Comparison of IOL master and ultrasound biomicroscopy in anterior chamber depth measurement

被引:0
作者
Zhou, Wen-Kai [1 ]
Li, Xiao-Yan [1 ]
Zhang, Jin-Song [1 ]
机构
[1] China Med Univ, Affiliated Hosp 4, Dept Ophthalmol, Liaoning Prov Coll,Key Lab Lens,Eye Hosp, Shenyang 110005, Liaoning Prov, Peoples R China
来源
INTERNATIONAL JOURNAL OF OPHTHALMOLOGY | 2009年 / 2卷 / 04期
关键词
anterior chamber depth; intraocular lens master; ultrasound biomicroscopy; BIOMETRY;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
AIM: To compare the measurement of anterior chamber depth (ACD) inclusive of corneal thickness using intraocular lens (IOL) master and ultrasound biomicroscopy (UBM) and evaluate the repeatability of each method. METHODS: Two consecutive measurements of ACD were prospectively performed using IOL master and UBM in 60 eyes in 60 individuals. Mean values were compared using the paired t test. For each individual, ACD measurements was performed 5 times to estimate the repeatability of each method by a coefficient of variation (CV). RESULTS: The mean ACD was 2.95? .25mm with the IOL master and 2.96 +/- 0.22mm with the UBM. This difference was not statistically significant (P =0.631). The coefficient of variation (CV) was 0.56%+/- 0.26% and 0.65%+/- 0.36% in IOL master and UBM, respectively. CONCLUSION: The mean ACD of IOL master is the same as UBM. The repeatability of IOL master is better than UBM.
引用
收藏
页码:352 / 354
页数:3
相关论文
共 7 条
[1]   Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment [J].
Dada, Tanuj ;
Sihota, Ramanjit ;
Gadia, Ritu ;
Aggarwal, Anand ;
Mandal, Subrata ;
Gupta, Viney .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (05) :837-840
[2]   Comparison of different techniques of anterior chamber depth and keratometric measurements [J].
Elbaz, Uri ;
Barkana, Yaniv ;
Gerber, Yariv ;
Avni, Isaac ;
Zadok, David .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (01) :48-53
[3]   Influence of operator experience on the performance of ultrasound biometry compared to optical biometry before cataract surgery [J].
Findl, O ;
Kriechbaum, K ;
Sacu, S ;
Kiss, B ;
Polak, K ;
Nepp, J ;
Schild, G ;
Rainer, G ;
Maca, S ;
Petternel, V ;
Lackner, B ;
Drexler, W .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (10) :1950-1955
[4]   Anterior chamber depth in Mongolians: Variation with age, sex, and method of measurement [J].
Foster, PJ ;
Alsbirk, PH ;
Baasanhu, J ;
Munkhbayar, D ;
Uranchimeg, D ;
Johnson, GJ .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1997, 124 (01) :53-60
[5]   Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations [J].
Holladay, JT .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1997, 23 (09) :1356-1370
[6]   Comparison of ultrasound pachymetry and partial coherence interferometry in the measurement of central corneal thickness [J].
Rainer, G ;
Petternel, V ;
Findl, O ;
Schmetterer, L ;
Skorpik, C ;
Luksch, A ;
Drexler, W .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (12) :2142-2145
[7]   A new non-contact optical device for ocular biometry [J].
Santodomingo-Rubido, J ;
Mallen, EAH ;
Gilmartin, B ;
Wolffsohn, JS .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2002, 86 (04) :458-462