Corneal Topographic Astigmatism Based on Total Corneal Power Data (CorT Total): A Benchmark for Total Corneal Astigmatism

被引:5
作者
Alpins, Noel [1 ,2 ]
Ong, James K. Y. [3 ]
Randleman, J. Bradley [4 ]
Nevyas-Wallace, Anita [5 ]
Stamatelatos, George [1 ]
机构
[1] NewVis Clin, Melbourne, Vic, Australia
[2] Univ Melbourne, Dept Ophthalmol, Melbourne, Vic, Australia
[3] ASSORT Pty Ltd, Melbourne, Vic, Australia
[4] Cleveland Clin Fdn, Cole Eye Inst, 9500 Euclid Ave, Cleveland, OH 44195 USA
[5] Nevyas Eye Associates, Delaware Valley Laser Surg Inst, Philadelphia, PA USA
关键词
corneal topographic astigmatism (CorT); corneal tomography; total corneal power; IN-SITU KERATOMILEUSIS; LIGHT-REFLEX; REFRACTIVE ASTIGMATISM; ORBSCAN-II; SURGERY; CENTRATION; AGREEMENT; ABLATION; OUTCOMES; VERTEX;
D O I
10.1097/ICO.0000000000002162
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate how closely manufacturer-provided measures of total corneal astigmatism correspond with the manifest refractive cylinder, as compared to a benchmark of corneal topographic astigmatism calculated on the basis of measured total corneal power (TCP) data (CorT Total). Methods: The SD of the ocular residual astigmatism magnitude (ORAsd) was evaluated for normal virgin eyes based on an optimized benchmark CorT Total and the various measures of total corneal astigmatism provided by 3 different Scheimpflug tomographers. Results: The CorT Total corresponded with the manifest refractive cylinder at least as well as all the measures of total corneal astigmatism provided by the tomographers [Sirius CorT Total ORAsd: 0.320D (standard error [SE] 0.017D), Sirius TCP 4 mm ORAsd: 0.324D (SE 0.017D); Pentacam CorT Total ORAsd: 0.338D (SE 0.027D), Pentacam total corneal refractive power apex zone 4 mm ORAsd: 0.337D (SE 0.029D); Galilei CorT Total ORAsd: 0.472D (SE 0.068D), and Galilei TCP2 ORAsd: 0.536D (SE 0.124D)]. The difference between CorT Total and best measure on each tomographer was not statistically significant (Sirius TCP 4 mm: P = 0.24, Pentacam total corneal refractive power apex zone 4 mm: P = 0.64, Galilei TCP2: P = 0.24). Most of the manufacturer-provided measures did not correspond closely with the manifest refractive cylinder. When there were multiple measures of total corneal astigmatism, those derived from a zone with a diameter of 4.0 mm corresponded best with the manifest refractive cylinder. Conclusions: The CorT Total is a reliable benchmark measure that can be used to assess how well other measures of total corneal astigmatism correspond with the manifest refractive cylinder.
引用
收藏
页码:431 / 436
页数:6
相关论文
共 50 条
  • [21] Influence of corneal astigmatism, corneal curvature and meridional differences on corneal hysteresis and corneal resistance factor
    Wong, Yin-zhi
    Lam, Andrew K. C.
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2011, 94 (05) : 418 - 424
  • [22] The Influence of Corneal Thickness on Surgically Induced Corneal Astigmatism Derived from Total Keratometry Measured by Anterior Segment Swept-Source OCT
    Zhao, Fangyu
    Yin, Yufan
    Pazo, Emmanuel Eric
    Tian, Fang
    Jiang, Yuanfeng
    Bu, Shaochong
    OPHTHALMOLOGY AND THERAPY, 2024, 13 (09) : 2381 - 2391
  • [23] Role of Hemidivisional Corneal Topographic Astigmatisms (CorTs) in the Regularization and Reduction of Irregular Astigmatism
    Alpins, Noel
    Ong, James K. Y.
    Stamatelatos, George
    CORNEA, 2018, 37 (03) : 386 - 393
  • [24] Correlation Between Anterior Corneal Elevation Differences in Main Meridians and Corneal Astigmatism
    Batres, Laura
    Pinero, David
    Carracedo, Gonzalo
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2020, 46 (02): : 99 - 104
  • [25] The Impact of Pterygium Excision on Corneal Astigmatism
    Khan, Faisal Aziz
    Niazi, Shafaq Pervez Khan
    Khan, Dilshad Alam
    JCPSP-JOURNAL OF THE COLLEGE OF PHYSICIANS AND SURGEONS PAKISTAN, 2014, 24 (06): : 404 - 407
  • [26] Anterior and Posterior Corneal Astigmatism after Refractive Lenticule Extraction for Myopic Astigmatism
    Kamiya, Kazutaka
    Shimizu, Kimiya
    Yamagishi, Mayumi
    Igarashi, Akihito
    Kobashi, Hidenaga
    JOURNAL OF OPHTHALMOLOGY, 2015, 2015
  • [27] Prediction of total corneal power in keratoconus using anterior surface data
    Camellin, Umberto
    Ninotta, Ivan
    Latino, Gianluigi
    Alibrandi, Angela
    Aragona, Pasquale
    Roszkowska, Anna M. M.
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2024, 107 (03) : 274 - 280
  • [28] Accuracy of Corneal Astigmatism Estimation by Neglecting the Posterior Corneal Surface Measurement
    Ho, Jau-Der
    Tsai, Ching-Yao
    Liou, Shiow-Wen
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (05) : 788 - 795
  • [29] Assessment of total corneal power after myopic corneal refractive surgery in Chinese eyes
    Hua, Yanjun
    Pan, Chao
    Wang, Qinmei
    INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (11) : 2467 - 2475
  • [30] Comparison of Corneal Power and Corneal Astigmatism of Different Diameter Zones Centered on the Pupil and Corneal Apex Using Scheimpflug Tomography
    Dong, Jing
    Zhang, Yaqin
    Zhou, Jing
    Zhang, Suhua
    Sun, Bin
    Tang, Maolong
    Wang, Xiaogang
    CORNEA, 2020, 39 (01) : 77 - 83