Determination of optimal electrode positions for transcranial direct current stimulation (tDCS)

被引:65
作者
Im, Chang-Hwan [1 ]
Jung, Hui-Hun [1 ]
Choi, Jung-Do [1 ]
Lee, Soo Yeol [2 ]
Jung, Ki-Young [3 ]
机构
[1] Yonsei Univ, Dept Biomed Engn, Wonju 220710, South Korea
[2] Kyung Hee Univ, Dept Biomed Engn, Suwon, South Korea
[3] Korea Univ, Coll Med, Med Ctr, Seoul 136705, South Korea
关键词
D O I
10.1088/0031-9155/53/11/N03
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The present study introduces a new approach to determining optimal electrode positions in transcranial direct current stimulation (tDCS). Electric field and 3D conduction current density were analyzed using 3D finite element method (FEM) formulated for a dc conduction problem. The electrode positions for minimal current injection were optimized by changing the Cartesian coordinate system into the spherical coordinate system and applying the (2+6) evolution strategy (ES) algorithm. Preliminary simulation studies applied to a standard three-layer head model demonstrated that the proposed approach is promising in enhancing the performance of tDCS.
引用
收藏
页码:N219 / N225
页数:7
相关论文
共 15 条
[1]   A controlled clinical trial of cathodal DC polarization in patients with refractory epilepsy [J].
Fregni, F ;
Thome-Souza, S ;
Nitsche, MA ;
Freedman, SD ;
Valente, KD ;
Pascual-Leone, A .
EPILEPSIA, 2006, 47 (02) :335-342
[2]   Transcranial direct current stimulation [J].
Fregni, F ;
Boggio, PS ;
Nitsche, PS ;
Pascual-Leone, A .
BRITISH JOURNAL OF PSYCHIATRY, 2005, 186 :446-447
[3]   Technology Insight: noninvasive brain stimulation in neurology - perspectives on the therapeutic potential of rTMS and tDCS [J].
Fregni, Felipe ;
Pascual-Leone, Alvaro .
NATURE CLINICAL PRACTICE NEUROLOGY, 2007, 3 (07) :383-393
[4]   Influence of tissue resistivities on neuromagnetic fields and electric potentials studied with a finite element model of the head [J].
Haueisen, J ;
Ramon, C ;
Eiselt, M ;
Brauer, H ;
Nowak, H .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997, 44 (08) :727-735
[5]   Predicted current densities in the brain during transcranial electrical stimulation [J].
Holdefer, RN ;
Sadleir, R ;
Russell, MJ .
CLINICAL NEUROPHYSIOLOGY, 2006, 117 (06) :1388-1397
[6]   Fast and robust localization of brain electrical sources using evolution strategies: Monte-carlo simulation and phantom experiment studies [J].
Im, CH ;
Jung, HK ;
Han, JY ;
Lee, HR ;
Lee, SY .
INTERNATIONAL JOURNAL OF APPLIED ELECTROMAGNETICS AND MECHANICS, 2004, 20 (3-4) :197-203
[7]   Computer-aided performance evaluation of a multichannel transcranial magnetic stimulation system [J].
Im, Chang-Hwan ;
Lee, Chany .
IEEE TRANSACTIONS ON MAGNETICS, 2006, 42 (12) :3803-3808
[8]  
Jin J., 2002, The Finite Element Method in Electromagnetics, V2
[9]   Monte Carlo simulation studies of EEG and MEG localization accuracy [J].
Liu, AK ;
Dale, AM ;
Belliveau, JW .
HUMAN BRAIN MAPPING, 2002, 16 (01) :47-62
[10]   Transcutaneous cranial electrical stimulation (Limoge's currents) decreases early buprenorphine analgesic requirements after abdominal surgery [J].
Mignon, A ;
Laudenbach, V ;
Guischard, F ;
Limoge, A ;
Desmonts, JM ;
Mantz, J .
ANESTHESIA AND ANALGESIA, 1996, 83 (04) :771-775