Tailoring CT Dose to Patient Size: Implementation of the Updated 2017 ACR Size-specific Diagnostic Reference Levels

被引:12
作者
Klosterkemper, Yan [1 ]
Appel, Elisabeth [1 ]
Thomas, Christoph [1 ]
Bethge, Oliver T. [1 ]
Aissa, Joel [1 ]
Kroepil, Patric [1 ]
Antoch, Gerald [1 ]
Boos, Johannes [1 ]
机构
[1] Univ Dusseldorf, Fac Med, Dept Diagnost & Intervent Radiol, Moorenstr 5, D-40225 Dusseldorf, Germany
关键词
Size specific dose estimates; SSDE; dose monitoring; radiation protection; TUBE CURRENT; VARIABILITY; SELECTION; INDEXES; IMPACT; ADULT; TABLE;
D O I
10.1016/j.acra.2018.03.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: To use an automatic computed tomography (CT) dose monitoring system to analyze the institutional chest and abdominopelvic CT dose data as regards the updated 2017 American College of Radiology (ACR) diagnostic reference levels (DRLs) based on water-equivalent diameter (Dw) and dose estimates (SSDE) to detect patient size subgroups in which CT dose can be optimized. Materials and Methods: All chest CT examinations performed between July 2016 and April 2017 with and without contrast material, CT of the pulmonary arteries, and abdominopelvic CT with and without contrast material were included in this retrospective study. Dw and SSDE were automatically calculated for all scans using a previously validated in-house developed Matlab software and stored into our CT dose monitoring system. CT dose data were analyzed as regards the updated ACR DRLs (size groups: 21-25 cm, 25-29 cm, 29-33 cm, 33-37 cm, 37-41 cm). SSDE and volumetric computed tomography dose index (CTDlvol) were used as CT dose parameter. Results: Overall, 30,002 CT examinations were performed in the study period, 3860 of which were included in the analysis (mean age 62.1 +/- 16.4 years, Dw 29.0 +/- 3.3 cm; n = 577 chest CT without contrast material, n = 628 chest CT with contrast material, n = 346 CT of chest pulmonary, n = 563 abdominopelvic CT without contrast material, n = 1746 abdominopelvic CT with contrast material). Mean SSDE and CTDlvol relative to the updated DRLs were 43.3 +/- 26.4 and 45.1 +/- 27.9% for noncontrast chest CT, 52.3 +/- 23.1 and 52.0 +/- 23.1% for contrast-enhanced chest CT, 68.8 +/- 29.5 and 70.0 +/- 31.0% for CT of pulmonary arteries, 41.9 +/- 29.2 and 43.3 +/- 31.3% for noncontrast abdominopelvic CT, and 56.8 +/- 22.2 and 58.8 +/- 24.4% for contrast-enhanced abdominopelvic CT. Lowest dose compared to the DRLs was found for the Dw group of 21-25 cm in noncontrast abdominopelvic CT (SSDE 30.4 +/- 21.8%, CTDlvol 30.8 +/- 21.4%). Solely the group of patients with a Dw of 37-41 cm undergoing noncontrast abdominopelvic CT exceeded the ACR DRL (SSDE 100.3 +/- 59.0%, CTDlvol 107.1 +/- 63.5%). Conclusions: On average, mean SSDE and CTDlvol of our institutional chest and abdominopelvic CT protocols were lower than the updated 2017 ACR DRLs. Size-specific subgroup analysis revealed a wide variability of SSDE and CTDlvol across CT protocols and patient size groups with a transgression of DRLs in noncontrast abdominopelvic CT of large patients (Dw 37-41 cm).
引用
收藏
页码:1624 / 1631
页数:8
相关论文
共 45 条
  • [21] Size-specific, scanner-independent organ dose estimates in contiguous axial and helical head CT examinations
    McMillan, Kyle
    Bostani, Maryam
    Cagnon, Chris
    Zankl, Maria
    Sepahdari, Ali R.
    McNitt-Gray, Michael
    MEDICAL PHYSICS, 2014, 41 (12)
  • [22] Size-specific dose estimates for pediatric head CT protocols based on the AAPM report TG-293
    Sapignoli, Sonia
    Roggio, Antonella
    Boschini, Alessio
    Guida, Federica
    Merlo, Carlo
    Paiusco, Marta
    Zorz, Alessandra
    De Monte, Francesca
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 100 : 26 - 30
  • [23] A SIMPLE METHOD FOR CALIBRATING PIXEL VALUES OF THE CT LOCALIZER RADIOGRAPH FOR CALCULATING WATER-EQUIVALENT DIAMETER AND SIZE-SPECIFIC DOSE ESTIMATE
    Anam, Choirul
    Fujibuchi, Toshioh
    Toyoda, Takatoshi
    Sato, Naoki
    Haryanto, Freddy
    Widita, Rena
    Arif, Idam
    Dougherty, Geoff
    RADIATION PROTECTION DOSIMETRY, 2018, 179 (02) : 158 - 168
  • [24] Accuracy of patient dose estimation in cone beam computed tomography in breast irradiation by size-specific dose estimates with position correction
    Ueno, Hiroyuki
    Matsubara, Kosuke
    Bou, Sayuri
    Hizume, Masato
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12):
  • [25] Local diagnostic reference level based on size-specific dose estimates: Assessment of pediatric abdominal/pelvic computed tomography at a Japanese national children’s hospital
    Rumi Imai
    Osamu Miyazaki
    Tetsuya Horiuchi
    Hideo Kurosawa
    Shunsuke Nosaka
    Pediatric Radiology, 2015, 45 : 345 - 353
  • [26] Breast dose reduction for chest CT by modifying the scanning parameters based on the pre-scan size-specific dose estimate (SSDE)
    Masafumi Kidoh
    Daisuke Utsunomiya
    Seitaro Oda
    Takeshi Nakaura
    Yoshinori Funama
    Hideaki Yuki
    Kenichiro Hirata
    Tomohiro Namimoto
    Daisuke Sakabe
    Masahiro Hatemura
    Yasuyuki Yamashita
    European Radiology, 2017, 27 : 2267 - 2274
  • [27] Volume computed tomography dose index (CTDIvol) and size-specific dose estimate (SSDE) for tube current modulation (TCM) in CT scanning
    Anam, C.
    Haryanto, F.
    Widita, R.
    Arif, I
    Dougherty, G.
    McLean, D.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2018, 16 (03): : 289 - 297
  • [28] Breast dose reduction for chest CT by modifying the scanning parameters based on the pre-scan size-specific dose estimate (SSDE)
    Kidoh, Masafumi
    Utsunomiya, Daisuke
    Oda, Seitaro
    Nakaura, Takeshi
    Funama, Yoshinori
    Yuki, Hideaki
    Hirata, Kenichiro
    Namimoto, Tomohiro
    Sakabe, Daisuke
    Hatemura, Masahiro
    Yamashita, Yasuyuki
    EUROPEAN RADIOLOGY, 2017, 27 (06) : 2267 - 2274
  • [29] Local diagnostic reference level based on size-specific dose estimates: Assessment of pediatric abdominal/pelvic computed tomography at a Japanese national children's hospital
    Imai, Rumi
    Miyazaki, Osamu
    Horiuchi, Tetsuya
    Kurosawa, Hideo
    Nosaka, Shunsuke
    PEDIATRIC RADIOLOGY, 2015, 45 (03) : 345 - 353
  • [30] Radiation doses and size-specific dose estimate from CT brain examinations according to head sizes in a tertiary hospital in Malaysia
    Kayun, Zunaide
    Karim, Muhammad Khalis Abdul
    Harun, Hanif Haspi
    Shaari, Abdul Halim
    Mahmud, Rozi
    Hamid, Hamzaini Abdul
    Sabarudin, Akmal
    Chew, Ming Tsuey
    RADIATION PHYSICS AND CHEMISTRY, 2021, 189