Mario .
INVESTIGATIVE RADIOLOGY, 2015, 50 (01) :24-31
[38]   Histopathology of breast cancer after magnetic resonance-guided high-intensity focused ultrasound and radiofrequency ablation [J].
Knuttel, Floortje M. ;
Waaijer, Laurien ;
Merckel, Laura G. ;
van den Bosch, Maurice A. A. J. ;
Witkamp, Arjen J. ;
Deckers, Roel ;
van Diest, Paul J. .
HISTOPATHOLOGY, 2016, 69 (02) :250-259
[39]   Boiling histotripsy lesion characterization on a clinical magnetic resonance imaging-guided high intensity focused ultrasound system [J].
Eranki, Avinash ;
Farr, Navid ;
Partanen, Ari ;
Sharma, Karun V. ;
Chen, Hong ;
Rossi, Christopher T. ;
Kothapalli, Satya V. V. N. ;
Oetgen, Matthew ;
Kim, AeRang ;
Negussie, Ayele H. ;
Woods, David ;
Wood, Bradford J. ;
Kim, Peter C. W. ;
Yarmolenko, Pavel S. .
PLOS ONE, 2017, 12 (03)
[40]   Efficacy, Efficiency, and Safety of Magnetic Resonance-Guided High-Intensity Focused Ultrasound for Ablation of Uterine Fibroids: Comparison with Ultrasound-Guided Method [J].
Wang, Yi ;
Wang, Zhi-Biao ;
Xu, Yong-Hua .
KOREAN JOURNAL OF RADIOLOGY, 2018, 19 (04) :724-732