Analysis of ecstasy tablets: comparison of reflectance and transmittance near infrared spectroscopy

被引:54
作者
Schneider, RC [1 ]
Kovar, KA [1 ]
机构
[1] Univ Tubingen, Dept Pharmaceut Anal, Inst Pharmaceut, D-72076 Tubingen, Germany
关键词
forensic science; NIR; MDMA; MDE; amphetamine; ecstasy;
D O I
10.1016/S0379-0738(03)00125-7
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
Calibration models for the quantitation of commonly used ecstasy substances have been developed using near infrared spectroscopy (NIR) in diffuse reflectance and in transmission mode by applying seized ecstasy tablets for model building and validation. The samples contained amphetamine, N-methyl-3,4-methylenedioxy-amphetamine (MDMA) and N-ethyl-3,4-methylenedioxy-amphetamine (MDE) in different concentrations. All tablets were analyzed using high performance liquid chromatography (HPLC) with diode array detection as reference method. We evaluated the performance of each NIR measurement method with regard to its ability to predict the content of each tablet with a low root mean square error of prediction (RMSEP). Best calibration models could be generated by using NIR measurement in transmittance mode with wavelength selection and 1/x-transformation of the raw data. The models build in reflectance mode showed higher RMSEPs using as data pretreatment, wavelength selection, 1/x-transformation and a second order Savitzky-Golay derivative with five point smoothing was applied to obtain the best models. To estimate the influence of inhomogeneities in the illegal tablets, a calibration of the destroyed, i.e. triturated samples was build and compared to the corresponding data of the whole tablets. The calibrations using these homogenized tablets showed lower RMSEPs. We can conclude that NIR analysis of ecstasy tablets in transmission mode is more suitable than measurement in diffuse reflectance to obtain quantification models for their active ingredients with regard to low errors of prediction. Inhomogeneities in the samples are equalized when measuring the tablets as powdered samples. (C) 2003 Published by Elsevier Science Ireland Ltd.
引用
收藏
页码:187 / 195
页数:9
相关论文
共 18 条
[1]   Drawbacks of the present standards for processing absorption spectra recorded linearly as a function of wavelength [J].
Antonov, L .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1997, 16 (09) :536-543
[2]   REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DATABASE OF RETENTION INDEXES AND UV SPECTRA OF TOXICOLOGICALLY RELEVANT SUBSTANCES AND ITS INTERLABORATORY USE [J].
BOGUSZ, M ;
ERKENS, M .
JOURNAL OF CHROMATOGRAPHY A, 1994, 674 (1-2) :97-126
[3]  
*BUND, 2000, RAUSCHG
[4]  
Chan K.B., 2001, MICROGRAM, V34, P237
[5]   Toxicological detection of the designer drug 3,4-methylenedioxyethylamphetamine (MDE, ''Eve'') and its metabolites in urine by gas chromatography mass spectrometry and fluorescence polarization immunoassay [J].
Ensslin, HK ;
Kovar, KA ;
Maurer, HH .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 1996, 683 (02) :189-197
[6]   Analysis of ''Ecstasy'' by capillary electrophoresis [J].
Frost, M ;
Kohler, H ;
Blaschke, G .
INTERNATIONAL JOURNAL OF LEGAL MEDICINE, 1996, 109 (02) :53-57
[7]   Vibrational spectrometry for the assessment of active substance in metoprolol tablets: A comparison between transmission and diffuse reflectance near-infrared spectrometry [J].
Gottfries, J ;
Depui, H ;
Fransson, M ;
Jongeneelen, M ;
Josefson, M ;
Langkilde, FW ;
Witte, DT .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1996, 14 (11) :1495-1503
[8]   Information depth of Vis-NIR light in polyethylene films using transmission and reflectance measurements [J].
Haanstra, WG ;
Hansen, W ;
Huys, MJG ;
Kip, BJ ;
Palmen, P ;
Roumen, J ;
Snieder, M ;
van de Weerdhof, T ;
Wiedemann, S ;
Wortel, VAL .
APPLIED SPECTROSCOPY, 1998, 52 (06) :863-868
[9]   Solid-phase microextraction/capillary gas chromatography for the profiling of confiscated ecstacy and amphetamine [J].
Kongshaug, KE ;
Pedersen-Bjergaard, S ;
Rasmussen, KE ;
Krogh, M .
CHROMATOGRAPHIA, 1999, 50 (3-4) :247-252
[10]  
MARTENS H, 1991, PARTIAL LEAST SQUARE