Determination of fluid leakages in the different screw-retained implant-abutment connections in a mechanical artificial mouth

被引:5
作者
Martin-Gili, D. [1 ]
Molmeneu, M. [2 ,3 ]
Fernandez, M. [2 ,3 ]
Punset, M. [2 ,3 ]
Giner, Ll [1 ]
Armengou, J. [4 ]
Gil, F. Javier [2 ,3 ]
机构
[1] Univ Int Catalunya, Fac Dent, Barcelona, Spain
[2] Tech Univ Catalonia UPC, Dept Mat Sci & Met Engn, Biomat Biomech & Tissue Engn Grp, ETSEIB, Barcelona 08028, Spain
[3] Biomed Res Networking Ctr Bioengn Biomat & Nanome, Barcelona, Spain
[4] Univ Int Catalunya, Sch Architecture, Barcelona, Spain
关键词
IN-VITRO EVALUATION; OSSEOINTEGRATED IMPLANTS; BACTERIAL-COLONIZATION; TITANIUM IMPLANTS; INTERFACE; INFLAMMATION;
D O I
10.1007/s10856-015-5544-x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study shows the potential risk of microfiltration between two different types of implant-abutment connections screwed at 45 Ncm: external and internal. For the first time the use of a mechanical artificial mouth is used with the values (compression and torsion loads with a frequency of 2 Hz) of the human chewing. The mechanical tests were performed with an artificial saliva at 37 degrees C. The microgap in the connection was measured by an Image Analysis software incorporated in a high resolution scanning electron microscopy. Implant connections were filled with methylene blue by using self-adjustable precision pipettes and the quantity of leakage was determined by high sensitivity spectometry. We showed that the internal connection has lower microgaps compared to the external ones and these microgaps increased with the number of mechanical cycles. The leakage of methylene blue was higher when the external connection was performed. Microgaps and the influence of the mechanical loads are very important for the long-term behavior avoiding the bacteria colonization in the dental implants. These aspects should be known by the implantologists.
引用
收藏
页数:7
相关论文
共 27 条
  • [1] A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW
    ADELL, R
    LEKHOLM, U
    ROCKLER, B
    BRANEMARK, PI
    [J]. INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06): : 387 - 416
  • [2] OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN
    ALBREKTSSON, T
    BRANEMARK, PI
    HANSSON, HA
    LINDSTROM, J
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02): : 155 - 170
  • [3] Microbial leakage through the implant-abutment interface of morse taper implants in vitro
    Aloise, Joao Pedro
    Curcio, Ricardo
    Laporta, Marcia Zorello
    Rossi, Liliane
    Alvares da Silva, Adriana Madeira
    Rapoport, Abrao
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (03) : 328 - 335
  • [4] Bacterial Leakage in Implants With Different Implant-Abutment Connections: An In Vitro Study
    Assenza, Bartolomeo
    Tripodi, Domenico
    Scarano, Antonio
    Perrotti, Vittoria
    Piattelli, Adriano
    Iezzi, Giovanna
    D'Ercole, Simonetta
    [J]. JOURNAL OF PERIODONTOLOGY, 2012, 83 (04) : 491 - 497
  • [5] Bone reactions to longstanding functional load at implants: an experimental study in dogs
    Berglundh, T
    Abrahamsson, I
    Lindhe, J
    [J]. JOURNAL OF CLINICAL PERIODONTOLOGY, 2005, 32 (09) : 925 - 932
  • [6] Binon PP, 1996, INT J PROSTHODONT, V9, P149
  • [7] Peri-implant inflammation defined by the implant-abutment interface
    Broggini, N
    McManus, LM
    Hermann, JS
    Medina, R
    Schenk, RK
    Buser, D
    Cochran, DL
    [J]. JOURNAL OF DENTAL RESEARCH, 2006, 85 (05) : 473 - 478
  • [8] Persistent acute inflammation at the implant-abutment interface
    Broggini, N
    McManus, LM
    Hermann, JS
    Medina, RU
    Oates, TW
    Schenk, RK
    Buser, D
    Mellonig, JT
    Cochran, DL
    [J]. JOURNAL OF DENTAL RESEARCH, 2003, 82 (03) : 232 - 237
  • [9] In vitro evaluation of the implant abutment connection sealing capability of different implant systems
    Coelho, P. G.
    Sudack, P.
    Suzuki, M.
    Kurtz, K. S.
    Romanos, G. E.
    Silva, N. R. F. A.
    [J]. JOURNAL OF ORAL REHABILITATION, 2008, 35 (12): : 917 - 924
  • [10] do Nascimento C, 2012, INT J ORAL MAX IMPL, V27, P551