This paper describes fabrication of a novel electrochemiluminescence (ECL) immunosensor array featuring capture-antibody-decorated single-wall carbon nanotube (SWCNT) forests residing in the bottoms of 10-mu L wells with hydrophobic polymer walls. Silica nanoparticles containing [Ru(bpy)(3)](2+) and secondary antibodies (RuBPY-silica-Ab(2)) are employed in this system for highly sensitive two-analyte detection. Antibodies to prostate specific antigen (PSA) and interleukin-6 (IL-6) were attached to the same RuBPY-silica-Ab(2) particle. The array was fabricated by forming the wells on a conductive pyrolytic graphite chip (1 in. x 1 in.) with a single connection to a potentiostat to achieve ECL. The sandwich immunoassay protocol employs antibodies attached to SWCNTs in the wells to capture analyte proteins. Then RuBPY-silica-Ab(2) is added to bind to the captured proteins. ECL is initiated in the microwells by electrochemical oxidation of tripropyl amine (TprA), which generates excited state [Ru(bpy)(3)](2+) in the 100-nm particles, and is measured with a charge-coupled device (CC D) camera. Separation of the analytical spots by the hydrophobic wall barriers enabled simultaneous immunoassays for two proteins in a single sample without cross-contamination. The detection limit.(DL) for PSA was 1 pg mL(-1) and for IL-6 was 0.25 pg mL(-1) (IL-6) in serum. Array determinations of PSA and IL-6 in patient serum were well-correlated with single-protein ELISAs. These microwell SWCNT immunoarrays provide a simple, sensitive approach to the detection of two or more proteins.
引用
收藏
页码:6698 / 6703
页数:6
相关论文
共 40 条
[1]   Mass spectrometry-based proteomics [J].
Aebersold, R ;
Mann, M .
NATURE, 2003, 422 (6928) :198-207
[2]   Microfluidic electrochemical immunoarray for ultrasensitive detection of two cancer biomarker proteins in serum [J].
Chikkaveeraiah, Bhaskara V. ;
Mani, Vigneshwaran ;
Patel, Vyomesh ;
Gutkind, J. Silvio ;
Rusling, James F. .
BIOSENSORS & BIOELECTRONICS, 2011, 26 (11) :4477-4483
[3]   Single-Wall Carbon Nanotube Forest Arrays for Immunoelectrochemical Measurement of Four Protein Biomarkers for Prostate Cancer [J].
Chikkaveeraiah, Bhaskara V. ;
Bhirde, Ashwin ;
Malhotra, Ruchika ;
Patel, Vyomesh ;
Gutkind, J. Silvio ;
Rusling, James F. .
ANALYTICAL CHEMISTRY, 2009, 81 (21) :9129-9134
[4]  
Debad JD, 2004, MG ELEC CH, P359
[5]   Multiplexed Sandwich Immunoassays Using Electrochemiluminescence Imaging Resolved at the Single Bead Level [J].
Deiss, Frederique ;
LaFratta, Christopher N. ;
Symer, Matthew ;
Blicharz, Timothy M. ;
Sojic, Neso ;
Walt, David R. .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2009, 131 (17) :6088-+
[6]   Electrogenerated Chemiluminescence [J].
Forster, Robert J. ;
Bertoncello, Paolo ;
Keyes, Tia E. .
ANNUAL REVIEW OF ANALYTICAL CHEMISTRY, 2009, 2 :359-385
[7]   Immunosensors: Electrochemical sensing and other engineering approaches [J].
Ghindilis, AL ;
Atanasov, P ;
Wilkins, M ;
Wilkins, E .
BIOSENSORS & BIOELECTRONICS, 1998, 13 (01) :113-131
[8]   Drivers of biodiagnostic development [J].
Giljohann, David A. ;
Mirkin, Chad A. .
NATURE, 2009, 462 (7272) :461-464
[9]  
GOLDSMITH S J, 1975, Seminars in Nuclear Medicine, V5, P125, DOI 10.1016/S0001-2998(75)80028-6
[10]   Mass Spectrometry-Based Biomarker Discovery: Toward a Global Proteome Index of Individuality [J].
Hawkridge, Adam M. ;
Muddiman, David C. .
ANNUAL REVIEW OF ANALYTICAL CHEMISTRY, 2009, 2 :265-277