Assessment of heparin binding to the AN69 ST hemodialysis membrane: I. Preclinical studies

被引:29
作者
Chanard, J
Lavaud, S
Paris, B
Toure, F
Rieu, P
Renaux, JL
Thomas, M
机构
[1] CHU, Serv Nephrol, F-51100 Reims, France
[2] Hosp Int & Ind, Lyon, France
[3] CNRS, FRE 2534, Reims, France
关键词
D O I
10.1097/01.mat.0000169119.06419.ed
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The AN69 ST membrane was designed to render the surface of the native polyacrylonitrile polymer less cationic. This was achieved by layering the membrane with the polycationic biopolymer polyethyleneimine. This new membrane is able to bind heparin to its surface, through electrical interactions, without altering the reactivity of the sulfonate groups of the membrane, regularly distributed in the membrane bulk. The kinetics of unfractionated or low-molecular-weight heparins were studied in vitro and in vivo in sheep. Encouraging results were obtained indicating that heparin-coated hemodialyzers are potent anticoagulants. Priming the AN69 ST membrane-equipped hemodialyzer with heparin, as in regular hemodialysis, could allow drastic reduction of heparin consumption in hemodialysis.
引用
收藏
页码:342 / 347
页数:6
相关论文
共 20 条
[1]  
Adam A, 2000, NEPHROLOGIE, V21, P163
[2]   New insights in dialysis membrane biocompatibility: relevance of adsorption properties and heparin binding [J].
Chanard, J ;
Lavaud, S ;
Randoux, C ;
Rieu, P .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2003, 18 (02) :252-257
[3]   ACTIVATION OF COMPLEMENT BY HEMODIALYSIS MEMBRANES - POLYACRYLONITRILE BINDS MORE C3A THAN CUPROPHAN [J].
CHEUNG, AK ;
PARKER, CJ ;
WILCOX, LA ;
JANATOVA, J .
KIDNEY INTERNATIONAL, 1990, 37 (04) :1055-1059
[4]   Effect of membrane composition and structure on solute removal and biocompatibility in hemodialysis [J].
Clark, WR ;
Hamburger, RJ ;
Lysaght, MJ .
KIDNEY INTERNATIONAL, 1999, 56 (06) :2005-2015
[5]   Contact system: A vascular biology modulator with anticoagulant, Profibrinolytic, antiadhesive, and proinflammatory attributes [J].
Colman, RW ;
Schmaier, AH .
BLOOD, 1997, 90 (10) :3819-3843
[6]   Blood coagulation [J].
Dahlback, B .
LANCET, 2000, 355 (9215) :1627-1632
[7]   Changes in platelet, granulocyte, and complement activation during cardiopulmonary bypass using heparin-coated equipment [J].
Fukutomi, M ;
Kobayashi, S ;
Niwaya, K ;
Hamada, Y ;
Kitamura, S .
ARTIFICIAL ORGANS, 1996, 20 (07) :767-776
[8]  
GARED P, 1997, ARTIF ORGANS, V21, P293
[9]   Antiinflammatory effect of heparin-coated circuits with leukocyte-depleting filters in coronary bypass surgery [J].
Hamada, Y ;
Kawachi, K ;
Nakata, T ;
Kohtani, T ;
Takano, S ;
Tsunooka, N .
ARTIFICIAL ORGANS, 2001, 25 (12) :1004-1008
[10]   Influence of the charge of low molecular weight proteins on their efficacy of filtration and/or adsorption on dialysis membranes with different intrinsic properties [J].
Moachon, N ;