First Clinical Photon-counting Detector CT System: Technical Evaluation

被引:322
|
作者
Rajendran, Kishore [1 ]
Petersilka, Martin [2 ]
Henning, Andre [2 ]
Shanblatt, Elisabeth R. [3 ]
Schmidt, Bernhard [2 ]
Flohr, Thomas G. [2 ]
Ferrero, Andrea [1 ]
Baffour, Francis [1 ]
Diehn, Felix E. [1 ]
Yu, Lifeng [1 ]
Rajiah, Prabhakar [1 ]
Fletcher, Joel G. [1 ]
Leng, Shuai [1 ]
McCollough, Cynthia H. [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[2] Siemens Healthineers, Forchheim, Germany
[3] Siemens Med Solut, Malvern, PA USA
基金
美国国家卫生研究院;
关键词
COMPUTED-TOMOGRAPHY; NOISE; ACCURACY;
D O I
10.1148/radiol.212579
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The first clinical CT system to use photon-counting detector (PCD) technology has become available for patient care. Purpose: To assess the technical performance of the PCD CT system with use of phantoms and representative participant examinations. Materials and Methods: Institutional review board approval and written informed consent from four participants were obtained. Technical performance of a dual-source PCD CT system was measured for standard and high-spatial-resolution (HR) collimations. Noise power spectrum, modulation transfer function, section sensitivity profile, iodine CT number accuracy in virtual monoenergetic images (VMIs), and iodine concentration accuracy were measured. Four participants were enrolled (between May 2021 and August 2021) in this prospective study and scanned using similar or lower radiation doses as their respective clinical examinations performed on the same day using energy-integrating detector (EID) CT. Image quality and findings from the participants' PCD CT and EID CT examinations were compared. Results: All standard technical performance measures met accreditation and regulatory requirements. Relative to filtered back-projection reconstructions, images from iterative reconstruction had lower noise magnitude but preserved noise power spectrum shape and peak frequency. Maximum in-plane spatial resolutions of 125 and 208 mu m were measured for HR and standard PCD CT scans, respectively. Minimum values for section sensitivity profile full width at half maximum measurements were 0.34 mm (0.2-mm nominal section thickness) and 0.64 mm (0.4-mm nominal section thickness) for HR and standard PCD CT scans, respectively. In a 120-kV standard PCD CT scan of a 40-cm phantom, VMI iodine CT numbers had a mean percentage error of 5.7%, and iodine concentration had root mean squared error of 0.5 mg/cm(3), similar to previously reported values for EID CT. VMIs, iodine maps, and virtual noncontrast images were created for a coronary CT angiogram acquired with 66-msec temporal resolution. Participant PCD CT images showed up to 47% lower noise and/or improved spatial resolution compared with EID CT. Conclusion: Technical performance of clinical photon-counting detector (PCD) CT is improved relative to that of a current stateof-the-art CT system. The dual-source PCD geometry facilitated 66-msec temporal resolution multienergy cardiac imaging. Study participant images illustrated the effect of the improved technical performance. (C) RSNA, 2022
引用
收藏
页码:130 / 138
页数:9
相关论文
共 50 条
  • [31] Clinical applications of photon counting detector CT
    McCollough, Cynthia H.
    Rajendran, Kishore
    Baffour, Francis I.
    Diehn, Felix E.
    Ferrero, Andrea
    Glazebrook, Katrina N.
    Horst, Kelly K.
    Johnson, Tucker F.
    Leng, Shuai
    Mileto, Achille
    Rajiah, Prabhakar Shantha
    Schmidt, Bernhard
    Yu, Lifeng
    Flohr, Thomas G.
    Fletcher, Joel G.
    EUROPEAN RADIOLOGY, 2023, 33 (08) : 5309 - 5320
  • [32] Metal Artifact Reduction in Photon-Counting Detector CT Quantitative Evaluation of Artifact Reduction Techniques
    Skornitzke, Stephan
    Mergen, Victor
    Biederer, Juergen
    Alkadhi, Hatem
    Do, Thuy D.
    Stiller, Wolfram
    Frauenfelder, Thomas
    Kauczor, Hans-Ulrich
    Euler, Andre
    INVESTIGATIVE RADIOLOGY, 2024, 59 (06) : 442 - 449
  • [33] Standardizing technical parameters and terms for abdominopelvic photon-counting CT: laying the groundwork for innovation and evidence sharing
    Leng, Shuai
    Toia, Giuseppe V.
    Hoodeshenas, Safa
    Ramirez-Giraldo, Juan Carlos
    Yagil, Yoad
    Maltz, Jonathan S.
    Boedeker, Kirsten
    Li, Ke
    Baffour, Francis
    Fletcher, Joel G.
    ABDOMINAL RADIOLOGY, 2024, 49 (09) : 3261 - 3273
  • [34] Quantum Iterative Reconstruction for Abdominal Photon-counting Detector CT Improves Image Quality
    Sartoretti, Thomas
    Landsmann, Anna
    Nakhostin, Dominik
    Eberhard, Matthias
    Roeren, Christian
    Mergen, Victor
    Higashigaito, Kai
    Raupach, Rainer
    Alkadhi, Hatem
    Euler, Andre
    RADIOLOGY, 2022, 303 (02)
  • [35] Photon-Counting Detector CT: Key Points Radiologists Should Know
    Esquivel, Andrea
    Ferrero, Andrea
    Mileto, Achille
    Baffour, Francis
    Horst, Kelly
    Rajiah, Prabhakar Shantha
    Inoue, Akitoshi
    Leng, Shuai
    McCollough, Cynthia
    Fletcher, Joel G.
    KOREAN JOURNAL OF RADIOLOGY, 2022, 23 (09) : 854 - 865
  • [36] Full field-of-view, high-resolution, photon-counting detector CT: technical assessment and initial patient experience
    Rajendran, Kishore
    Petersilka, Martin
    Henning, Andre
    Shanblatt, Elisabeth
    Marsh, Jeffrey, Jr.
    Thorne, Jamison
    Schmidt, Bernhard
    Flohr, Thomas
    Fletcher, Joel
    McCollough, Cynthia
    Leng, Shuai
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (20):
  • [37] First Experience With a Whole-Body Spectral Photon-Counting CT Clinical Prototype
    Si-Mohamed, Salim A.
    Boccalini, Sara
    Villien, Marjorie
    Yagil, Yoad
    Erhard, Klaus
    Boussel, Loic
    Douek, Philippe C.
    INVESTIGATIVE RADIOLOGY, 2023, 58 (07) : 459 - 471
  • [38] Interweaving Network: A Novel Monochromatic Image Synthesis Method for a Photon-Counting Detector CT System
    Zheng, Ao
    Yang, Hongkai
    Zhang, Li
    Xing, Yuxiang
    IEEE ACCESS, 2020, 8 : 217701 - 217710
  • [39] Evaluation of a preclinical photon-counting CT prototype for pulmonary imaging
    Kopp, Felix K.
    Daerr, Heiner
    Si-Mohamed, Salim
    Sauter, Andreas P.
    Ehn, Sebastian
    Fingerle, Alexander A.
    Brendel, Bernhard
    Pfeiffer, Franz
    Roessl, Ewald
    Rummeny, Ernst J.
    Pfeiffer, Daniela
    Proksa, Roland
    Douek, Philippe
    Noel, Peter B.
    SCIENTIFIC REPORTS, 2018, 8
  • [40] Clinical commissioning of the first point-of-care spectral photon-counting CT for the upper extremities
    Manzano, Lucia Gallego
    Monnin, Pascal
    Sayous, Yann
    Becce, Fabio
    Damet, Jerome
    Viry, Anais
    MEDICAL PHYSICS, 2023, 50 (05) : 2844 - 2859