Effects of Acute Pump Speed Changes on Cerebral Hemodynamics in Patients With an Implantable Continuous-Flow Left Ventricular Assist Devices

被引:0
作者
Konarik, Miroslav [1 ,2 ]
Sramko, Marek [3 ,4 ]
Dorazilova, Zora [3 ]
Blaha, Martin [3 ]
Netuka, Ivan [1 ,5 ]
Ivak, Peter [1 ,6 ,7 ]
Maly, Jiri [1 ]
Szarszoi, Ondrej [1 ,8 ]
机构
[1] Inst Clin & Expt Med, Dept Cardiac Surg, Videnska 1958-9, Prague 14021, Czech Republic
[2] Charles Univ Prague, Fac Med 1, Dept Physiol, Prague, Czech Republic
[3] Inst Clin & Expt Med, Dept Cardiol, Prague, Czech Republic
[4] Charles Univ Prague, Fac Med 1, Dept Internal Med 2, Prague, Czech Republic
[5] Charles Univ Prague, Fac Med 1, Dept Surg 2, Dept Cardiovasc Surg, Prague, Czech Republic
[6] Charles Univ Prague, Fac Med 3, Dept Physiol, Prague, Czech Republic
[7] Charles Univ Prague, Fac Med 1, Dept Surg 2, Dept Cardiovasc Surg, Prague, Czech Republic
[8] Charles Univ Prague, Fac Med 2, Dept Pathophysiol, Prague, Czech Republic
关键词
LVAD; Continuous flow; Plethysmography; Cerebral autoregulation; AUTOREGULATION; PULSATILITY;
D O I
10.33549/physiolres.934738
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Mechanical circulatory support (MCS) with an implantable left ventricular assist device (LVAD) is an established therapeutic option for advanced heart failure. Most of the currently used LVADs generate a continuous stream of blood that decreases arterial pulse pressure. This study investigated whether a change of the pulse pressure during different pump speed settings would affect cerebral autoregulation and thereby affect cerebral blood flow (CBF). The study included 21 haemodynamically stable outpatients with a continuous-flow LVAD (HeartMate II, Abbott, USA) implanted a median of 6 months before the study (interquartile range 3 to 14 months). Arterial blood pressure (measured by finger plethysmography) was recorded simultaneously with CBF (measured by transcranial Doppler ultrasound) during baseline pump speed (8900 rpm [IQR 8800; 9200]) and during minimum and maximum tolerated pump speeds (8000 rpm [IQR 8000; 8200] and 9800 rpm [IQR 9800; 10 000]). An increase in LVAD pump speed by 800 rpm [IQR 800; 1000] from the baseline lead to a significant decrease in arterial pulse pressure and cerebral blood flow pulsatility (relative change -24 % and -32 %, both p < 0.01), but it did not affect mean arterial pressure and mean CBF velocity (relative change 1 % and -1.7 %, p=0.1 and 0.7). In stable patients with a continuous-flow LVAD, changes of pump speed settings within a clinically used range did not impair static cerebral autoregulation and cerebral blood flow.
引用
收藏
页码:831 / 839
页数:9
相关论文
共 50 条
  • [31] Surface mapping demonstrates compatibility of implantable loop monitor with a continuous-flow left ventricular assist device
    Li, Angela
    Atteya, Gourg
    Vullaganti, Sirish
    Mitra, Ramanak
    ESC HEART FAILURE, 2021, 8 (04): : 3392 - 3396
  • [32] Evaluation of anticoagulation and nonsurgical major bleeding in recipients of continuous-flow left ventricular assist devices
    Veasey, Tara M.
    Floroff, Catherine K.
    Strout, Sara E.
    McElray, Krista L.
    Brisco-Bacik, Meredith A.
    Cook, Jennifer L.
    Toole, John M.
    Craig, Michael L.
    Van Bakel, Adrian B.
    Meadows, Holly B.
    Uber, Walter E.
    ARTIFICIAL ORGANS, 2019, 43 (08) : 736 - 744
  • [33] Prognostic impact of implantable cardioverter defibrillators and associated adverse events in patients with continuous flow left ventricular assist devices
    Pausch, Jonas
    Mersmann, Julian
    Bhadra, Oliver D.
    Barten, Markus J.
    Toennis, Tobias
    Yildirim, Yalin
    Pecha, Simon
    Reichenspurner, Hermann
    Bernhardt, Alexander M.
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 10
  • [34] Effects of Continuous-Flow Left Ventricular Assist Device Therapy on Peripheral Vascular Function
    Kaya, Ersin
    Kocabas, Umut
    Simsek, Evrim
    Nalbantgil, Sanem
    Kahraman, Umit
    Engin, Cagatay
    Yagdi, Tahir
    Ozbaran, Mustafa
    Akilli, Azem
    ASAIO JOURNAL, 2022, 68 (02) : 214 - 219
  • [35] Noninvasive Blood Pressure Monitor Designed for Patients With Heart Failure Supported with Continuous-Flow Left Ventricular Assist Devices
    Sajgalik, Pavol
    Kremen, Vaclav
    Fabian, Vratislav
    Maltais, Simon
    Stulak, John M.
    Kushwaha, Sudhir S.
    Joyce, Lyle D.
    Schirger, John A.
    Johnson, Bruce D.
    ASAIO JOURNAL, 2019, 65 (02) : 127 - 133
  • [36] Pulsatile Pressure Delivery of Continuous-Flow Left Ventricular Assist Devices Is Markedly Reduced Relative to Heart Failure Patients
    Rosenbaum, Andrew N.
    Rossman, Timothy L.
    Reddy, Yogesh N.
    Villavicencio, Mauricio A.
    Stulak, John M.
    Spencer, Philip J.
    Kushwaha, Sudhir S.
    Behfar, Atta
    ASAIO JOURNAL, 2023, 69 (05) : 445 - 450
  • [37] The Predicament of Gastrointestinal Bleeding in Patients With a Continuous-Flow Left Ventricular Assist Device Pathophysiology, Evaluation, and Management
    Berg, Dana
    Lebovics, Edward
    Kai, Masashi
    Spielvogel, David
    CARDIOLOGY IN REVIEW, 2019, 27 (05) : 222 - 229
  • [38] Left Ventricular Decompression During Speed Optimization Ramps in Patients Supported by Continuous-Flow Left Ventricular Assist Devices: Device-Specific Performance Characteristics and Impact on Diagnostic Algorithms
    Uriel, Nir
    Levin, Allison P.
    Sayer, Gabriel T.
    Mody, Kanika P.
    Thomas, Sunu S.
    Adatya, Sirtaz
    Yuzefpolskaya, Melana
    Garan, Arthur R.
    Breskin, Alexander
    Takayama, Hiroo
    Colombo, Paulo C.
    Naka, Yoshifumi
    Burkhoff, Daniel
    Jorde, Ulrich P.
    JOURNAL OF CARDIAC FAILURE, 2015, 21 (10) : 785 - 791
  • [39] Prior hematologic conditions carry a high morbidity and mortality in patients supported with continuous-flow left ventricular assist devices
    Fried, Justin
    Levin, Allison P.
    Mody, Kanika M.
    Garan, Arthur R.
    Yuzefpolsakaya, Melana
    Takayama, Hiroo
    Diuguid, David L.
    Naka, Yoshifumi
    Jorde, Ulrich
    Uriel, Nir
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2014, 33 (11) : 1119 - 1125
  • [40] The value of echocardiographic examination in predicting right ventricular heart failure in patients after the implantation of continuous-flow left ventricular assist devices
    Paluszkiewicz, Lech
    Boergermann, Jochen
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2018, 27 (06) : 931 - 937