Tailoring CT Dose to Patient Size: Implementation of the Updated 2017 ACR Size-specific Diagnostic Reference Levels

被引:12
|
作者
Klosterkemper, Yan [1 ]
Appel, Elisabeth [1 ]
Thomas, Christoph [1 ]
Bethge, Oliver T. [1 ]
Aissa, Joel [1 ]
Kroepil, Patric [1 ]
Antoch, Gerald [1 ]
Boos, Johannes [1 ]
机构
[1] Univ Dusseldorf, Fac Med, Dept Diagnost & Intervent Radiol, Moorenstr 5, D-40225 Dusseldorf, Germany
关键词
Size specific dose estimates; SSDE; dose monitoring; radiation protection; TUBE CURRENT; VARIABILITY; SELECTION; INDEXES; IMPACT; ADULT; TABLE;
D O I
10.1016/j.acra.2018.03.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: To use an automatic computed tomography (CT) dose monitoring system to analyze the institutional chest and abdominopelvic CT dose data as regards the updated 2017 American College of Radiology (ACR) diagnostic reference levels (DRLs) based on water-equivalent diameter (Dw) and dose estimates (SSDE) to detect patient size subgroups in which CT dose can be optimized. Materials and Methods: All chest CT examinations performed between July 2016 and April 2017 with and without contrast material, CT of the pulmonary arteries, and abdominopelvic CT with and without contrast material were included in this retrospective study. Dw and SSDE were automatically calculated for all scans using a previously validated in-house developed Matlab software and stored into our CT dose monitoring system. CT dose data were analyzed as regards the updated ACR DRLs (size groups: 21-25 cm, 25-29 cm, 29-33 cm, 33-37 cm, 37-41 cm). SSDE and volumetric computed tomography dose index (CTDlvol) were used as CT dose parameter. Results: Overall, 30,002 CT examinations were performed in the study period, 3860 of which were included in the analysis (mean age 62.1 +/- 16.4 years, Dw 29.0 +/- 3.3 cm; n = 577 chest CT without contrast material, n = 628 chest CT with contrast material, n = 346 CT of chest pulmonary, n = 563 abdominopelvic CT without contrast material, n = 1746 abdominopelvic CT with contrast material). Mean SSDE and CTDlvol relative to the updated DRLs were 43.3 +/- 26.4 and 45.1 +/- 27.9% for noncontrast chest CT, 52.3 +/- 23.1 and 52.0 +/- 23.1% for contrast-enhanced chest CT, 68.8 +/- 29.5 and 70.0 +/- 31.0% for CT of pulmonary arteries, 41.9 +/- 29.2 and 43.3 +/- 31.3% for noncontrast abdominopelvic CT, and 56.8 +/- 22.2 and 58.8 +/- 24.4% for contrast-enhanced abdominopelvic CT. Lowest dose compared to the DRLs was found for the Dw group of 21-25 cm in noncontrast abdominopelvic CT (SSDE 30.4 +/- 21.8%, CTDlvol 30.8 +/- 21.4%). Solely the group of patients with a Dw of 37-41 cm undergoing noncontrast abdominopelvic CT exceeded the ACR DRL (SSDE 100.3 +/- 59.0%, CTDlvol 107.1 +/- 63.5%). Conclusions: On average, mean SSDE and CTDlvol of our institutional chest and abdominopelvic CT protocols were lower than the updated 2017 ACR DRLs. Size-specific subgroup analysis revealed a wide variability of SSDE and CTDlvol across CT protocols and patient size groups with a transgression of DRLs in noncontrast abdominopelvic CT of large patients (Dw 37-41 cm).
引用
收藏
页码:1624 / 1631
页数:8
相关论文
共 45 条
  • [1] Implementation of Institutional Size-Specific Diagnostic Reference Levels for CT Angiography
    Klosterkemper, Yan
    Thomas, Christoph
    Bethge, Oliver T.
    Appel, Elisabeth
    Aissa, Joel
    Boeven, Judith
    Antoch, Gerald
    Boos, Johannes
    ACADEMIC RADIOLOGY, 2019, 26 (12) : 1661 - 1667
  • [2] Estimation of radiation dose associated with bone SPECT/CT and establishing local diagnostic reference levels using size-specific dose estimate
    Alrehily, Faisal
    Alshamrani, Abdullah F.
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2023, 16 (01)
  • [3] Establishing local diagnostic reference levels for computed tomography examinations using size-specific dose estimates
    Alrehily, Faisal A.
    Alqahtani, Khalid S.
    Aljohani, Marwan H.
    Alharbi, Nawaf S.
    Alharbi, Reyan M.
    Abdulaal, Osamah M.
    Alshamrani, Abdullah F.
    Alsaedi, Abdulrahman S.
    Al-Murshedi, Sadeq H.
    Alhazmi, Fahad H.
    SAUDI MEDICAL JOURNAL, 2023, 44 (08) : 761 - 766
  • [4] Size-Specific Dose Estimates for Evaluation of Individual Patient Dose in CT Protocol for Renal Colic
    Waszczuk, Lukasz A.
    Guzinski, Maciej
    Czarnecka, Anna
    Sasiadek, Marek J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 205 (01) : 100 - 105
  • [5] Comparison of central, peripheral, and weighted size-specific dose in CT
    Choirul, Anam
    Adhianto, Dwi
    Sutanto, Heri
    Adi, Kusworo
    Ali, Mohd Hanafi
    Rae, William Ian Duncombe
    Fujibuchi, Toshioh
    Dougherty, Geoff
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2020, 28 (04) : 695 - 708
  • [6] Institutional computed tomography diagnostic reference levels based on water-equivalent diameter and size-specific dose estimates
    Boos, Johannes
    Thomas, Christoph
    Appel, Elisabeth
    Klosterkemper, Yan
    Schleich, Christoph
    Aissa, Joel
    Bethge, Oliver Th
    Antoch, Gerald
    Kroepil, Patric
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (02) : 536 - 548
  • [7] Local diagnostic reference level based on size-specific dose estimates: case of chest-abdomen-pelvis CT examination in a Moroccan hospital
    Ouaggou, Ibrahim Ait
    Bakari, Dikra
    Zerfaoui, Mustapha
    Rrhioua, Abdeslem
    Oulhouq, Yassine
    Benmessaoud, Mounir
    Ouhadda, Hassan
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2024,
  • [8] Patient size as a parameter for determining Diagnostic Reference Levels for paediatric Computed Tomography (CT) procedures
    Satharasinghe, Duminda
    Jeyasugiththan, Jeyasingam
    Wanninayake, W. M. N. M. B.
    Pallewatte, A. S.
    Samarasinghe, R. A. N. K. K.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 102 : 55 - 65
  • [9] How accurate is size-specific dose estimate in pediatric body CT examinations?
    Karmazyn, Boaz
    Ai, Huisi
    Klahr, Paul
    Ouyang, Fangqian
    Jennings, S. Gregory
    PEDIATRIC RADIOLOGY, 2016, 46 (09) : 1234 - 1240
  • [10] Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size
    Kawashima, Hiroki
    Ichikawa, Katsuhiro
    Hanaoka, Shinsuke
    Matsubara, Kosuke
    Takata, Tadanori
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04): : 246 - 251