High-Pitch 128-Slice Dual-Source CT for the Assessment of Coronary Stents in a Phantom Model

被引:16
|
作者
Donati, Olivio F. [1 ]
Burg, Matthias C. [2 ]
Desbiolles, Lotus [1 ]
Karlo, Christoph [1 ]
Stolzmann, Paul [1 ,3 ]
Bunck, Alexander [2 ]
Baumueller, Stephan [1 ]
Marincek, Borut [1 ]
Seifarth, Harald [2 ]
Alkadhi, Hatem [1 ,3 ]
Leschka, Sebastian [1 ]
Maintz, David [2 ]
机构
[1] Univ Zurich Hosp, Inst Diagnost Radiol, CH-8091 Zurich, Switzerland
[2] Univ Munster, Dept Clin Radiol, D-4400 Munster, Germany
[3] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Cardiac MR PET CT Program, Boston, MA USA
关键词
Stent; computed tomography; high pitch; coronary artery disease; dual source; PACLITAXEL-ELUTING STENT; IN-VITRO EVALUATION; BARE-METAL STENT; COMPUTED-TOMOGRAPHY; ANGIOGRAPHY; PATENCY; RESTENOSIS;
D O I
10.1016/j.acra.2010.06.017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: To evaluate the quality of stent lumen delineation using dual-source computed tomography (DSCT) in the standard-pitch mode (SP) as compared to the high-pitch mode (HP) in a phantom study. Materials and Methods: Forty different coronary stents placed in plastic tubes filled with contrast agent were imaged with a second generation DSCT system in a SP (pitch 0.23) and HP (pitch 3.4) mode in orientations of 0, 450, and 90 relative to the z-axis. Two observers independently measured the in-stent lumen and the attenuation values in the center of the stents. The artificial lumen narrowing (ALN) was calculated using the measured in-stent lumen and the nominal diameter of the plastic tube. Results: lnterobserver correlation was excellent for in-stent lumen (0.86) and attenuation measurements (0.91). There was no significant difference neither for ALN (SP: 54.7-62.8%; HP: 55.8-64.0%) nor attenuation (SP: 356-395 Hounsfield units [HU]; HP: 352-384 HU) between SP and HP mode. For both modes, the orientation of the stent relative to the z-axis significantly affected ALN and attenuation (each P < .001). CT volume dose index was significantly lower using HP mode as compared to SP mode (P < .001). Conclusion: The HP mode in DSCT provides visualization of the coronary in-stent lumen comparable to that measured in SP mode while reducing applied radiation dose in a stationary phantom model.
引用
收藏
页码:1366 / 1374
页数:9
相关论文
共 50 条
  • [1] Accuracy of 128-slice dual-source CT using high-pitch spiral mode for the assessment of coronary stents: First in vivo experience
    Yang Xia
    Yang Junjie
    Zhou Ying
    He Bai
    Wang Qi
    Jin Qinhua
    Chen Yundai
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (04) : 617 - 622
  • [2] Coronary artery stent imaging with 128-slice dual-source CT using high-pitch spiral acquisition in a cardiac phantom: comparison with the sequential and low-pitch spiral mode
    Wolf, Florian
    Leschka, Sebastian
    Loewe, Christian
    Homolka, Peter
    Plank, Christina
    Schernthaner, Ruediger
    Bercaczy, Dominik
    Goetti, Robert
    Lammer, Johannes
    Friedrich, Guy
    Marincek, Borut
    Alkadhi, Hatem
    Feuchtner, Gudrun
    EUROPEAN RADIOLOGY, 2010, 20 (09) : 2084 - 2091
  • [3] Coronary artery stent imaging with 128-slice dual-source CT using high-pitch spiral acquisition in a cardiac phantom: comparison with the sequential and low-pitch spiral mode
    Florian Wolf
    Sebastian Leschka
    Christian Loewe
    Peter Homolka
    Christina Plank
    Ruediger Schernthaner
    Dominik Bercaczy
    Robert Goetti
    Johannes Lammer
    Guy Friedrich
    Borut Marincek
    Hatem Alkadhi
    Gudrun Feuchtner
    European Radiology, 2010, 20 : 2084 - 2091
  • [4] Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode
    Alkadhi, Hatem
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Plass, Andre
    Scheffel, Hans
    Feuchtner, Gudrun
    Falk, Volkmar
    Marincek, Borut
    Leschka, Sebastian
    HEART, 2010, 96 (12) : 933 - 938
  • [5] Assessment of the Coronary Artery in High-Pitch, Dual-Source CT Aortography Without ECG Synchronization
    Choi, Hyeong Gi
    Park, Mi Jung
    Choi, Ho Cheol
    Choi, Hye Young
    Shin, Hwa Seon
    Na, Jae Boem
    Cho, Jae Min
    Choi, Dae Seob
    IRANIAN JOURNAL OF RADIOLOGY, 2017, 14 (03)
  • [6] Diagnostic accuracy of 128-slice dual-source CT coronary angiography: a randomized comparison of different acquisition protocols
    Neefjes, Lisan A.
    Rossi, Alexia
    Genders, Tessa S. S.
    Nieman, Koen
    Papadopoulou, Stella L.
    Dharampal, Anoeshka S.
    Schultz, Carl J.
    Weustink, Annick C.
    Dijkshoorn, Marcel L.
    ten Kate, Gert-Jan R.
    Dedic, Admir
    van Straten, Marcel
    Cademartiri, Filippo
    Hunink, M. G. Myriam
    Krestin, Gabriel P.
    de Feyter, Pim J.
    Mollet, Nico R.
    EUROPEAN RADIOLOGY, 2013, 23 (03) : 614 - 622
  • [7] Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience
    Leschka, Sebastian
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Schertler, Thomas
    Scheffel, Hans
    Plass, Andre
    Falk, Volkmar
    Feuchtner, Gudrun
    Marincek, Borut
    Alkadhi, Hatem
    EUROPEAN RADIOLOGY, 2009, 19 (12) : 2896 - 2903
  • [8] Quantification and Limitation of CT Coronary Angiography Using 128-Slice Dual-Source CT at High Heart Rates: A Simulation Study
    Zhao, Lei
    Liu, Aishi
    Guo, Youmin
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2018, 8 (06) : 1201 - 1206
  • [9] Comparison of image quality and radiation dose of different pulmonary CTA protocols on a 128-slice CT: high-pitch dual source CT, dual energy CT and conventional spiral CT
    De Zordo, Tobias
    von Lutterotti, Klemens
    Dejaco, Christian
    Soegner, Peter F.
    Frank, Renate
    Aigner, Friedrich
    Klauser, Andrea S.
    Pechlaner, Christoph
    Schoepf, U. Joseph
    Jaschke, Werner R.
    Feuchtner, Gudrun M.
    EUROPEAN RADIOLOGY, 2012, 22 (02) : 279 - 286
  • [10] Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease
    Nie, Pei
    Wang, Ximing
    Cheng, Zhaoping
    Ji, Xiaopeng
    Duan, Yanhua
    Chen, Jiuhong
    EUROPEAN RADIOLOGY, 2012, 22 (10) : 2057 - 2066