Design and Hemocompatibility Analysis of a Double-Suction Injection Suspension Blood Pump Using Computational Fluid Dynamics Methods

被引:6
作者
Wu, Yue [1 ]
Zhu, Liangfan [1 ]
Luo, Yun [1 ]
机构
[1] Shanghai Jiao Tong Univ, Inst Biomed Mfg & Life Qual Engn, Sch Mech Engn, 800 Dongchuan Rd, Shanghai 200240, Peoples R China
关键词
Blood pump; Double suction; Injection suspension; Hemocompatibility; Computational fluid dynamics; Left ventricular assist device; VENTRICULAR ASSIST DEVICES; DAMAGE PREDICTION; HEMOLYSIS; MODEL; IMPELLER; BEARINGS;
D O I
10.1111/aor.12888
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The blood pump has become a possible solution to heart diseases. For the prevention of device failure and hemocompatibility problems, a rotary pump with suspended bearing is a preferred solution. In our previous work, a novel injection suspension method has been introduced to levitate the rotor. The suspension method is totally passive. This study aims to apply this suspension method to a double-suction pump, and the property of the pump was investigated using computational fluid dynamics (CFD) methods. The flow field of the pump is simulated based on the SST k- turbulent model. The characteristic curves of the pump were calculated. At the nominal working point of 5 L/min, 100 mm Hg, the suspension force acting on the rotor was detected, which could reach 0.46 N with a gap of 150 mu m. We compared the pump with a previously developed single-suction injection pump to evaluate the blood compatibility of the double-suction design. The average scalar shear stress values were 3.13 Pa for the double-suction pump and 7.10 Pa for the single-suction pump. Larger volumes in the single-suction pump were exposed to shear stresses higher than 10 Pa. Thresholds for the von Willebrand factor cleavage, platelet activation, and hemolysis were defined to be 9 Pa, 50 Pa, and 150 Pa, respectively. The volume fractions for the double-suction pump are lower for all thresholds. The normalized index of hemolysis (NIH) values for the two pumps were calculated to be 0.008 g/100 L and 0.016 g/100 L. Results proved that the double-suction pump has a better hemocompatibility compared with the single-suction pump.
引用
收藏
页码:979 / 987
页数:9
相关论文
共 31 条
  • [11] HEUSER G, 1980, BIORHEOLOGY, V17, P17
  • [12] Third-generation blood pumps with mechanical noncontact magnetic bearings
    Hoshi, H
    Shinshi, T
    Takatani, S
    [J]. ARTIFICIAL ORGANS, 2006, 30 (05) : 324 - 338
  • [13] Karassik I., 2012, Centrifugal pumps
  • [14] Geometric Optimization of a Step Bearing for a Hydrodynamically Levitated Centrifugal Blood Pump for the Reduction of Hemolysis
    Kosaka, Ryo
    Yada, Toru
    Nishida, Masahiro
    Maruyama, Osamu
    Yamane, Takashi
    [J]. ARTIFICIAL ORGANS, 2013, 37 (09) : 778 - 785
  • [15] Kumar A, 2011, COMPREHENSIVE BIOMAT, P389
  • [16] Lobanoff V.S., 2013, CENTRIFUGAL PUMPS DE
  • [17] Magnetically suspended centrifugal blood pump with a self bearing motor
    Masuzawa, T
    Onuma, H
    Kim, SJ
    Okada, Y
    [J]. ASAIO JOURNAL, 2002, 48 (04) : 437 - 442
  • [18] Effect of Impeller Geometry on Lift-Off Characteristics and Rotational Attitude in a Monopivot Centrifugal Blood Pump
    Nishida, Masahiro
    Nakayama, Kento
    Sakota, Daisuke
    Kosaka, Ryo
    Maruyama, Osamu
    Kawaguchi, Yasuo
    Kuwana, Katsuyuki
    Yamane, Takashi
    [J]. ARTIFICIAL ORGANS, 2016, 40 (06) : E89 - E101
  • [19] The Effects of Ambulatory Accelerations on the Stability of a Magnetically Suspended Impeller for an Implantable Blood Pump
    Paul, Gordon
    Rezaienia, Mohammed Amin
    Rahideh, Akbar
    Munjiza, Ante
    Korakianitis, Theodosios
    [J]. ARTIFICIAL ORGANS, 2016, 40 (09) : 867 - 876
  • [20] Computational fluid dynamics analysis of hydrodynamic bearings of the VentrAssist rotary blood pump
    Qian, Y
    Bertram, CD
    [J]. ARTIFICIAL ORGANS, 2000, 24 (06) : 488 - 491