Dosimetry evaluation of the GammaPod stereotactic radiosurgery device based on established AAPM and IAEA protocols

被引:5
|
作者
Becker, Stewart J. [1 ]
Culberson, Wesley S. [2 ]
Poirier, Yannick [1 ]
Mutaf, Yildirim [1 ]
Niu, Ying [3 ]
Nichols, Elizabeth M. [1 ]
Yi, Byongyong [1 ]
机构
[1] Univ Maryland, Sch Med, Dept Radiat Oncol, Baltimore, MD 21201 USA
[2] Univ Wisconsin, Dept Med Phys, Madison, WI 53705 USA
[3] MedStar Georgetown Univ Hosp, Washington, DC 20007 USA
关键词
breast; cancer; IAEA TRS 483; stereotactic radiosurgery; TG-21; HIGH-ENERGY PHOTON; CYLINDRICAL CHAMBERS; CALIBRATION; PHYSICS;
D O I
10.1002/mp.14197
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The GammaPod is a novel dedicated prone breast stereotactic radiosurgery (SRS) device recently developed at the University of Maryland Medical Center. This device utilizes multiple rotating Co-60 sources to create highly conformal dose distributions for breast treatments, including boosts, partial breast irradiation, or presurgery SRS. However, due to its small field sizes and nonstandard geometry, existing calibration protocols cannot be directly applied. In this study, we adapt and implement the American Association of Physicists in Medicine Task Group 21 (TG-21) and International Atomic Energy Agency (IAEA) Technical Report Series 483 (TRS 483) protocols for reference dose measurements for the GammaPod. This represents the first published dosimetric investigation GammaPod and is meant to serve as a reference to future users commissioning and calibrating these devices. Methods Reference dose measurements were performed following the TG-21/IAEA TRS 483 protocols using an ADCL-calibrated Exradin A1SL thimble chamber in a polymethyl methacrylate (PMMA) breast-mimicking phantom. Monte Carlo calculations and measurements were also performed in water to determine chamber-specifickPMMAQmsr,Q0fmsr,frefquality conversion factor converting reference field size (f(ref)) to machine-specific field sizes (f(msr)) (25-mm) as well askPMMAfclin,fmsr, the conversion factor from the (f(msr)) to the clinical field size (f(clin)) (15mm). Verification was performed using the thermoluminescent dosimeter remote monitoring service from the Imaging and Radiation Oncology Core (IROC) in Houston, TX. Results The (f(ref)) to (f(msr)) chamber-specific factorkPMMAQmsr,Q0fmsr,frefwas 0.992 while the (f(msr)) to (f(clin)) chamber-specifickPMMAfclin,fmsrfactor was 1.014. The radiation absorbed dose to water measured in the PMMA phantom based on the TG-21/IAEA TRS 483 formalism agreed with IROC values to within 1% and 2% for the 25- and 15-mm collimators, respectively. Conclusion We successfully implemented the TG-21 and TRS 483 reference dosimetry protocols for the GammaPod. These results show agreement between measurements performed with different reference dosimetry protocols and independent thermoluminescent measurements.
引用
收藏
页码:3614 / 3620
页数:7
相关论文
共 15 条
  • [1] Recommendations on the practice of calibration, dosimetry, and quality assurance for gamma stereotactic radiosurgery: Report of AAPM Task Group 178
    Petti, Paula L.
    Rivard, Mark J.
    Alvarez, Paola E.
    Bednarz, Greg
    Daniel Bourland, J.
    DeWerd, Larry A.
    Drzymala, Robert E.
    Johansson, Jonas
    Kunugi, Keith
    Ma, Lijun
    Meltsner, Sheridan G.
    Neyman, Gennady
    Seuntjens, Jan P.
    Shiu, Almon S.
    Goetsch, Steven J.
    MEDICAL PHYSICS, 2021, 48 (07) : E733 - E770
  • [2] Dosimetric and geometric evaluation of a novel stereotactic radiotherapy device for breast cancer: The GammaPod™
    Mutaf, Yildirim D.
    Zhang, Jin
    Yu, Cedric X.
    Yi, Byong Yong
    Prado, Karl
    D'Souza, Warren D.
    Regine, William F.
    Feigenberg, Steven J.
    MEDICAL PHYSICS, 2013, 40 (04)
  • [3] A round-robin gamma stereotactic radiosurgery dosimetry interinstitution comparison of calibration protocols
    Drzymala, R. E.
    Alvarez, P. E.
    Bednarz, G.
    Bourland, J. D.
    DeWerd, L. A.
    Ma, L.
    Meltsner, S. G.
    Neyman, G.
    Novotny, J., Jr.
    Petti, P. L.
    Rivard, M. J.
    Shiu, A. S.
    Goetsch, S. J.
    MEDICAL PHYSICS, 2015, 42 (11) : 6745 - 6756
  • [4] A dosimetric evaluation of the IAEA-AAPM TRS483 code of practice for dosimetry of small static fields used in conventional linac beams and comparison with IAEA TRS-398, AAPM TG51, and TG51 Addendum protocols
    Huq, M. Saiful
    Hwang, Min-Sig
    Teo, Troy P.
    Jang, Si Young
    Heron, Dwight E.
    Lalonde, Ronald J.
    MEDICAL PHYSICS, 2018, 45 (09) : 4257 - 4273
  • [5] Technical note: Consistency of IAEA's TRS-483 and AAPM's extended TG-51 protocols for clinical reference dosimetry of the CyberKnife M6 machine
    Duchaine, Jasmine
    Markel, Daniel
    Ley, Jean-Luc
    Beliveau-Nadeau, Dominique
    Zerouali, Karim
    Doucet, Robert
    Bouchard, Hugo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (05):
  • [6] The use of high field strength and parallel imaging techniques for MRI-based gel dosimetry in stereotactic radiosurgery
    Seimenis, I.
    Moutsatsos, A.
    Petrokokkinos, L.
    Kantemiris, I.
    Benekos, O.
    Efstathopoulos, E.
    Papagiannis, P.
    Spevacek, V.
    Semnicka, J.
    Dvorak, P.
    JOURNAL OF INSTRUMENTATION, 2009, 4
  • [7] Evaluation of linear accelerator-based stereotactic radiosurgery in the management of glomus jugulare tumors
    Sager, Omer
    Beyzadeoglu, Murat
    Dincoglan, Ferret
    Gamsiz, Hakan
    Demiral, Selcuk
    Uysal, Bora
    Oysul, Kaan
    Dirican, Bahar
    Sirin, Salt
    TUMORI, 2014, 100 (02) : 184 - 188
  • [8] Evaluation of linear accelerator-based stereotactic radiosurgery in the management of meningiomas: a single center experience
    Dincoglan, F.
    Beyzadeoglu, M.
    Sager, O.
    Uysal, B.
    Demiral, S.
    Gamsiz, H.
    Dirican, B.
    JOURNAL OF BUON, 2013, 18 (03): : 717 - 722
  • [9] Dosimetry study comparing single-isocenter high-definition dynamic radiosurgery with multiple-isocenter cone-based stereotactic radiosurgery in treating multiple brain metastases
    Ji, Tianlong
    Li, Guang
    Song, Yaowen
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2025, 57 (03)
  • [10] Technical Note: Evaluation of the systematic accuracy of a frameless, multiple image modality guided, linear accelerator based stereotactic radiosurgery system
    Wen, N.
    Snyder, K. C.
    Scheib, S. G.
    Schmelzer, P.
    Qin, Y.
    Li, H.
    Siddiqui, M. S.
    Chetty, I. J.
    MEDICAL PHYSICS, 2016, 43 (05) : 2527 - 2537