Comparison of small field of view gamma camera systems for scintimammography

被引:32
作者
Hruska, CB [1 ]
O'Connor, MK [1 ]
Collins, DA [1 ]
机构
[1] Mayo Clin, Nucl Med Sect, Dept Radiol, Rochester, MN 55905 USA
关键词
breast; compact gamma camera; scintimammography;
D O I
10.1097/00006231-200505000-00008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective To compare the performance characteristics of three compact gamma camera systems for scintimammography. Methods A multi-crystal Nal system (Dilon 6800), a multicrystal CsI system (Digirad 2020tc Imager) and a prototype cadmium zinc telluride (CZT) detector were evaluated using a breast phantom with simulated tumours of 1.8-9.8 mm in diameter and a tumour to background ratio of 5.6: 1. A conventional Anger-based gamma camera was also evaluated for comparative purposes. Results The conventional gamma camera had the highest intrinsic sensitivity of the four systems and utilized the highest resolution collimator. However, the prototype CZT and the CsI systems yielded the best spatial resolution over the range 0-5cm from the collimator face and better visualized small (5-7 mm) tumours. Conclusion In addition to the advantage of close proximity to the breast, two of the three compact detectors offered superior performance to a conventional gamma camera for scintimammography. (c) 2005 Lippincott Williams & Wilkins.
引用
收藏
页码:441 / 445
页数:5
相关论文
共 15 条
[1]   Evaluation of a high-resolution, breast-specific, small-field-of view gamma camera for the detection of breast cancer [J].
Brem, RF ;
Kieper, DA ;
Rapelyea, JA ;
Majewski, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 497 (01) :39-45
[2]  
Brem RF, 2002, J NUCL MED, V43, P909
[3]   A discrete scintillation camera module using silicon photodiode readout of CsI(Tl) crystals for breast cancer imaging [J].
Gruber, GJ ;
Moses, WW ;
Derenzo, SE ;
Wang, NW ;
Beuville, E ;
Ho, MH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :1063-1068
[4]   Development of a miniature scintillation camera using an NaI(Tl) scintillator and PSPMT for scintimammography [J].
Kim, JH ;
Choi, Y ;
Joo, KS ;
Sihn, BS ;
Chong, JW ;
Kim, SE ;
Lee, KH ;
Choe, YS ;
Kim, BT .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3481-3488
[5]   Optimization of dedicated scintimammography procedure using detector prototypes and compressible phantoms [J].
Majewski, S ;
Kieper, D ;
Curran, E ;
Keppel, C ;
Kross, B ;
Palumbo, A ;
Popov, V ;
Weisenberger, AG ;
Welch, B ;
Wojcik, R ;
Williams, MB ;
Goode, AR ;
More, M ;
Zhang, G .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) :822-829
[6]  
Maublant J, 1996, J NUCL MED, V37, P922
[7]   Evaluation of breast tumor detectability with two dedicated, compact scintillation cameras [J].
McElroy, DP ;
Hoffman, EJ ;
MacDonald, L ;
Patt, BE ;
Iwanczyk, JS ;
Yamaguchi, Y ;
Levin, CS .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (03) :794-802
[8]  
Mueller B, 2003, J NUCL MED, V44, P602
[9]  
*NAT EL MAN ASS, 1994, PERF MEAS SCINT CAM
[10]   Multi-PSPMT scintillation camera [J].
Pani, R ;
Soluri, A ;
Scafe, R ;
Pergola, A ;
Pellegrini, R ;
De Vincentis, G ;
Trotta, G ;
Scopinaro, F .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (03) :702-708