Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT

被引:165
|
作者
Pelgrim, Gert Jan [1 ]
van Hamersvelt, Robbert W. [2 ]
Willemink, Martin J. [2 ]
Schmidt, Bernhard T. [3 ]
Flohr, Thomas [3 ]
Schilham, Arnold [2 ]
Milles, Julien [4 ]
Oudkerk, Matthijs [1 ]
Leiner, Tim [2 ]
Vliegenthart, Rozemarijn [1 ,5 ]
机构
[1] Univ Groningen, Univ Med Ctr Groningen, Ctr Med Imaging North East Netherlands, POB EB44,Hanzepl 1, NL-9713 GZ Groningen, Netherlands
[2] Univ Utrecht, Med Ctr, Dept Radiol, Utrecht, Netherlands
[3] Siemens Healthcare GmbH, Forchheim, Germany
[4] Philips Healthcare, Best, Netherlands
[5] Univ Groningen, Univ Med Ctr Groningen, Dept Radiol, Groningen, Netherlands
关键词
Tomography; x-ray computed; Absorptiometry; photon; Myocardial perfusion imaging; Iodine; Phantoms; imaging; COMPUTED-TOMOGRAPHY; IMAGE-QUALITY; PHANTOM;
D O I
10.1007/s00330-017-4752-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To determine the accuracy of iodine quantification with dual energy computed tomography (DECT) in two high-end CT systems with different spectral imaging techniques. Five tubes with different iodine concentrations (0, 5, 10, 15, 20 mg/ml) were analysed in an anthropomorphic thoracic phantom. Adding two phantom rings simulated increased patient size. For third-generation dual source CT (DSCT), tube voltage combinations of 150Sn and 70, 80, 90, 100 kVp were analysed. For dual layer CT (DLCT), 120 and 140 kVp were used. Scans were repeated three times. Median normalized values and interquartile ranges (IQRs) were calculated for all kVp settings and phantom sizes. Correlation between measured and known iodine concentrations was excellent for both systems (R = 0.999-1.000, p < 0.0001). For DSCT, median measurement errors ranged from -0.5% (IQR -2.0, 2.0%) at 150Sn/70 kVp and -2.3% (IQR -4.0, -0.1%) at 150Sn/80 kVp to -4.0% (IQR -6.0, -2.8%) at 150Sn/90 kVp. For DLCT, median measurement errors ranged from -3.3% (IQR -4.9, -1.5%) at 140 kVp to -4.6% (IQR -6.0, -3.6%) at 120 kVp. Larger phantom sizes increased variability of iodine measurements (p < 0.05). Iodine concentration can be accurately quantified with state-of-the-art DECT systems from two vendors. The lowest absolute errors were found for DSCT using the 150Sn/70 kVp or 150Sn/80 kVp combinations, which was slightly more accurate than 140 kVp in DLCT.
引用
收藏
页码:3904 / 3912
页数:9
相关论文
共 50 条
  • [21] Iodine quantification to distinguish hepatic neuroendocrine tumor metastasis from hepatocellular carcinoma at dual-source dual-energy liver CT
    Kaltenbach, Benjamin
    Wichmann, Julian L.
    Pfeifer, Sophia
    Albrecht, Moritz H.
    Booz, Christian
    Lenga, Lukas
    Hammerstingl, Renate
    Dangelo, Tommaso
    Vogl, Thomas J.
    Martin, Simon S.
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 105 : 20 - 24
  • [22] Advances of dual source, dual-energy imaging in postmortem CT
    Persson, Anders
    Jackowski, Christian
    Engstrom, Elias
    Zachrisson, Helene
    EUROPEAN JOURNAL OF RADIOLOGY, 2008, 68 (03) : 446 - 455
  • [23] Multivendor Comparison of Quantification Accuracy of Iodine Concentration and Attenuation Measurements by Dual-Energy CT: A Phantom Study
    Chen, Yong
    Zhong, Jingyu
    Wang, Lan
    Shi, Xiaomeng
    Chang, Rui
    Fan, Jing
    Jiang, Jiang
    Xia, Yihan
    Yan, Fuhua
    Yao, Weiwu
    Zhang, Huan
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 219 (05) : 827 - 839
  • [24] Intrinsic limitations on quantification accuracy of dual energy CT at low dose levels
    Cruz-Bastida, Juan P.
    Zhang, Ran
    Li, Ke
    Chen, Guang-Hong
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [25] Diagnostic accuracy of third-generation dual-source dual-energy CT: a prospective trial and protocol for clinical implementation
    Tim Nestler
    Kai Nestler
    Andreas Neisius
    Hendrik Isbarn
    Christopher Netsch
    Stephan Waldeck
    Hans U. Schmelz
    Christian Ruf
    World Journal of Urology, 2019, 37 : 735 - 741
  • [26] Diagnostic accuracy of third-generation dual-source dual-energy CT: a prospective trial and protocol for clinical implementation
    Nestler, Tim
    Nestler, Kai
    Neisius, Andreas
    Isbarn, Hendrik
    Netsch, Christopher
    Waldeck, Stephan
    Schmelz, Hans U.
    Ruf, Christian
    WORLD JOURNAL OF UROLOGY, 2019, 37 (04) : 735 - 741
  • [27] Initial experience of dual-energy lung perfusion CT using a dual-source CT system in children
    Goo, Hyun Woo
    PEDIATRIC RADIOLOGY, 2010, 40 (09) : 1536 - 1544
  • [28] Initial experience of dual-energy lung perfusion CT using a dual-source CT system in children
    Hyun Woo Goo
    Pediatric Radiology, 2010, 40 : 1536 - 1544
  • [29] Accuracy of iodine quantification in dual-layer spectral CT: Influence of iterative reconstruction, patient habitus and tube parameters
    Sauter, Andreas P.
    Kopp, Felix K.
    Muenzel, Daniela
    Dangelmaier, Julia
    Renz, Martin
    Renger, Bernhard
    Braren, Rickmer
    Fingerle, Alexander A.
    Rummeny, Ernst J.
    Noeel, Peter B.
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 102 : 83 - 88
  • [30] In vivo measurement of iron concentration using dual-source, dual-energy CT
    Weavers, Paul T.
    Jacobsen, Megan
    Liu, Xin
    Morin, Richard L.
    McCollough, Cynthia H.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622