Commissioning and beam characterization of the first gantry-mounted accelerator pencil beam scanning proton system

被引:40
|
作者
Kang, M. [1 ]
Cessac, Rob [2 ]
Pang, D. [1 ]
机构
[1] Georgetown Univ Hosp, Dept Radiat Med, Washington, DC 20007 USA
[2] Mevion Med Syst Inc, Littleton, MA USA
关键词
adaptive aperture; beam commissioning; compart PBS system; gantry-mounted accelerator; HYPERSCAN; pencil beam scanning; MONTE-CARLO; THERAPY; INTERPLAY; LUNG; DELIVERY; CANCER; MOTION; PATIENT; IMPACT; TUMOR;
D O I
10.1002/mp.13972
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To present and discuss beam characteristics and commissioning process of the first gantry-mounted accelerator single-room pencil beam scanning (PBS) proton system. Methods: The Mevion HYPERSCAN employs a design configuration with a synchrocyclotron mounted on the gantry to eliminate the traditional beamline and a nozzle that contains the dosimetry monitoring chambers, the energy modulator [energy selector (ES)], and an adaptive aperture (AA). To characterize the beam, we measured the integrated depth dose (IDD) for 12 energies. from the highest energy of 227 MeV down to 28 MeV with a range difference similar to 2 cm between the adjacent energies, using a large radius Bragg peak chamber; single spot profiles in air at five locations along the beam central axis using radiochromic EBT3 film and cross compared with a scintillation detector; and determined the output using a densely packed spot map. To access the performance of AA, we measured interleaf leakage and the penumbra reduction effect. Monte Carlo simulation using TOPAS was performed to study spot size variation along the beam path, beam divergence, and energy spectrum. Results: This proton system is calibrated to deliver 1 Gy dose at 5 cm depth in water using the highest beam energy by delivering 1 MU/spot to a 10 x 10 cm(2) map with a 2.5 mm spot spacing. The spot size in air at isocenter for a maximum range beam of 227 MeV is 4.1 mm. This system is able to reduce the beam range all the way to the patient surface; the lowest energy beam measured was 28 MeV which has a spot size of 15.7 mm. The beam divergence is 2.4 mrad at 227 MeV and 52.7 mrad for the superficial 28 MeV beam. The binary design of the ES has resulted in shifts of the effective SSD toward the isocenter as the energy is modulated lower. The pristine Bragg peaks have a constant 80%-80% width of 8.4 mm at all energies. The interleaf leakage of the AA is less than 1.5% at the highest energy. The AA reduces field penumbras. For a 10x10 optimized field, the 227 MeV beam penumbra measured at isocenter with a 5 cm air gap went from 6.8 to 4.7 mm and the 28 MeV beam penumbra went from 21.4 to 7.5 mm. Conclusions: The HYPERSCAN proton system has a unique design, which is reflected in the Bragg peak shapes, the variation of spot sizes with energy and the penumbra sharpening effect of the AA. The combination of the ES and AA makes PBS implementation possible without using a beam transport line and supplemental range shifter devices. In commissioning the TPS and designing plans these differences need to be considered. (C) 2020 American Association of Physicists in Medicine
引用
收藏
页码:3496 / 3510
页数:15
相关论文
共 50 条
  • [1] High spatiotemporal resolution scintillation imaging of pulsed pencil beam scanning proton beams produced by a gantry-mounted synchrocyclotron
    Goddu, S. Murty
    Hao, Yao
    Ji, Zhen
    Setianegara, Jufri
    Liu, Fengwei
    Green, Winter
    Sobotka, Lee G.
    Zhao, Tianyu
    Perkins, Stephanie
    Darafsheh, Arash
    MEDICAL PHYSICS, 2024, 51 (07) : 4996 - 5006
  • [2] Beam Position Projection Algorithms in Proton Pencil Beam Scanning
    Nesteruk, Konrad P.
    Bradley, Stephen G.
    Kooy, Hanne M.
    Clasie, Benjamin M.
    CANCERS, 2024, 16 (11)
  • [3] Comparison of multi-institutional Varian ProBeam pencil beam scanning proton beam commissioning data
    Langner, Ulrich W.
    Eley, John G.
    Dong, Lei
    Langen, Katja
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (03): : 96 - 107
  • [4] Beam characteristics of the first clinical 360° rotational single gantry room scanning pencil beam proton treatment system and comparisons against a multi-room system
    Shang, Charles
    Evans, Grant
    Rahman, Mushfiqur
    Lin, Liyong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (09): : 266 - 271
  • [5] Proton pencil beam scanning for mediastinal lymphoma: the impact of interplay between target motion and beam scanning
    Zeng, C.
    Plastaras, J. P.
    Tochner, Z. A.
    White, B. M.
    Hill-Kayser, C. E.
    Hahn, S. M.
    Both, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (07) : 3013 - 3029
  • [6] Gantry-based pencil beam scanning proton therapy for uveal melanoma: IMPT versus proton arc therapy
    Qi, Hang
    Hu, Lei
    Huang, Sheng
    Lee, Yen-Po
    Yu, Francis
    Chen, Qing
    Yang, Yunjie
    Kang, Minglei
    Zhai, Huifang
    Vermeulen, Milo
    Shim, Andy
    Park, Peter
    Ding, Xuanfeng
    Zhou, Jun
    Abramson, David H.
    Francis, Jasmine H.
    Simone, Charles B.
    Barker, Christopher A.
    Lin, Haibo
    RADIATION ONCOLOGY, 2025, 20 (01)
  • [7] Proton pencil beam scanning for mediastinal lymphoma: treatment planning and robustness assessment
    Zeng, Chuan
    Plastaras, John P.
    James, Paul
    Tochner, Zelig A.
    Hill-Kayser, Christine E.
    Hahn, Stephen M.
    Both, Stefan
    ACTA ONCOLOGICA, 2016, 55 (9-10) : 1132 - 1138
  • [8] Impact of Intrafraction and Residual Interfraction Effect on Prostate Proton Pencil Beam Scanning
    Tang, Shikui
    Deville, Curtiland
    Tochner, Zelig
    Wang, Ken Kang-Hsin
    McDonough, James
    Vapiwala, Neha
    Both, Stefan
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 (05): : 1186 - 1194
  • [9] Contour scanning, multi-leaf collimation and the combination thereof for proton pencil beam scanning
    Winterhalter, Carla
    Meier, Gabriel
    Oxley, David
    Weber, Damien C.
    Lomax, Antony J.
    Safai, Sairos
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (01)
  • [10] Biophysical characterization of collimated and uncollimated fields in pencil beam scanning proton therapy
    Nabha, Racell
    De Saint-Hubert, Marijke
    Marichal, Joachim
    Esser, Johannes
    Van Hoey, Olivier
    Baeumer, Christian
    Verbeek, Nico
    Struelens, Lara
    Sterpin, Edmond
    Tabury, Kevin
    Marek, Lukas
    Granja, Carlos
    Timmermann, Beate
    Vanhavere, Filip
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (06)