On the heating of inductively coupled resonators (stents) during MRI examinations

被引:12
作者
Busch, M
Vollmann, W
Bertsch, T
Wetzler, R
Bornstedt, A
Schnackenburg, B
Schnorr, J
Kivelitz, D
Taupitz, M
Grönemeyer, D
机构
[1] Univ Witten Herdecke, Gronemeyer Inst Mikrotherapie, D-44799 Bochum, Germany
[2] Univ Appl Sci, TFH, Berlin, Germany
[3] Deutsch Herzzentrum Berlin, Berlin, Germany
[4] Humboldt Univ, Fak Med, Charite, Inst Radiol, D-1040 Berlin, Germany
[5] Redis GmbH, Bochum, Germany
[6] Med Enzymes AG, Berlin, Germany
关键词
stent; active MRI implants; SAR; temperature effects; MRI artifacts;
D O I
10.1002/mrm.20618
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Stents that have been implanted to preserve the results of vascular dilatation are frequently affected by in-stent restenosis, which ideally should be followed up by a noninvasive diagnostic modality. Active MRI stents can enable this kind of follow-up, while normal metallic stents can not. The prototype stents investigated in this study were designed as electric resonating circuits without a direct connection to the MR imager, and function as inductively coupled transmit coils. The model of a long solenoid coil is used to describe the additional power loss caused by such resonators. The theoretically estimated temperature increase is verified by measurements for different resonators and discussed for worst-case conditions. The RF power absorption of an active resonator is negligible compared to the total power absorbed during MRI. The local temperature increase observed for prototypes embedded in phantoms is in a range that excludes direct tissue damage. However, ruptures in the conducting structure of a resonator can cause hot spots, which may establish a high local temperature. This hazard can be reduced by designing resonators with a low quality (Q) factor or by setting the circuit slightly off resonance; however, this would lower the nominal amplification for which the resonator was designed.
引用
收藏
页码:775 / 782
页数:8
相关论文
共 38 条
[1]  
ATHEY T W, 1982, Magnetic Resonance Imaging, V1, P149, DOI 10.1016/0730-725X(82)90206-5
[2]   Improved lumen visualization in metallic vascular implants by reducing RF artifacts [J].
Bartels, LW ;
Bakker, CJG ;
Viergever, MA .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (01) :171-180
[3]   MR imaging of vascular stents: effects of susceptibility, flow, and radiofrequency eddy currents [J].
Bartels, LW ;
Smits, HFM ;
Bakker, CJG ;
Viergever, MA .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (03) :365-371
[4]   ESTIMATING RADIOFREQUENCY POWER DEPOSITION IN BODY NMR IMAGING [J].
BOTTOMLEY, PA ;
REDINGTON, RW ;
EDELSTEIN, WA ;
SCHENCK, JF .
MAGNETIC RESONANCE IN MEDICINE, 1985, 2 (04) :336-349
[5]  
BUDINGER TF, 1983, CLIN MAGNETIC RESONA, P303
[6]   Artifact-free in-stent lumen visualization by standard magnetic resonance angiography using a new metallic magnetic resonance imaging stent [J].
Buecker, A ;
Spuentrup, E ;
Ruebben, A ;
Günther, RW .
CIRCULATION, 2002, 105 (15) :1772-1775
[7]  
BUETTNER UW, 1992, BIOL EFFECTS SAFETY, P649
[8]  
BUSCH M, 2001, Patent No. 6280385
[9]   Finite volume analysis of temperature effects induced by active MRI implants with cylindrical symmetry: 1. Properly working devices [J].
Busch, Martin H. J. ;
Vollmann, Wolfgang ;
Schnorr, Joerg ;
Groenemeyer, Dietrich H. W. .
BIOMEDICAL ENGINEERING ONLINE, 2005, 4 (1)
[10]  
*CDRH MAGN RES WOR, PRIM MED DEV INT MAG