Using of high precision multimeter digitizer for high transient current calibration

被引:0
作者
Roccato, P. E. [1 ]
Capra, P. P. [1 ]
Bellavia, L. [1 ]
Cinnirella, L. [1 ]
Lanzillotti, M. [1 ]
机构
[1] Ist Nazl Ric Metrol INRiM, Str Cacce 91, I-10135 Turin, Italy
来源
2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018) | 2018年
关键词
Short-circuit currents; fault currents; measurement techniques; measurement uncertainty; current measurements; uncertainty; PERFORMANCES;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The calibration of transient current sensors is fundamental for the reliability of high current, short-circuit tests. The test can be performed with prospective or withstand current, but in both cases, the dynamic behavior of the current sensors has to be investigated, due to his influence on the scale factor value. The requirements of the IEC 62475 standard can be achieved with a calibration in transient conditions, performed with an approved reference measuring system. In this work a most performance measuring system is presented, for the improvement of INRIM high transient current measuring capability.
引用
收藏
页数:2
相关论文
共 5 条
  • [1] Agilent Technologies, 2014, 3458A MULT US GUID
  • [2] Evaluation of Flexible Rogowski Coil Performances in Power Frequency Applications
    Chiampi, Mario
    Crotti, Gabriella
    Morando, Andrea
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (03) : 854 - 862
  • [3] Improvement of Agilent 3458A Performances in Wideband Complex Transfer Function Measurement
    Crotti, Gabriella
    Giordano, Domenico
    Luiso, Mario
    Pescetto, Paolo
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (06) : 1108 - 1116
  • [4] IEC, 2010, 62475 IEC
  • [5] Swerlein L., 1991, 10PPM ACCURAGE DIGIT