Vascular access surveillance: an ongoing controversy

被引:80
|
作者
Paulson, William D. [2 ,3 ]
Moist, Louise [4 ]
Lok, Charmaine E. [1 ,5 ]
机构
[1] Toronto Gen Hosp, Dept Med, Div Nephrol, Toronto, ON M5G 2C4, Canada
[2] Georgia Hlth Sci Univ, Dept Med, Charlie Norwood VA Med Ctr, Augusta, GA USA
[3] Georgia Hlth Sci Univ, Dept Med, Nephrol Sect, Augusta, GA USA
[4] Univ Western Ontario, Div Nephrol, London, ON, Canada
[5] Univ Toronto, Toronto, ON, Canada
关键词
arteriovenous fistula; arteriovenous graft; clinical practice guidelines; vascular access; PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY; VENOUS NEOINTIMAL HYPERPLASIA; HEMODIALYSIS GRAFT THROMBOSIS; BLOOD-FLOW; ARTERIOVENOUS-FISTULAS; PHYSICAL-EXAMINATION; LUMINAL DIAMETERS; SYNTHETIC GRAFTS; CONTROLLED-TRIAL; EARLY FAILURE;
D O I
10.1038/ki.2011.337
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Hemodialysis vascular access surveillance continues to be widely recommended despite ongoing controversy as to its benefit in prolonging access patency compared with clinical monitoring alone. The most common screening tests are access blood flow and dialysis venous pressure measurements. When surveillance test results cross a predetermined threshold, accesses are referred for intervention with correction of stenosis to reduce future thrombosis and prolong access survival. Current surveillance strategies have four components: (1) underlying condition; (2) screening test; (3) intervention; and (4) outcomes. However, limitations exist within each component that may prevent achieving the desired outcomes. This review discusses these limitations and their consequences. To date, randomized controlled trials have not consistently shown that surveillance improves outcomes in grafts, and there is limited evidence that surveillance reduces thrombosis without prolonging the life of native fistulae. In conclusion, current evidence does not support the concept that all accesses should undergo routine surveillance with intervention. Kidney International (2012) 81, 132-142; doi:10.1038/ki.2011.337; published online 5 October 2011
引用
收藏
页码:132 / 142
页数:11
相关论文
共 50 条
  • [1] Controversial Vascular Access Surveillance Mandate
    Paulson, William D.
    Work, Jack
    SEMINARS IN DIALYSIS, 2010, 23 (01) : 92 - 94
  • [2] The role of regular surveillance on maintenance of patency of vascular access
    Khairy, Ahmed
    Rashed, Ahmed M.
    Shahat, Mohamed
    Ali, Haitham
    EGYPTIAN JOURNAL OF SURGERY, 2022, 41 (03) : 1002 - 1008
  • [3] Doppler ultrasound and calcification score: improving vascular access surveillance
    Marques, Maria Guedes
    Botelho, Carlos
    Maia, Pedro
    Ibeas, Jose
    Ponce, Pedro
    RENAL FAILURE, 2015, 37 (09) : 1425 - 1429
  • [4] Monitoring and Surveillance of Hemodialysis Access
    Koirala, Nischal
    Anvari, Evamaria
    McLennan, Gordon
    SEMINARS IN INTERVENTIONAL RADIOLOGY, 2016, 33 (01) : 25 - 30
  • [5] Why vascular access trials on flow surveillance failed
    Krivitski, Nikolai
    JOURNAL OF VASCULAR ACCESS, 2014, 15 : S15 - S19
  • [6] Vascular access surveillance: Evaluation of combining dynamic venous pressure and vascular access blood flow measurements
    Hoeben, H
    Abu-Alfa, AK
    Reilly, RF
    Aruny, JE
    Bouman, K
    Perazella, MA
    AMERICAN JOURNAL OF NEPHROLOGY, 2003, 23 (06) : 403 - 408
  • [7] Vascular Access Monitoring and Surveillance: An Update
    Valliant, Amanda
    McComb, Kathryn
    ADVANCES IN CHRONIC KIDNEY DISEASE, 2015, 22 (06) : 446 - 452
  • [8] Imaging in Vascular Access
    Murphy, Eoin A.
    Ross, Rose A.
    Jones, Robert G.
    Gandy, Stephen J.
    Aristokleous, Nicolas
    Salsano, Marco
    Weir-McCall, Jonathan R.
    Matthew, Shona
    Houston, John Graeme
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2017, 8 (03) : 255 - 272
  • [9] Vascular sound visualization system is useful for monitoring and surveillance of vascular access
    Tsuboi, Masato
    Suzuki, Hiroaki
    Kawai, Hirokazu
    Ejima, Toru
    Mitsuishi, Fukumi
    JOURNAL OF VASCULAR ACCESS, 2022, 23 (03) : 390 - 397
  • [10] Haemodialysis vascular access and fistula surveillance methods in The Netherlands
    Ezzahiri, R
    Lemson, MS
    Kitslaar, PJEHM
    Leunissen, KML
    Tordoir, JHM
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 1999, 14 (09) : 2110 - 2115