关键词
LEFT-HEART BYPASS; ASSIST DEVICE; VASOCONSTRICTIVE FUNCTION; ANIMAL-MODEL; NONPULSATILE; FAILURE; FLOW; PRESSURE; DYSFUNCTION; PROGRESS;
D O I
10.1097/MAT.0b013e3181e7bf3c
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Debate exists regarding the merits and limitations of continuous versus pulsatile flow mechanical circulatory support. To characterize the hemodynamic differences between each mode of support, we investigated the acute effects of continuous versus pulsatile unloading of the failing left ventricle in a bovine model. Heart failure was induced in male calves (n = 14). During an acute study, animals were instrumented through thoracotomy for hemodynamic measurement. A continuous flow (n = 8) and/or pulsatile flow (n = 8) left ventricular assist device (LVAD) was implanted and studied during maximum support (similar to 5 L/min) and moderate support (similar to 2-3 L/min) modes. Pulse pressure (PP), surplus hemodynamic energy (SHE), and (energy equivalent pressure [EEP]/mean aortic pressure (MAP) - 1) x 100% were derived to characterize hemodynamic energy profiles during the different support modes. Standard hemodynamic parameters of cardiac performance were also derived. Data were analyzed by repeated measures one-way analysis of variance within groups and unpaired Student's t-tests across groups. During maximum and moderate continuous unloading, PP, SHE, and (EEP/MAP - 1) x 100% were significantly decreased compared with baseline and compared with pulsatile unloading. As a result, continuous unloading significantly altered left ventricular peak systolic pressure, aortic systolic and diastolic pressure, +/- dP/dt, and rate x pressure product, whereas pulsatile unloading preserved a normal profile of physiologic values. As continuous unloading increased, the pressure-volume relationship collapsed, and the aortic valve remained closed. In contrast, as pulsatile unloading increased, a comparable decrease in left ventricular volumes was noted. However, a normal range of left ventricular pressures was preserved. Continuous unloading deranged the physiologic profile of myocardial and vascular hemodynamic energy utilization, whereas pulsatile unloading preserved more normal physiologic values. These findings may have important implications for chronic LVAD therapy. ASAIO Journal 2010; 56:410-416.
引用
收藏
页码:410 / 416
页数:7
相关论文
共 36 条
  • [1] EFFECTS OF GRADED PULSATILE PRESSURE ON REFLEX VASOMOTOR RESPONSES ELICITED BY CHANGES OF MEAN PRESSURE IN PERFUSED CAROTID SINUS-AORTIC ARCH REGIONS OF DOG
    ANGELLJA.JE
    DALY, DB
    [J]. JOURNAL OF PHYSIOLOGY-LONDON, 1971, 214 (01): : 51 - +
  • [2] BARTOLI CR, 2010, ARTIF ORGAN IN PRESS
  • [3] The role of bridge to transplantation: should LVAD patients be transplanted?
    Birks, EJ
    Yacoub, MH
    Banner, NR
    Khaghani, A
    [J]. CURRENT OPINION IN CARDIOLOGY, 2004, 19 (02) : 148 - 153
  • [4] Left ventricular assist device and drug therapy for the reversal of heart failure
    Birks, Emma J.
    Tansley, Patrick D.
    Hardy, James
    George, Robert S.
    Bowles, Christopher T.
    Burke, Margaret
    Banner, Nicholas R.
    Khaghani, Asghar
    Yacoub, Magdi H.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2006, 355 (18) : 1873 - 1884
  • [5] Gastrointestinal bleeding rates in recipients of nonpulsatile and pulsatile left ventricular assist devices
    Crow, Sheri
    John, Ranjit
    Boyle, Andrew
    Shumway, Sara
    Liao, Kenneth
    Colvin-Adams, Monica
    Toninato, Carol
    Missov, Emil
    Pritzker, Marc
    Martin, Cindy
    Garry, Daniel
    Thomas, William
    Joyce, Lyle
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 137 (01) : 208 - 215
  • [6] Left ventricular assist system as a bridge to myocardial recovery
    Frazier, OH
    Myers, TJ
    [J]. ANNALS OF THORACIC SURGERY, 1999, 68 (02) : 734 - 741
  • [7] Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease
    Geisen, Ulrich
    Heilmann, Claudia
    Beyersdorf, Friedhelm
    Benk, Christoph
    Berchtold-Herz, Michael
    Schlensak, Christian
    Budde, Ulrich
    Zieger, Barbara
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2008, 33 (04) : 679 - 684
  • [8] Left ventricular and myocardial perfusion responses to volume unloading and afterload reduction in a computer simulation
    Giridharan, GA
    Ewert, DL
    Pantalos, GM
    Gillars, KJ
    Litwak, KN
    Gray, LA
    Koenig, SC
    [J]. ASAIO JOURNAL, 2004, 50 (05) : 512 - 518
  • [9] Medical progress - Implantable left ventricular assist devices
    Goldstein, DJ
    Oz, MC
    Rose, EA
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (21) : 1522 - 1533
  • [10] Hemodynamic and exercise performance with Pulsatile and continuous-flow left ventricular assist devices
    Haft, Jonathan
    Armstrong, William
    Dyke, David B.
    Aaronson, Keith D.
    Koelling, Todd M.
    Farrar, David J.
    Pagani, Francis D.