Confirming nasogastric tube placement: Is the colorimeter as sensitive and specific as X-ray? A diagnostic accuracy study

被引:15
|
作者
Mordiffi, Siti Zubaidah [1 ]
Goh, Mien Li [1 ]
Phua, Jason [2 ]
Chan, Yiong-Huak [3 ]
机构
[1] Natl Univ Singapore Hosp, Dept Nursing, Evidence Based Nursing Unit, 5 Lower Kent Ridge Rd,Main Bldg 1,Level 6, Singapore 119074, Singapore
[2] Natl Univ Hlth Syst, Univ Med Cluster, Div Resp & Crit Care Med, Singapore, Singapore
[3] Natl Univ Hlth Syst, Yong Loo Lin Sch Med, Biostat Unit, Singapore, Singapore
关键词
Colorimeter; Enteral nutrition; General ward setting; Hospital; Inpatient; Nasogastric tube; Nursing; Sensitivity; Specificity; Radiographic imaging; X-ray; COMPLICATIONS; POSITION;
D O I
10.1016/j.ijnurstu.2016.06.011
中图分类号
R47 [护理学];
学科分类号
1011 ;
摘要
Background: The effect of delivering enteral nutrition or medications via a nasogastric tube that is inadvertently located in the tracheobronchial tract can cause respiratory complications. Although radiographic examination is accepted as the gold standard for confirming the position of patients' enteral tubes, it is costly, involves risks of radiation, and is not failsafe. Studies using carbon dioxide sensors to detect inadvertent nasogastric tube placements have been conducted in intensive care settings. However, none involved patients in general wards. Objective: The objective of this study was to ascertain the diagnostic measure of colorimeter, with radiographic examination as the reference standard, to confirm the location of nasogastric tubes in patients. Design: A prospective observational study of a diagnostic test. Setting: This study was conducted in the general wards of an approximately 1100-bed acute care tertiary hospital of an Academic Medical Center in Singapore. Participants: Adult patients with nasogastric tubes admitted to the general wards were recruited into the study. Methods: The colorimeter was attached to the nasogastric tube to detect for the presence of carbon dioxide, suggestive of a tracheobronchial placement. The exact location of the nasogastric tube was subsequently confirmed by a radiographic examination. Results: A total of 192 tests were undertaken. The colorimeter detected carbon dioxide in 29 tested nasogastric tubes, of which radiographic examination confirmed that four tubes were located in the tracheobronchial tract. The colorimeter failed to detect carbon dioxide in one nasogastric tube that was located in the tracheobronchial tract, thus, demonstrating a sensitivity of 0.80 [95% CI (0.376, 0.964)]. The colorimeter detected absence of carbon dioxide in 163 tested nasogastric tubes in which radiographic examination confirmed 160 gastrointestinal and one tracheobronchial placements, demonstrating a specificity of 0.865 [95% CI (0.808, 0.907)]. The colorimeter detected one tracheobronchial nasogastric tube placement that the radiographic examination was misinterpreted. Conclusion: The study found that the use of the colorimeter in the general ward setting was not 100% sensitive or specific in ascertaining the location of a nasogastric tube as previously reported by many studies undertaken in intensive care settings. This is the first study on the use of a colorimeter to confirm the placement of a nasogastric tube in adult patients in the general ward setting. More research on the use of a colorimeter in the general ward setting and its potential use in certain processes for confirming the placement of a nasogastric tube is warranted. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:248 / 257
页数:10
相关论文
共 31 条
  • [21] Diagnostic accuracy of computer aided reading of chest x-ray in screening for pulmonary tuberculosis in comparison with Gene-Xpert
    Nishtar, Tahira
    Burki, Shamsullah
    Ahmad, Fatima Sultan
    Ahmad, Tabish
    PAKISTAN JOURNAL OF MEDICAL SCIENCES, 2022, 38 (01) : 62 - 68
  • [22] Artificial intelligence for X-ray scaphoid fracture detection: a systematic review and diagnostic test accuracy meta-analysis
    Kraus, Matan
    Anteby, Roi
    Konen, Eli
    Eshed, Iris
    Klang, Eyal
    EUROPEAN RADIOLOGY, 2024, 34 (07) : 4341 - 4351
  • [23] Determination of molybdenum target parameters for transmission X-ray tube: A Geant4 simulation study
    Mohammad Mehdi Nasseri
    Pramana, 2019, 92
  • [24] Multivariate analysis of factors associated with the successful prediction of initial blind placement of a nasointestinal tube in the stomach based on X-ray imaging: a retrospective, single-center study
    Zheng, Zihao
    Wang, Jiangbo
    Shao, Ziqiang
    Cai, Hanhui
    Lu, Lihai
    Tang, Siyu
    Xu, Shuting
    Gong, Fangxiao
    Ye, Sen
    Yang, Xianghong
    Liu, Jingquan
    BMC GASTROENTEROLOGY, 2024, 24 (01)
  • [25] Determination of molybdenum target parameters for transmission X-ray tube: A Geant4 simulation study
    Nasseri, Mohammad Mehdi
    PRAMANA-JOURNAL OF PHYSICS, 2019, 92 (04):
  • [26] X-Ray and CT Scan Based Prediction of Best Fit Tracheostomy Tube-A Pilot Study
    Corbett, Mel
    Hughes, Isobel
    O'Shea, John
    Davey, Matthew G.
    Savage, Jane
    Hughes, Joseph
    Wallis, Fintan
    DIAGNOSTICS, 2020, 10 (08)
  • [27] Determination of Tungsten Target Parameters for Transmission X-ray Tube: A Simulation Study Using Geant4
    Nasseri, Mohammad M.
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2016, 48 (03) : 795 - 798
  • [28] The diagnostic yield and clinical impact of a chest X-ray after percutaneous dilatational tracheostomy: a prospective cohort study
    Haddad, S. H.
    Aldawood, A. S.
    Arabi, Y. M.
    ANAESTHESIA AND INTENSIVE CARE, 2007, 35 (03) : 393 - 397
  • [29] Specific surface, crystallinity, and dissolution of lyophilized fibrinogen. A study by combined small- and wide-angle X-ray scattering (SWAXS)
    Wahl, Verena
    Saurugger, Eva
    Khinast, Johannes
    Laggner, Peter
    EUROPEAN JOURNAL OF PHARMACEUTICS AND BIOPHARMACEUTICS, 2015, 89 : 374 - 382
  • [30] Reappraisal of Pediatric Diastatic Skull Fractures in the 3-Dimensional CT Era: Clinical Characteristics and Comparison of Diagnostic Accuracy of Simple Skull X-Ray, 2-Dimensional CT, and 3-Dimensional CT
    Sim, Sook Young
    Kim, Hyun Gi
    Yoon, Soo Han
    Choi, Jong Wook
    Cho, Sung Min
    Choi, Mi Sun
    WORLD NEUROSURGERY, 2017, 108 : 399 - 406