Development and validation of HPLC method for the determination of Candesartan in human plasma

被引:0
|
作者
Hamidi, Shaista [1 ]
Beg, Anwar Ejaz [2 ]
Muhammad, Iyad Naeem [1 ]
Hassan, Sohail [3 ]
Hassan, Amir [3 ]
Akram, Arfa [4 ]
Alladin, Neyama [5 ]
机构
[1] Univ Karachi, Fac Pharm & Pharmaceut Sci, Dept Pharmaceut, Karachi, Pakistan
[2] Ziauddin Univ, Dept Pharmaceut, Fac Pharm, Karachi, Pakistan
[3] Univ Karachi, Fac Pharm & Pharmaceut Sci, Dept Pharmaceut Chem, Karachi, Pakistan
[4] FUUAST, Dept Pharmaceut Chem, Fac Pharm, Karachi, Pakistan
[5] Univ Karachi, Int Ctr Chem & Biol Sci, Karachi, Pakistan
关键词
Candesartan; HPLC; human plasma; UV-Fluorescence; CILEXETIL;
D O I
暂无
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Candesartan (CAN), an ARB-blocker, antihypertensive, was analyzed in human plasma by a simple, accurate and precise RP-HPLC (reverse phase-High performance liquid chromatography assay method which was then validated for its accuracy, specificity and precision. The mobile phase has a constitution of acetone, diethylamine and distilled water, while Phosphoric acid was used to adjust the pH to 2.5 +/- 0.1. This mobile phase was run at 1.1ml/min and the fluorescence wavelength was set to 392 nm. A C-18 HPLC, column particle size (5 mu m) Mediterranean Sea (R) L x 1.D. 25cm x 4.6 mm (Supelcosil), with auto sampler injection volume of 30 mu l ,an internal standard Valsartan was utilized for chromatographic detection. Candesartan took a retention time of 6 +/- 0.5 minutes. This method was validated by the parameters of selectivity, accuracy, precision, repeatability, reproducibility, recovery, linearity and stability. Candesartan's calibration curves were found to be linear in the range of 200ng/ml to 3.125ng/ml and the coefficient of determination (r(2)) was found to be 0.99. Analytical recovery obtained was above 88%. Hence, this method has been found to be useful for determining Candesartan in plasma.
引用
收藏
页码:2323 / 2327
页数:5
相关论文
共 50 条
  • [1] DEVELOPMENT AND VALIDATION OF HPLC METHOD FOR THE DETERMINATION OF LOSARTAN IN HUMAN PLASMA
    Veldandi, Uday Kiran
    Yellu, Narsimha Reddy
    Malothu, Nagulu
    Devarakonda, Rama Krishna
    REVIEWS IN ANALYTICAL CHEMISTRY, 2010, 29 (01) : 59 - 67
  • [2] Development and validation of HPLC method for the determination of Cefpodoxime Proxetil in human plasma
    Bashir, Lubna
    Shoaib, Muhammad Harris
    Naz, Shazia
    Yousuf, Rabia Ismail
    Jabeen, Sabahat
    Israr, Fozia
    Siddiqui, Fahad
    PAKISTAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2017, 30 (05) : 1603 - 1607
  • [3] HPLC method development and validation for the determination of Cefaclor in human plasma
    Naz, Shazia
    Shoaib, Muhammad Harris
    Bashir, Lubna
    Yousuf, Rabia Ismail
    Anjum, Fakhsheena
    Siddiqui, Fahad
    Yaseen, Saima
    PAKISTAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2017, 30 (05) : 1645 - 1649
  • [4] Development and Validation of an HPLC Method for Determination of Carbamazepine in Human Plasma
    Volonte, Maria G.
    Vinas, Maria A.
    Gorriti, Carolina E.
    Escales, Maria C.
    Sanchez, Laura A.
    Esperanza Ruiz, M.
    LATIN AMERICAN JOURNAL OF PHARMACY, 2009, 28 (01): : 80 - 85
  • [5] DEVELOPMENT AND VALIDATION OF A HPLC/MS/MS METHOD FOR THE DETERMINATION OF ALBUTEROL IN HUMAN PLASMA
    Snehalatha, B.
    Shankar, D. Gowri
    Rathod, Sudha
    PHARMACOPHORE, 2014, 5 (05): : 647 - 656
  • [6] Development and validation of an HPLC method for the determination of gatifloxacin stability in human plasma
    Al-Dgither, S
    Alvi, SN
    Hammami, MA
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2006, 41 (01) : 251 - 255
  • [7] Development and Validation of a RP-HPLC Method for Determination of Citicoline Monosodium in Human Plasma
    Bindaiya, Shailendra K.
    Sahu, Kapendra
    Bhaisare, Mukesh
    Karthikeyan, C.
    Moorthy, N. S. H. N.
    Mehta, Farhad F.
    Trivedi, Piyush
    LATIN AMERICAN JOURNAL OF PHARMACY, 2011, 30 (04): : 794 - 798
  • [8] Development and Validation of a HPLC Method for Determination of Sertraline in Human Plasma and its Application to Bioequivalence Study
    Xu, Zhi-sheng
    Zhu, Hai-yan
    Tang, Cong-rong
    LATIN AMERICAN JOURNAL OF PHARMACY, 2014, 33 (02): : 326 - 331
  • [9] Development and Validation of Reverse Phase HPLC Method for the Determination of Midazolam in Human Plasma
    Deepakumari, H.N.
    Prashanth, M.K.
    Revanasiddappa, H.D.
    Analytical Chemistry Letters, 2014, 4 (04) : 213 - 223
  • [10] Development and validation of an HPLC method for simultaneous determination of trimethoprim and sulfamethoxazole in human plasma
    Sayar, Esin
    Sahin, Selma
    Cevheroglu, Semsettin
    Hincal, A. Atilla
    EUROPEAN JOURNAL OF DRUG METABOLISM AND PHARMACOKINETICS, 2010, 35 (1-2) : 41 - 46