Reproducibility of ultrasonic fetal volume blood flow measurements

被引:11
作者
Brodzki, J
Gardiner, HM
Eriksson, A
Stale, H
Marsál, K
机构
[1] Univ London Imperial Coll Sci Technol & Med, Sch Med, Royal Brompton Hosp & Natl Heart & Lung Inst, Dept Paediat Cardiol, London, England
[2] Univ Lund, Lund Inst Technol, Dept Elect Measurements, Lund, Sweden
[3] Univ Lund Hosp, Dept Obstet & Gynecol, Malmo, Sweden
[4] Univ Lund, Malmo Univ Hosp, Dept Obstet & Gynecol, Malmo, Sweden
来源
CLINICAL PHYSIOLOGY | 1998年 / 18卷 / 05期
关键词
Doppler ultrasound; fetal blood flow; intraobserver agreement; ultrasound; velocity; vessel diameter;
D O I
10.1046/j.1365-2281.1998.00127.x
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
The intraobserver reproducibility of ultrasonic volume blood flow measurements in the human fetus was evaluated in this study. A new approach, simultaneous measurement of the vessel diameter and the flow velocity with a pulsed-wave Doppler ultrasound synchronized with a real-time ultrasound phase-locked echo-tracking system, was used to estimate volume blood flow (VBF) in the fetal descending aorta. Measurements were performed in a longitudinal study on 20 normally grown fetuses. Intraobserver reproducibility of repeated estimations of mean blood flow velocities throughout gestation was very good, with high values of intraclass correlation coefficient (IntraCC 0.80-0.91) and low values of coefficient of variation (CV 4-11%). The IntraCC of repeated vessel diameter measurements throughout gestation was low (0.30-0.68), whereas the values of CV were acceptable (< 12%), with the exception of the period between 140 and 167 gestational days (CV > 12%). The lower reproducibility of vessel diameter measurement contributed directly to the relatively low reproducibility of VBF estimations overall (IntraCC 0.25-0.70; CV 17-28%), as these are calculated from a formula using both flow velocity and vessel diameter. Nevertheless, the synchronized approach gives absolute values of vessel diameter, flow velocity and VBF comparable with values reported in the human fetus previously. The new method provides, by taking the vessel wall pulsations into consideration and by measuring diameter and velocity simultaneously, a more complete information on fetal haemodynamics and fetal physiology.
引用
收藏
页码:479 / 485
页数:7
相关论文
共 17 条
[1]   A NOTE ON THE USE OF THE INTRACLASS CORRELATION-COEFFICIENT IN THE EVALUATION OF AGREEMENT BETWEEN 2 METHODS OF MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
COMPUTERS IN BIOLOGY AND MEDICINE, 1990, 20 (05) :337-340
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   A NEW VIEW OF INTER-OBSERVER AGREEMENT [J].
BURDOCK, EI ;
FLEISS, JL ;
HARDESTY, AS .
PERSONNEL PSYCHOLOGY, 1963, 16 (04) :373-384
[4]  
CAMPBELL S, 1995, COLOR ATLAS DOPPLER, P128
[5]   ULTRASONIC MEASUREMENT OF HUMAN-FETAL BLOOD-FLOW [J].
EIKNES, SH ;
MARSAL, K ;
BRUBAKK, AO ;
KRISTOFFERSON, K ;
ULSTEIN, M .
JOURNAL OF BIOMEDICAL ENGINEERING, 1982, 4 (01) :28-36
[6]  
EIKNES SH, 1984, ULTRASOUND MED BIOL, V10, P329, DOI 10.1016/0301-5629(84)90167-4
[7]  
ENGLAND JM, 1975, MED RES STAT EPIDEIM, P4
[8]  
GILL R W, 1979, Ultrasound in Medicine and Biology, V5, P223, DOI 10.1016/0301-5629(79)90014-0
[9]   NONINVASIVE MEASUREMENT OF PULSATILE VESSEL DIAMETER CHANGE AND ELASTIC PROPERTIES IN HUMAN ARTERIES - A METHODOLOGICAL STUDY [J].
HANSEN, F ;
BERGQVIST, D ;
MANGELL, P ;
RYDEN, A ;
SONESSON, B ;
LANNE, T .
CLINICAL PHYSIOLOGY, 1993, 13 (06) :631-643
[10]  
LINDSTROM K, 1987, FETAL PHYSL MEASUREM, P217