First Clinical Photon-counting Detector CT System: Technical Evaluation

被引:322
|
作者
Rajendran, Kishore [1 ]
Petersilka, Martin [2 ]
Henning, Andre [2 ]
Shanblatt, Elisabeth R. [3 ]
Schmidt, Bernhard [2 ]
Flohr, Thomas G. [2 ]
Ferrero, Andrea [1 ]
Baffour, Francis [1 ]
Diehn, Felix E. [1 ]
Yu, Lifeng [1 ]
Rajiah, Prabhakar [1 ]
Fletcher, Joel G. [1 ]
Leng, Shuai [1 ]
McCollough, Cynthia H. [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[2] Siemens Healthineers, Forchheim, Germany
[3] Siemens Med Solut, Malvern, PA USA
基金
美国国家卫生研究院;
关键词
COMPUTED-TOMOGRAPHY; NOISE; ACCURACY;
D O I
10.1148/radiol.212579
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The first clinical CT system to use photon-counting detector (PCD) technology has become available for patient care. Purpose: To assess the technical performance of the PCD CT system with use of phantoms and representative participant examinations. Materials and Methods: Institutional review board approval and written informed consent from four participants were obtained. Technical performance of a dual-source PCD CT system was measured for standard and high-spatial-resolution (HR) collimations. Noise power spectrum, modulation transfer function, section sensitivity profile, iodine CT number accuracy in virtual monoenergetic images (VMIs), and iodine concentration accuracy were measured. Four participants were enrolled (between May 2021 and August 2021) in this prospective study and scanned using similar or lower radiation doses as their respective clinical examinations performed on the same day using energy-integrating detector (EID) CT. Image quality and findings from the participants' PCD CT and EID CT examinations were compared. Results: All standard technical performance measures met accreditation and regulatory requirements. Relative to filtered back-projection reconstructions, images from iterative reconstruction had lower noise magnitude but preserved noise power spectrum shape and peak frequency. Maximum in-plane spatial resolutions of 125 and 208 mu m were measured for HR and standard PCD CT scans, respectively. Minimum values for section sensitivity profile full width at half maximum measurements were 0.34 mm (0.2-mm nominal section thickness) and 0.64 mm (0.4-mm nominal section thickness) for HR and standard PCD CT scans, respectively. In a 120-kV standard PCD CT scan of a 40-cm phantom, VMI iodine CT numbers had a mean percentage error of 5.7%, and iodine concentration had root mean squared error of 0.5 mg/cm(3), similar to previously reported values for EID CT. VMIs, iodine maps, and virtual noncontrast images were created for a coronary CT angiogram acquired with 66-msec temporal resolution. Participant PCD CT images showed up to 47% lower noise and/or improved spatial resolution compared with EID CT. Conclusion: Technical performance of clinical photon-counting detector (PCD) CT is improved relative to that of a current stateof-the-art CT system. The dual-source PCD geometry facilitated 66-msec temporal resolution multienergy cardiac imaging. Study participant images illustrated the effect of the improved technical performance. (C) RSNA, 2022
引用
收藏
页码:130 / 138
页数:9
相关论文
共 50 条
  • [21] Investigation of abdominal artery delineation by photon-counting detector CT
    Ota, Takashi
    Onishi, Hiromitsu
    Itoh, Toshihide
    Fukui, Hideyuki
    Tsuboyama, Takahiro
    Nakamoto, Atsushi
    Enchi, Yukihiro
    Tatsumi, Mitsuaki
    Tomiyama, Noriyuki
    RADIOLOGIA MEDICA, 2024, 129 (09): : 1265 - 1274
  • [22] New Horizons in Vascular Imaging With Photon-Counting Detector CT
    Wildberger, Joachim E.
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2023, 58 (07) : 499 - 504
  • [23] Soft Reconstruction Kernels Improve HCC Imaging on a Photon-Counting Detector CT
    Graafen, D.
    Mueller, L.
    Halfmann, M. C.
    Stoehr, F.
    Foerster, F.
    Dueber, C.
    Yang, Y.
    Kloeckner, R.
    Emrich, T.
    ACADEMIC RADIOLOGY, 2023, 30 : S143 - S154
  • [24] Evaluation of Coronary Plaques and Stents with Conventional and Photon-counting CT: Benefits of High-Resolution Photon-counting CT
    Rajagopal, Jayasai R.
    Farhadi, Faraz
    Richards, Taylor
    Nikpanah, Moozhan
    Sahbaee, Pooyan
    Shanbhag, Sujata M.
    Bandettini, W. Patricia
    Saboury, Babak
    Malayeri, Ashkan A.
    Pritchard, William F.
    Jones, Elizabeth C.
    Samei, Ehsan
    Chen, Marcus Y.
    RADIOLOGY-CARDIOTHORACIC IMAGING, 2021, 3 (05):
  • [25] Computed tomography with a full FOV photon-counting detector in a clinical setting, the first experience
    Ferda, Jiri
    Vendis, Tomas
    Flohr, Thomas
    Schmidt, Bernhard
    Henning, Andre
    Ulzheimer, Stefan
    Pecen, Ladislav
    Ferdova, Eva
    Baxa, Jan
    Mirka, Hynek
    EUROPEAN JOURNAL OF RADIOLOGY, 2021, 137
  • [26] Evaluation of conventional imaging performance in a research whole-body CT system with a photon-counting detector array
    Yu, Zhicong
    Leng, Shuai
    Jorgensen, Steven M.
    Li, Zhoubo
    Gutjahr, Ralf
    Chen, Baiyu
    Halaweish, Ahmed F.
    Kappler, Steffen
    Yu, Lifeng
    Ritman, Erik L.
    McCollough, Cynthia H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (04): : 1572 - 1595
  • [27] A Universal Protocol for Abdominal CT Examinations Performed on a Photon-Counting Detector CT System A Feasibility Study
    Zhou, Wei
    Michalak, Gregory J.
    Weaver, Jayse M.
    Gong, Hao
    Yu, Lifeng
    McCollough, Cynthia H.
    Leng, Shuai
    INVESTIGATIVE RADIOLOGY, 2020, 55 (04) : 226 - 232
  • [28] An Experimental Evaluation of Material Separability in Photon-counting CT
    Rajagopal, Jayasai R.
    Farhadi, Faraz
    Negussie, Ayele H.
    Abadi, Ehsan
    Sahbaee, Pooyan
    Saboury, Babak
    Malayeri, Ashkan A.
    Pritchard, William F.
    Jones, Elizabeth C.
    Samei, Ehsan
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [29] Photon-counting CT: Review of initial clinical results
    van der Bie, Judith
    van Straten, Marcel
    Booij, Ronald
    Bos, Daniel
    Dijkshoorn, Marcel L.
    Hirsch, Alexander
    Sharma, Simran P.
    Oei, Edwin H. G.
    Budde, Ricardo P. J.
    EUROPEAN JOURNAL OF RADIOLOGY, 2023, 163
  • [30] Photon-counting CT review
    Flohr, Thomas
    Petersilka, Martin
    Henning, Andre
    Ulzheimer, Stefan
    Ferda, Jiri
    Schmidt, Bernhard
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 : 126 - 136