Commissioning and beam characterization of the first gantry-mounted accelerator pencil beam scanning proton system

被引:44
作者
Kang, M. [1 ]
Cessac, Rob [2 ]
Pang, D. [1 ]
机构
[1] Georgetown Univ Hosp, Dept Radiat Med, Washington, DC 20007 USA
[2] Mevion Med Syst Inc, Littleton, MA USA
关键词
adaptive aperture; beam commissioning; compart PBS system; gantry-mounted accelerator; HYPERSCAN; pencil beam scanning; MONTE-CARLO; THERAPY; INTERPLAY; LUNG; DELIVERY; CANCER; MOTION; PATIENT; IMPACT; TUMOR;
D O I
10.1002/mp.13972
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To present and discuss beam characteristics and commissioning process of the first gantry-mounted accelerator single-room pencil beam scanning (PBS) proton system. Methods: The Mevion HYPERSCAN employs a design configuration with a synchrocyclotron mounted on the gantry to eliminate the traditional beamline and a nozzle that contains the dosimetry monitoring chambers, the energy modulator [energy selector (ES)], and an adaptive aperture (AA). To characterize the beam, we measured the integrated depth dose (IDD) for 12 energies. from the highest energy of 227 MeV down to 28 MeV with a range difference similar to 2 cm between the adjacent energies, using a large radius Bragg peak chamber; single spot profiles in air at five locations along the beam central axis using radiochromic EBT3 film and cross compared with a scintillation detector; and determined the output using a densely packed spot map. To access the performance of AA, we measured interleaf leakage and the penumbra reduction effect. Monte Carlo simulation using TOPAS was performed to study spot size variation along the beam path, beam divergence, and energy spectrum. Results: This proton system is calibrated to deliver 1 Gy dose at 5 cm depth in water using the highest beam energy by delivering 1 MU/spot to a 10 x 10 cm(2) map with a 2.5 mm spot spacing. The spot size in air at isocenter for a maximum range beam of 227 MeV is 4.1 mm. This system is able to reduce the beam range all the way to the patient surface; the lowest energy beam measured was 28 MeV which has a spot size of 15.7 mm. The beam divergence is 2.4 mrad at 227 MeV and 52.7 mrad for the superficial 28 MeV beam. The binary design of the ES has resulted in shifts of the effective SSD toward the isocenter as the energy is modulated lower. The pristine Bragg peaks have a constant 80%-80% width of 8.4 mm at all energies. The interleaf leakage of the AA is less than 1.5% at the highest energy. The AA reduces field penumbras. For a 10x10 optimized field, the 227 MeV beam penumbra measured at isocenter with a 5 cm air gap went from 6.8 to 4.7 mm and the 28 MeV beam penumbra went from 21.4 to 7.5 mm. Conclusions: The HYPERSCAN proton system has a unique design, which is reflected in the Bragg peak shapes, the variation of spot sizes with energy and the penumbra sharpening effect of the AA. The combination of the ES and AA makes PBS implementation possible without using a beam transport line and supplemental range shifter devices. In commissioning the TPS and designing plans these differences need to be considered. (C) 2020 American Association of Physicists in Medicine
引用
收藏
页码:3496 / 3510
页数:15
相关论文
共 34 条
[1]   Radiation oncology in the era of precision medicine [J].
Baumann, Michael ;
Krause, Mechthild ;
Overgaard, Jens ;
Debus, Juergen ;
Bentzen, Soren M. ;
Daartz, Juliane ;
Richter, Christian ;
Zips, Daniel ;
Bortfeld, Thomas .
NATURE REVIEWS CANCER, 2016, 16 (04) :234-249
[2]   The world's first single-room proton therapy facility: Two-year experience [J].
Contreras, Jessika ;
Zhao, Tianyu ;
Perkins, Stephanie ;
Sun, Baozhou ;
Goddu, Sreekrishna ;
Mutic, Sasa ;
Bottani, Beth ;
Endicott, Sharon ;
Michalski, Jeff ;
Robinson, Clifford ;
Tsien, Christina ;
Huang, Jiayi ;
Fischer-Valuck, Benjamin W. ;
Hallahan, Dennis ;
Klein, Eric ;
Bradley, Jeffrey .
PRACTICAL RADIATION ONCOLOGY, 2017, 7 (01) :E71-E76
[3]  
Dong L, 2017, AAPM ED SESS DENV CO
[4]   Interplay effects in proton scanning for lung: a 4D Monte Carlo study assessing the impact of tumor and beam delivery parameters [J].
Dowdell, S. ;
Grassberger, C. ;
Sharp, G. C. ;
Paganetti, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (12) :4137-4156
[5]  
Fagundes M., 2015, Int J Part Ther, V1, P827, DOI [DOI 10.14338/IJPT-14-00025.1, 10.14338/IJPT-14-00025.1]
[6]   Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking [J].
Farr, J. B. ;
Mascia, A. E. ;
Hsi, W. -C. ;
Allgower, C. E. ;
Jesseph, F. ;
Schreuder, A. N. ;
Wolanski, M. ;
Nichiporov, D. F. ;
Anferov, V. .
MEDICAL PHYSICS, 2008, 35 (11) :4945-4954
[7]   Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas MD Anderson Cancer Center, Proton Therapy Center, Houston [J].
Gillin, Michael T. ;
Sahoo, Narayan ;
Bues, Martin ;
Ciangaru, George ;
Sawakuchi, Gabriel ;
Poenisch, Falk ;
Arjomandy, Bijan ;
Martin, Craig ;
Titt, Uwe ;
Suzuki, Kazumichi ;
Smith, Alfred R. ;
Zhu, X. Ronald .
MEDICAL PHYSICS, 2010, 37 (01) :154-163
[8]   Motion Interplay as a Function of Patient Parameters and Spot Size in Spot Scanning Proton Therapy for Lung Cancer [J].
Grassberger, Clemens ;
Dowdell, Stephen ;
Lomax, Antony ;
Sharp, Greg ;
Shackleford, James ;
Choi, Noah ;
Willers, Henning ;
Paganetti, Harald .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 86 (02) :380-386
[9]   Validation and clinical implementation of an accurate Monte Carlo code for pencil beam scanning proton therapy [J].
Huang, Sheng ;
Kang, Minglei ;
Souris, Kevin ;
Ainsley, Christopher ;
Solberg, Timothy D. ;
McDonough, James E. ;
Simone, Charles B. ;
Lin, Liyong .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05) :558-572
[10]   A study of the beam-specific interplay effect in proton pencil beam scanning delivery in lung cancer [J].
Kang, Minglei ;
Huang, Sheng ;
Solberg, Timothy D. ;
Mayer, Rulon ;
Thomas, Andy ;
Teo, Boon-Keng Kevin ;
McDonough, James E. ;
Simone, Charles B., II ;
Lin, Liyong .
ACTA ONCOLOGICA, 2017, 56 (04) :531-540