Preliminary surgical result of cervical spine reconstruction with a dynamic plate and titanium mesh cage

被引:0
作者
Chung, Dae-Yeong [1 ]
Cho, Dae-Chul [1 ]
Lee, Sun-Ho [1 ]
Sung, Joo-Kyung [1 ]
机构
[1] Kyungpook Natl Univ, Dept Neurosurg, Sch Med, Taegu 700721, South Korea
关键词
cervical corpectomy; dynamic plate; titanium mesh cage;
D O I
暂无
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective : The objective of this study was to validate the effects of a titanium mesh cage and dynamic plating in anterior cervical stabilization after corpectomy. Methods : A retrospective study was performed on 31 consecutive patients, who underwent anterior cervical reconstruction with a titanium mesh cage and dynamic plating, from March 2004 to February 2006. Twenty-four patients had 1-level and 7 had 2-level corpectomies. Ten patients underwent surgery with a cage of 10-mm diameter and 21 with 13-mm diameter. Neurological status and outcomes were assessed according to Odom's criteria. Sagittal angle, coronal angle, settling ratio, sagittal displacement, and cervical lordosis were used to evaluate the radiological outcomes. Results : In overall, 26 (83.9%) of 31 showed excellent or good outcomes. Thirteen percent (4 cases) of the patients developed surgical complications, such as hoarseness, transient dysphagia, or nerve root palsy. Seven (22.6%) patients had reconstruction failure : 5 (20.8%) in the 1-level. corpectomy group and 2 (28.5%) in the 2-level corpectomy group. Revisions were required in 2 patients with plate pullout due to significant instability. However, none of 5 patients who demonstrated cage displacement or screw pullout, underwent a revision. Radiographs revealed bony consolidation in 96.3% of the patients, including 6 patients with implantation failure during the follow-up period. Conclusion : Based on our preliminary results, the titanium mesh cage and dynamic plating was effective for cervical reconstruction after corpectomy. The anterior cervical reconstruction performed with dynamic plates is considered to reduce stress shielding and greater graft compression that is afforded by the unique plate design.
引用
收藏
页码:111 / 116
页数:6
相关论文
共 28 条
  • [1] Barnes Bryan, 2002, Neurosurg Focus, V12, pE13
  • [2] Anterior cervical arthrodesis using DOC dynamic stabilization implant for improvement in sagittal angulation and controlled settling
    Bose, B
    [J]. JOURNAL OF NEUROSURGERY, 2003, 98 (01) : 8 - 13
  • [3] Caspar W, 1998, J SPINAL DISORD, V11, P1
  • [4] Cervical spondylotic myelopathy: 10 years of prospective outcome analysis of anterior decompression and fusion
    Chagas, H
    Domingues, F
    Aversa, A
    Fonseca, ALV
    de Souza, JM
    [J]. SURGICAL NEUROLOGY, 2005, 64 : S30 - S36
  • [5] Cheng N S, 2005, J Orthop Surg (Hong Kong), V13, P223
  • [6] Efficacy and safety of the use of titanium mesh cages and anterior cervical plates for interbody fusion after anterior cervical corpectomy
    Chuang, Hao-Che
    Cho, Der-Yang
    Chang, Cheng-Siu
    Lee, Wen-Yuen
    Chen Jung-Chung
    Lee, Han-Chung
    Chen, Chun-Chung
    [J]. SURGICAL NEUROLOGY, 2006, 65 (05): : 464 - 471
  • [7] Connolly PJ, 1996, J SPINAL DISORD, V9, P202
  • [8] Use of cylindrical titanium mesh and locking plates in anterior cervical fusion - Technical note
    Das, K
    Couldwell, WT
    Sava, G
    Taddonio, RF
    [J]. JOURNAL OF NEUROSURGERY, 2001, 94 (01) : 174 - 178
  • [9] Early failures following cervical corpectomy reconstruction with titanium mesh cages and anterior plating
    Daubs, MD
    [J]. SPINE, 2005, 30 (12) : 1402 - 1406
  • [10] Anterior cervical plating reverses load transfer through multilevel strut-grafts
    DiAngelo, DJ
    Foley, KT
    Vossel, KA
    Rampersaud, YR
    Jansen, TH
    [J]. SPINE, 2000, 25 (07) : 783 - 795