Cascaded systems analysis of anatomic noise in digital mammography and dual-energy digital mammography

被引:6
|
作者
Tanguay, Jesse [1 ]
Lalonde, Robert [1 ]
Bjarnason, Thorarin A. [2 ,3 ,4 ]
Yang, Chang-Ying Joseph [2 ,5 ]
机构
[1] Ryerson Univ, Dept Phys, Toronto, ON, Canada
[2] Univ British Columbia Okanagan, Dept Phys, Kelowna, BC, Canada
[3] Interior Hlth, Med Imaging, Kelowna, BC, Canada
[4] Univ British Columbia, Dept Radiol, Vancouver, BC, Canada
[5] BC Canc Kelowna, Kelowna, BC, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
mammography; breast imaging; image quality; noise power spectrum; medical imaging; dual-energy imaging; DETECTIVE QUANTUM EFFICIENCY; OPTIMIZATION; TOMOSYNTHESIS; SIGNAL; MODEL; DETECTABILITY; PROPAGATION; VALIDATION; PARAMETERS; SPECTRA;
D O I
10.1088/1361-6560/ab3fcd
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In x-ray based imaging of the breast, contrast between fibroglandular (Fg) tissue and adipose (Ad) tissue is a source of anatomic noise. The goal of this work was to validate by simulation and experiment a mathematical framework for modelling the Fg component of anatomic noise in digital mammograpy (DM) and dual-energy (DE) DM. Our mathematical framework unifies and generalizes existing approaches. We compared mathematical predictions directly with empirical measurements of the anatomic noise power spectrum of the CIRS BR3D structured breast phantom using two clinical mammography systems and four beam qualities. Our simulation and experimental results showed agreement with mathematical predictions. As a demonstration of utility, we used our mathematical framework in a theoretical spectral optimization of DM for the task of detecting breast masses. Our theoretical optimization showed that the optimal tube voltage for DM may be higher than that based on predictions that do not account for anatomic noise, in agreement with recent theoretical findings. Additionally, our theoretical optimization predicts that filtering tungsten-anode x-ray spectra with rhodium has little influence on lesion detectability, in contrast with previous findings. The mathematical methods validated in this work can be incorporated easily into cascaded systems analysis of breast imaging systems and will be useful when optimizating novel techniques for x-ray-based imaging of the breast.
引用
收藏
页数:21
相关论文
共 50 条
  • [1] A linear systems approach to modeling anatomic noise in digital mammography
    Lalonde, Robert
    Tanguay, Jesse
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [2] Cascaded systems analysis of noise and detectability in dual-energy cone-beam CT
    Gang, Grace J.
    Zbijewski, Wojciech
    Stayman, J. Webster
    Siewerdsen, Jeffrey H.
    MEDICAL PHYSICS, 2012, 39 (08) : 5145 - 5156
  • [3] Algorithmic scatter correction in dual-energy digital mammography
    Chen, Xi
    Nishikawa, Robert M.
    Chan, Suk-tak
    Lau, Beverly A.
    Zhang, Lei
    Mou, Xuanqin
    MEDICAL PHYSICS, 2013, 40 (11)
  • [4] Contrast-enhanced dual-energy digital subtraction mammography: Optimization of the beam energy
    Kwan, ALC
    Boone, JM
    Le-Petross, H
    Lindfors, KK
    Seibert, JA
    Lewin, JM
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 1317 - 1321
  • [5] Anatomical noise in contrast-enhanced digital mammography. Part II. Dual-energy imaging
    Hill, Melissa L.
    Mainprize, James G.
    Carton, Ann-Katherine
    Saab-Puong, Sylvie
    Iordache, Razvan
    Muller, Serge
    Jong, Roberta A.
    Dromain, Clarisse
    Yaffe, Martin J.
    MEDICAL PHYSICS, 2013, 40 (08)
  • [6] Improved microcalcification visualization using dual-energy digital mammography
    Tsai, Chia-Jung
    Chen, Ran-Chou
    Peng, Hui-Ling
    Hsu, Wen-Lin
    Lee, Jason J. S.
    ACTA RADIOLOGICA, 2013, 54 (06) : 614 - 621
  • [7] Dual-energy digital mammography for calcification imaging: theory and implementation
    Kappadath, SC
    Shaw, C
    Lai, CJ
    Liu, XM
    Whitman, G
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 751 - 760
  • [8] Dual-energy contrast-enhanced digital mammography: Examination protocol
    Barra, F. R.
    Ribeiro, A. C.
    Mathieu, O. D.
    Rodrigues, A. C.
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2014, 95 (03) : 351 - 352
  • [9] Dual-energy digital mammography for calcification imaging: Scatter and nonuniformity corrections
    Kappadath, SC
    Shaw, CC
    MEDICAL PHYSICS, 2005, 32 (11) : 3395 - 3408
  • [10] DUAL-ENERGY MAMMOGRAPHY - A DETECTOR ANALYSIS
    BOONE, JM
    SHABER, GS
    TECOTZKY, M
    MEDICAL PHYSICS, 1990, 17 (04) : 665 - 675