Wear Comparison Between a Highly Cross-Linked Polyethylene and Conventional Polyethylene Against a Zirconia Femoral Head Minimum 5-Year Follow-Up

被引:17
|
作者
Fukui, Kiyokazu [1 ]
Kaneuji, Ayumi [1 ]
Sugimori, Tanzo [1 ]
Ichiseki, Toru [1 ]
Kitamura, Kenji [1 ]
Matsumoto, Tadami [1 ]
机构
[1] Kanazawa Med Univ, Dept Orthoped Surg, Kanazawa, Ishikawa, Japan
关键词
cross-linked polyethylene; polyethylene wear; total hip arthroplasty; zirconia femoral head; TOTAL HIP-ARTHROPLASTY; CERAMIC HEAD; RADIOSTEREOMETRIC ANALYSIS; UNCEMENTED SOCKETS; ALUMINA; FRACTURE; JOINTS; METAL; REPLACEMENT; SIMULATOR;
D O I
10.1016/j.arth.2009.11.005
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Highly cross-linked polyethylene (HXLPE) was developed to reduce wear of articular bearing surface in total hip arthroplasty patients. Several studies have shown reduced wear of HXLPE compared with conventional polyethylene; however, these studies had used HXLPE in combination with a Co-Cr metal head. The purpose of this study was to compare the 5-year in vivo wear of HXLPE with that of conventional PE using a zirconia femoral head. Forty-five hips with a Trilogy HXLPE (Zimmer, Warsaw, Ind) were matched and compared with a control group of 20 conventional Trilogy PE hips. The 2-dimensional linear wear rate was significantly less in the HXLPE group between 1 and 5 years postoperation (P < .001). The results show that HXLPE reduces short-term polyethylene wear against not only a Co-Cr head but also a zirconia head.
引用
收藏
页码:45 / 49
页数:5
相关论文
共 50 条
  • [1] Wear Comparison Between Conventional and Highly Cross-Linked Polyethylene Against a Zirconia Head A Concise Follow-Up, at an Average 10 Years, of a Previous Report
    Fukui, Kiyokazu
    Kaneuji, Ayumi
    Sugimori, Tanzo
    Ichiseki, Toru
    Matsumoto, Tadami
    JOURNAL OF ARTHROPLASTY, 2013, 28 (09) : 1654 - 1658
  • [2] Highly Cross-Linked Polyethylene in Total Hip Arthroplasty for Osteonecrosis of the Femoral Head: A Minimum 5-Year Follow-Up Study
    Min, Byung-Woo
    Lee, Kyung-Jae
    Song, Kwang-Soon
    Bae, Ki-Cheor
    Cho, Chul-Hyun
    JOURNAL OF ARTHROPLASTY, 2013, 28 (03) : 526 - 530
  • [3] Femoral Head Size and Wear of Highly Cross-linked Polyethylene at 5 to 8 Years
    Lachiewicz, Paul F.
    Heckman, Daniel S.
    Soileau, Elizabeth S.
    Mangla, Jimmi
    Martell, John M.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (12) : 3290 - 3296
  • [4] Minimum Five-Year Follow-Up Wear Measurement of Longevity Highly Cross-Linked Polyethylene Cup Against Cobalt-Chromium or Zirconia Heads
    Nakahara, Ichiro
    Nakamura, Nobuo
    Nishii, Takashi
    Miki, Hidenobu
    Sakai, Takashi
    Sugano, Nobuhiko
    JOURNAL OF ARTHROPLASTY, 2010, 25 (08) : 1182 - 1187
  • [5] Fixation and Wear With a Contemporary Acetabular Component and Cross-Linked Polyethylene at Minimum 10-Year Follow-Up
    Bedard, Nicholas A.
    Callaghan, John J.
    Stefl, Michael D.
    Willman, Tyler J.
    Liu, Steve S.
    Goetz, Devon D.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (10) : 1961 - 1969
  • [6] Highly Cross-Linked Polyethylene in Cementless Total Hip Arthroplasty in Patients with Previous Acetabular Fractures: A Minimum 5-Year Follow-Up Study
    Min, Byung-Woo
    Bae, Ki-Cheor
    Cho, Chul-Hyun
    Son, Eun-Suck
    Lee, Kyung-Jae
    Lee, Si-Wook
    INDIAN JOURNAL OF ORTHOPAEDICS, 2020, 54 (SUPPL 2) : 239 - 245
  • [7] Highly Cross-Linked Versus Conventional Polyethylene in Posterior-Stabilized Total Knee Arthroplasty at a Mean 5-Year Follow-up
    Meneghini, R. Michael
    Lovro, Luke R.
    Smits, Shelly A.
    Ireland, Philip H.
    JOURNAL OF ARTHROPLASTY, 2015, 30 (10) : 1736 - 1739
  • [8] Highly Cross-Linked Polyethylene in Cementless Total Hip Arthroplasty in Patients with Previous Acetabular Fractures: A Minimum 5-Year Follow-Up Study
    Byung-Woo Min
    Ki-Cheor Bae
    Chul-Hyun Cho
    Eun-Suck Son
    Kyung-Jae Lee
    Si-Wook Lee
    Indian Journal of Orthopaedics, 2020, 54 : 239 - 245
  • [9] No Revisions Attributable to Wear of Highly Cross-Linked Polyethylene Liners: A Long-Term Follow-Up Study
    Wright, Breydan H.
    Hadley, Matthew L.
    Harmer, Joshua R.
    Fruth, Kristin M.
    Sierra, Rafael J.
    Couch, Cory G.
    JOURNAL OF ARTHROPLASTY, 2024, 39 (09) : S347 - S352
  • [10] In vivo wear performance of highly cross-linked polyethylene vs. yttria stabilized zirconia and alumina stabilized zirconia at a mean seven-year follow-up
    Hasegawa, Masahiro
    Sudo, Akihiro