Between-examiner reproducibility in manual differential leukocyte counting

被引:23
作者
Fuentes-Arderiu, Xavier [1 ]
Garcia-Panyella, Mariona [1 ]
Dot-Bach, Dolors [1 ]
机构
[1] Univ Hosp Bellvitge, IDIBELL, Clin Lab, Catalonia 08907, Spain
关键词
imprecision; reproducibility; laboratory practice; leukocytes; quality control; statistics;
D O I
10.1007/s00769-007-0323-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In manual-visual differential leukocyte counts, the counting itself may be different depending on the examiner. We estimated the between-examiner (B-E) reproducibility in differential leukocyte counting. During 2 months, daily, after performing manual-visual differential leukocyte counts, two slides with blood films were retained. Next day these two blood films were re-examined by different examiners. For each type of leukocyte the B-E reproducibility was estimated from the 58 pairs of data obtained by four technicians. The B-E coefficient of variation for each type of leukocyte was: basophilocytes, 263.2%; eosinophilocytes, 68.8%; lymphocytes, 32.5%; metamyelocytes, 69.6%; monocytes, 55.0%; myelocytes, 132.5%; and neutrophilocytes, 6.6%. For each type of leukocyte the coefficient of variation is the sought estimate of the average reproducibility due to the "human factor" as a whole, without regard to any examiner in particular. Knowledge of this component of day-to-day imprecision, and its follow-up, may be used to decide corrective actions (training of the examiner, etc.) and to set critical differences to interpret the significance of changes in serial results.
引用
收藏
页码:643 / 645
页数:3
相关论文
共 7 条
  • [1] Between-examiner reproducibility in manual differential leukocyte counting
    Xavier Fuentes-Arderiu
    Mariona García-Panyella
    Dolors Dot-Bach
    Accreditation and Quality Assurance, 2007, 12 : 643 - 645
  • [2] Standardized manual palpation of myofascial trigger points in relation to neck/shoulder pain; the influence of clinical experience on inter-examiner reproducibility
    Myburgh, Corrie
    Lauridsen, Henrik Hein
    Larsen, Anders H.
    Hartvigsen, Jan
    MANUAL THERAPY, 2011, 16 (02) : 136 - 140
  • [3] Relationship between total and differential leukocyte counts and isolated coronary artery ectasia
    Kocaman, Sinan Altan
    Tapoy, Guelten
    Sahinarslan, Asife
    Cengel, Atiye
    CORONARY ARTERY DISEASE, 2008, 19 (05) : 307 - 310
  • [4] Relationship between total and differential leukocyte counts and isolated coronary artery ectasia
    Kocaman, S. A.
    Tacoy, G.
    Sahinarslan, A.
    Timurkaynak, T.
    Cengel, A.
    CIRCULATION, 2008, 118 (12) : E180 - E180
  • [5] How Much Does the Platelet Aggregate Influence the Total Leukocyte Count? Comparison Between Manual and Automated Impedance Methods in Domestic Cats
    Zanfagnini, Leticia Gomes
    Kassab, Siham
    Campos, Diefrey
    de Oliveira Chaves, Jessica Karoline
    Pacheco, Acacio Duarte
    ACTA SCIENTIAE VETERINARIAE, 2021, 49
  • [6] Intra- and inter-tester reproducibility of venous occlusion plethysmography: comparison between a manual and a semi-automatic method of blood flow analysis
    Silva, Bruno Moreira
    Neves, Fabricia Junqueira
    Rocha, Natalia Galito
    Cagy, Mauricio
    de Souza, Marcio Nogueira
    Lucas da Nobrega, Antonio Claudio
    PHYSIOLOGICAL MEASUREMENT, 2009, 30 (11) : 1267 - 1279
  • [7] Auto-segmentation, radiomic reproducibility, and comparison of radiomics between manual and AI-derived segmentations for coronary arteries in cardiac [18F]NaF PET/CT images
    Suning Li
    Jake Kendrick
    Martin A. Ebert
    Ghulam Mubashar Hassan
    Nathaniel Barry
    Keaton Wright
    Sing Ching Lee
    Jamie W. Bellinge
    Carl Schultz
    EJNMMI Physics, 12 (1)