Effects of dose loading conditions and device geometry on the transport and aerosolization in dry powder inhalers: A simulation study

被引:7
|
作者
Sulaiman, Mostafa [1 ]
Liu, Xiaoyu [1 ]
Sundaresan, Sankaran [1 ]
机构
[1] Princeton Univ, Dept Chem & Biol Engn, Princeton, NJ 08544 USA
关键词
Dry Powder Inhaler; Discrete Element Method; Computational Fluid Dynamics; Dose loading; Fine Particle Fraction; CLASSIFIER TECHNOLOGY ACT; MICRON-SIZED PARTICLES; CARRIER PARTICLES; DISPERSION MECHANISMS; DELIVERY PERFORMANCE; AIR-FLOW; DESIGN; INHALATION; DRUG; DEPOSITION;
D O I
10.1016/j.ijpharm.2021.121219
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The transport and aerosolization of particles are studied in several different dry powder inhaler geometries via Computational Fluid Dynamics-Discrete Element Method (CFD-DEM) simulations. These simulations combine Large Eddy Simulation of gas with Discrete Element Model simulation of all the carrier particles and a representative subset of the active pharmaceutical ingredient (API) particles. The purpose of the study is to probe the dominant mechanism leading to the release of the API particles and to demonstrate the value of the CFD-DEM simulations where one tracks the motion of all the carrier and API particles. Simulations are performed at different inhalation rates and initial dose loading conditions for the screen-haler geometry, a simple cylindrical tube inhaler, and five different geometry modifications that took the form of bumpy walls and baffles. These geometry modifications alter the residence time of the powder sample in the inhaler, pressure drop across the inhaler, the severity of gas-carrier interactions, and the number of collisions experienced by the carrier particles, all of which are quantified. The quality of aerosolization is found to correlate with the average air-carrier slip velocity, while collisions played only a secondary role. Some geometry modifications improved aerosolization quality with very little increase in the pressure drop across the device.
引用
收藏
页数:13
相关论文
共 36 条
  • [1] Device and Measurement Method of Dose in Dry Powder Inhalers
    Esteban Gorgojo, I.
    Calvo, A.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2020, 201
  • [2] Effects of Device and Formulation on In Vitro Performance of Dry Powder Inhalers
    Adams, Wallace P.
    Lee, Sau L.
    Plourde, Robert
    Lionberger, Robert A.
    Bertha, Craig M.
    Doub, William H.
    Bovet, Jean-Marc
    Hickey, Anthony J.
    AAPS JOURNAL, 2012, 14 (03): : 400 - 409
  • [3] Effects of Device and Formulation on In Vitro Performance of Dry Powder Inhalers
    Wallace P. Adams
    Sau L. Lee
    Robert Plourde
    Robert A. Lionberger
    Craig M. Bertha
    William H. Doub
    Jean-Marc Bovet
    Anthony J. Hickey
    The AAPS Journal, 2012, 14 : 400 - 409
  • [4] Effect of Device Design and Formulation on the In Vitro Comparability for Multi-Unit Dose Dry Powder Inhalers
    Shur, Jagdeep
    Saluja, Bhawana
    Lee, Sau
    Tibbatts, James
    Price, Robert
    AAPS JOURNAL, 2015, 17 (05): : 1105 - 1116
  • [5] Effect of Device Design and Formulation on the In Vitro Comparability for Multi-Unit Dose Dry Powder Inhalers
    Jagdeep Shur
    Bhawana Saluja
    Sau Lee
    James Tibbatts
    Robert Price
    The AAPS Journal, 2015, 17 : 1105 - 1116
  • [6] Continued investigation into the influence of loaded dose on the performance of dry powder inhalers: Surface smoothing effects
    El-Sabawi, Dina
    Edge, Stephen
    Price, Robert
    Young, Paul M.
    DRUG DEVELOPMENT AND INDUSTRIAL PHARMACY, 2006, 32 (10) : 1135 - 1138
  • [7] DRY POWDER INHALERS ARE BIOEQUIVALENT TO METERED-DOSE INHALERS - A STUDY USING A NEW URINARY ALBUTEROL (SALBUTAMOL) ASSAY TECHNIQUE
    HINDLE, M
    NEWTON, DAG
    CHRYSTYN, H
    CHEST, 1995, 107 (03) : 629 - 633
  • [8] Influence of modified induction port, modified DUSA assembly and device air-inlet geometry on the aerosolization pattern of a dry powder inhaler
    Mehta, Piyush P.
    Kadam, Shivajirao S.
    Pawar, Atmaram P.
    JOURNAL OF DRUG DELIVERY SCIENCE AND TECHNOLOGY, 2020, 55
  • [9] Effect of Device Design on the Aerosolization of a Carrier-Based Dry Powder Inhaler—a Case Study on Aerolizer® Foradile®
    Qi Tony Zhou
    Zhenbo Tong
    Patricia Tang
    Mauro Citterio
    Runyu Yang
    Hak-Kim Chan
    The AAPS Journal, 2013, 15 : 511 - 522
  • [10] Impact of critical handling errors on delivered budesonide/formoterol dose in dry powder inhalers - An in vitro study
    Loof, Thomas
    Telg, Gunilla
    Stratelis, Georgios
    EUROPEAN RESPIRATORY JOURNAL, 2016, 48