Cascaded systems analysis of anatomic noise in digital mammography and dual-energy digital mammography

被引:6
作者
Tanguay, Jesse [1 ]
Lalonde, Robert [1 ]
Bjarnason, Thorarin A. [2 ,3 ,4 ]
Yang, Chang-Ying Joseph [2 ,5 ]
机构
[1] Ryerson Univ, Dept Phys, Toronto, ON, Canada
[2] Univ British Columbia Okanagan, Dept Phys, Kelowna, BC, Canada
[3] Interior Hlth, Med Imaging, Kelowna, BC, Canada
[4] Univ British Columbia, Dept Radiol, Vancouver, BC, Canada
[5] BC Canc Kelowna, Kelowna, BC, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
mammography; breast imaging; image quality; noise power spectrum; medical imaging; dual-energy imaging; DETECTIVE QUANTUM EFFICIENCY; OPTIMIZATION; TOMOSYNTHESIS; SIGNAL; MODEL; DETECTABILITY; PROPAGATION; VALIDATION; PARAMETERS; SPECTRA;
D O I
10.1088/1361-6560/ab3fcd
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In x-ray based imaging of the breast, contrast between fibroglandular (Fg) tissue and adipose (Ad) tissue is a source of anatomic noise. The goal of this work was to validate by simulation and experiment a mathematical framework for modelling the Fg component of anatomic noise in digital mammograpy (DM) and dual-energy (DE) DM. Our mathematical framework unifies and generalizes existing approaches. We compared mathematical predictions directly with empirical measurements of the anatomic noise power spectrum of the CIRS BR3D structured breast phantom using two clinical mammography systems and four beam qualities. Our simulation and experimental results showed agreement with mathematical predictions. As a demonstration of utility, we used our mathematical framework in a theoretical spectral optimization of DM for the task of detecting breast masses. Our theoretical optimization showed that the optimal tube voltage for DM may be higher than that based on predictions that do not account for anatomic noise, in agreement with recent theoretical findings. Additionally, our theoretical optimization predicts that filtering tungsten-anode x-ray spectra with rhodium has little influence on lesion detectability, in contrast with previous findings. The mathematical methods validated in this work can be incorporated easily into cascaded systems analysis of breast imaging systems and will be useful when optimizating novel techniques for x-ray-based imaging of the breast.
引用
收藏
页数:21
相关论文
共 51 条
[1]  
Abbey C. K., 2000, Handbook of Medical Imaging, V1, P629
[2]   Including the effect of motion artifacts in noise and performance analysis of dual-energy contrast-enhanced mammography [J].
Allec, N. ;
Abbaszadeh, S. ;
Scott, C. C. ;
Lewin, J. M. ;
Karim, K. S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (24) :8405-8425
[3]   Evaluating noise reduction techniques while considering anatomical noise in dual-energy contrast-enhanced mammography [J].
Allec, Nicholas ;
Abbaszadeh, Shiva ;
Scott, Chris C. ;
Karim, Karim S. ;
Lewin, John M. .
MEDICAL PHYSICS, 2013, 40 (05)
[4]  
[Anonymous], Probability, Random Variables and Stochastic Processes
[5]   X-ray spectrum optimization of full-field digital mammography: Simulation and phantom study [J].
Bernhardt, Philipp ;
Mertelmeier, Thomas ;
Hoheisel, Martin .
MEDICAL PHYSICS, 2006, 33 (11) :4337-4349
[6]   Estimation of the noisy component of anatomical backgrounds [J].
Bochud, FO ;
Valley, JF ;
Verdun, FR ;
Hessler, C ;
Schnyder, P .
MEDICAL PHYSICS, 1999, 26 (07) :1365-1370
[7]   The importance of anatomical noise in mammography [J].
Bochud, FO ;
Verdun, FR ;
Valley, JF ;
Hessler, C ;
Moeckli, R .
IMAGE PERCEPTION: MEDICAL IMAGING 1997, 1997, 3036 :74-80
[8]   Glandular breast dose for monoenergetic and high-energy X-ray beams: Monte Carlo assessment [J].
Boone, JM .
RADIOLOGY, 1999, 213 (01) :23-37
[9]   Scatter/primary in mammography: Comprehensive results [J].
Boone, JM ;
Lindfors, KK ;
Cooper, VN ;
Seibert, JA .
MEDICAL PHYSICS, 2000, 27 (10) :2408-2416
[10]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874