Triaxial fluxgate gradiometer of high stability and linearity

被引:33
作者
Merayo, JMG [1 ]
Brauer, P [1 ]
Primdahl, F [1 ]
机构
[1] Tech Univ Denmark, Orsted DTU Measurement & Instrumentat Syst, DK-2800 Lyngby, Denmark
关键词
gradiometry; magnetometry; fluxgate; calibration; geomagnetism; magnetic gradient; linearity;
D O I
10.1016/j.sna.2004.11.014
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A novel highly stable magnetic fluxgate vector gradiometer is presented in this paper. It is based on two triaxial fluxgate sensors with the Compact Spherical Coil (CSC) feedback to demonstrate the feasibility of such instrument (the full gradient instrument will consist of at least four triaxial sensors for measuring all the components of the gradient tensor). The sensors have been designed and constructed for the geopotential German satellite CHAMP, and are based on the instrument flying on the Danish satellite Orsted dedicated to measure the Earth's magnetic field with very high precision. The transducers are of the ringcore type with very low noise and high thermal stability. They use amorphous metal magnetic core (Vitrovac 6025). The cores have been annealed in two different processes with different temperatures and stress. With this instrument, three components of the gradient tensor can be measured with a precision better than 100 pT(rms)/m. The noise of the gradiometer is 93 pT(rms)/m in the band 0.01-10 Hz (30.1 pT(rms)/(Hz(1/2) m) at 1 Hz). (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:71 / 77
页数:7
相关论文
共 8 条
[1]   Transverse field effect in fluxgate sensors [J].
Brauer, P ;
Merayo, JMG ;
Nielsen, OV ;
Primdahl, F ;
Petersen, JR .
SENSORS AND ACTUATORS A-PHYSICAL, 1997, 59 (1-3) :70-74
[2]  
JORGENSEN J, 1998, P TETH TECHN INT M N, P239
[3]   Scalar calibration of vector magnetometers [J].
Merayo, JMG ;
Brauer, P ;
Primdahl, F ;
Petersen, JR ;
Nielsen, OV .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2000, 11 (02) :120-132
[4]   A portable single axis magnetic gradiometer [J].
Merayo, JMG ;
Petersen, JR ;
Nielsen, OV ;
Primdahl, F ;
Brauer, P .
SENSORS AND ACTUATORS A-PHYSICAL, 2001, 93 (03) :185-196
[5]   The orthogonalization of magnetic systems [J].
Merayo, JMG ;
Primdahl, F ;
Brauer, P ;
Risbo, T ;
Olsen, N ;
Sabaka, T .
SENSORS AND ACTUATORS A-PHYSICAL, 2001, 89 (03) :185-196
[6]   DEVELOPMENT, CONSTRUCTION AND ANALYSIS OF THE ORSTED FLUXGATE MAGNETOMETER [J].
NIELSEN, OV ;
PETERSEN, JR ;
PRIMDAHL, F ;
BRAUER, P ;
HERNANDO, B ;
FERNANDEZ, A ;
MERAYO, JMG ;
RIPKA, P .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1995, 6 (08) :1099-1115
[7]   COMPACT SPHERICAL COIL FOR FLUXGATE MAGNETOMETER VECTOR FEEDBACK [J].
PRIMDAHL, F ;
JENSEN, PA .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1982, 15 (02) :221-226
[8]  
RIPKA, 2001, MAGNETIC SENSORS MAN