Design and Hemocompatibility Analysis of a Double-Suction Injection Suspension Blood Pump Using Computational Fluid Dynamics Methods

被引:6
作者
Wu, Yue [1 ]
Zhu, Liangfan [1 ]
Luo, Yun [1 ]
机构
[1] Shanghai Jiao Tong Univ, Inst Biomed Mfg & Life Qual Engn, Sch Mech Engn, 800 Dongchuan Rd, Shanghai 200240, Peoples R China
关键词
Blood pump; Double suction; Injection suspension; Hemocompatibility; Computational fluid dynamics; Left ventricular assist device; VENTRICULAR ASSIST DEVICES; DAMAGE PREDICTION; HEMOLYSIS; MODEL; IMPELLER; BEARINGS;
D O I
10.1111/aor.12888
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The blood pump has become a possible solution to heart diseases. For the prevention of device failure and hemocompatibility problems, a rotary pump with suspended bearing is a preferred solution. In our previous work, a novel injection suspension method has been introduced to levitate the rotor. The suspension method is totally passive. This study aims to apply this suspension method to a double-suction pump, and the property of the pump was investigated using computational fluid dynamics (CFD) methods. The flow field of the pump is simulated based on the SST k- turbulent model. The characteristic curves of the pump were calculated. At the nominal working point of 5 L/min, 100 mm Hg, the suspension force acting on the rotor was detected, which could reach 0.46 N with a gap of 150 mu m. We compared the pump with a previously developed single-suction injection pump to evaluate the blood compatibility of the double-suction design. The average scalar shear stress values were 3.13 Pa for the double-suction pump and 7.10 Pa for the single-suction pump. Larger volumes in the single-suction pump were exposed to shear stresses higher than 10 Pa. Thresholds for the von Willebrand factor cleavage, platelet activation, and hemolysis were defined to be 9 Pa, 50 Pa, and 150 Pa, respectively. The volume fractions for the double-suction pump are lower for all thresholds. The normalized index of hemolysis (NIH) values for the two pumps were calculated to be 0.008 g/100 L and 0.016 g/100 L. Results proved that the double-suction pump has a better hemocompatibility compared with the single-suction pump.
引用
收藏
页码:979 / 987
页数:9
相关论文
共 31 条
  • [1] Ariff M, 2009, 7 INT C CFD MIN PROC
  • [2] Dynamic characteristics of a magnetically levitated impeller in a centrifugal blood pump
    Asama, Junichi
    Shinshi, Tadahiko
    Hoshi, Hideo
    Takatani, Setsuo
    Shimokohbe, Akira
    [J]. ARTIFICIAL ORGANS, 2007, 31 (04) : 301 - 311
  • [3] 3-DIMENSIONAL NUMERICAL PREDICTION OF STRESS LOADING OF BLOOD PARTICLES IN A CENTRIFUGAL PUMP
    BLUDSZUWEIT, C
    [J]. ARTIFICIAL ORGANS, 1995, 19 (07) : 590 - 596
  • [4] MODEL FOR A GENERAL MECHANICAL BLOOD DAMAGE PREDICTION
    BLUDSZUWEIT, C
    [J]. ARTIFICIAL ORGANS, 1995, 19 (07) : 583 - 589
  • [5] A CFD model for the prediction of haemolysis in micro axial left ventricular assist devices
    Carswell, D.
    McBride, D.
    Croft, T. N.
    Slone, A. K.
    Cross, M.
    Foster, G.
    [J]. APPLIED MATHEMATICAL MODELLING, 2013, 37 (06) : 4199 - 4207
  • [6] Acquired von Willebrand Syndrome in Continuous-Flow Ventricular Assist Device Recipients
    Crow, Sheri
    Chen, Dong
    Milano, Carmelo
    Thomas, William
    Joyce, Lyle
    Piacentino, Valentino, III
    Sharma, Riti
    Wu, Jogin
    Arepally, Gowthami
    Bowles, Dawn
    Rogers, Joseph
    Villamizar-Ortiz, Nestor
    [J]. ANNALS OF THORACIC SURGERY, 2010, 90 (04) : 1263 - 1269
  • [7] A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis Index
    Fraser, Katharine H.
    Zhang, Tao
    Taskin, M. Ertan
    Griffith, Bartley P.
    Wu, Zhongjun J.
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2012, 134 (08):
  • [8] Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease
    Geisen, Ulrich
    Heilmann, Claudia
    Beyersdorf, Friedhelm
    Benk, Christoph
    Berchtold-Herz, Michael
    Schlensak, Christian
    Budde, Ulrich
    Zieger, Barbara
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2008, 33 (04) : 679 - 684
  • [9] ESTIMATION OF SHEAR STRESS-RELATED BLOOD DAMAGE IN HEART-VALVE PROSTHESES - INVITRO COMPARISON OF 25 AORTIC VALVES
    GIERSIEPEN, M
    WURZINGER, LJ
    OPITZ, R
    REUL, H
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1990, 13 (05) : 300 - 306
  • [10] The power-law mathematical model for blood damage prediction: Analytical developments and physical inconsistencies
    Grigioni, M
    Daniele, C
    Morbiducci, U
    D'Avenio, G
    Di Benedetto, G
    Barbaro, V
    [J]. ARTIFICIAL ORGANS, 2004, 28 (05) : 467 - 475