BENEFIT OF INTRAOPERATIVE OPTICAL COHERENCE TOMOGRAPHY FOR VITREOMACULAR SURGERY IN HIGHLY MYOPIC EYES

被引:18
作者
Bruyere, Elsa [1 ]
Philippakis, Elise [1 ]
Dupas, Benedicte [1 ]
Nguyen-Kim, Pauline [1 ]
Tadayoni, Ramin [1 ]
Couturier, Aude [1 ]
机构
[1] Univ Paris 07, Hop Lariboisiere, AP HP, Dept Ophthalmol,Sorbonne Paris Cite, Paris, France
来源
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES | 2018年 / 38卷 / 10期
关键词
high myopia; intraoperative optical coherence tomography; myopic foveoschisis; myopic macular hole; myopic traction maculopathy; myopic vitreoretinal diseases; vitreoretinal surgery; INTERNAL LIMITING MEMBRANE; MACULAR HOLE; RETINAL-DETACHMENT; FLAP TECHNIQUE; VITRECTOMY; HEMORRHAGE; DYNAMICS; REMOVAL; PIONEER; UTILITY;
D O I
10.1097/IAE.0000000000001827
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To report the feasibility and information provided by intraoperative optical coherence tomography (iOCT) during vitreomacular surgery in highly myopic eyes. Methods: Retrospective observational case series on consecutive highly myopic eyes that underwent vitreomacular surgery with iOCT for epiretinal membrane (ERM), macular hole, and myopic foveoschisis. The main outcome was the qualitative and quantitative assessment of retinal changes: detection of persistent epiretinal structures, new openings, central macular thickness, and macular hole diameters after each step of the surgical procedure. Quantitative measurements (in pixels) were manually obtained on iOCT video screen captures. Results: Twenty-two eyes were included: six ERMs, 10 macular holes, and 6 with myopic foveoschisis. An unsuspected postpeeling macular opening was detected by iOCT in 2/22 eyes. Intraoperative optical coherence tomography also allowed for detecting the presence of residual fragments of the vitreous cortex in 6/12 eyes after surgically induced posterior vitreous detachment. Intraoperative optical coherence tomography detected residual fragments of the internal limiting membrane in 5/21 eyes after internal limiting membrane peeling, and residual fragments of ERM in 3/15 eyes with ERM. Quantitative analysis did not find any significant change in central macular thickness and macular hole diameters before and after ERM and internal limiting membrane peeling. Conclusion: In highly myopic eyes, iOCT could help assess undetected macular openings and otherwise posterior vitreous status and epiretinal structure peeling.
引用
收藏
页码:2035 / 2044
页数:10
相关论文
共 30 条
[1]   Anatomical and visual outcomes in high myopic macular hole (HM-MH) without retinal detachment: a review [J].
Alkabes, Micol ;
Pichi, Francesco ;
Nucci, Paolo ;
Massaro, Domenico ;
Dutra Medeiros, Marco ;
Corcostegui, Borja ;
Mateo, Carlos .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2014, 252 (02) :191-199
[2]   FEASIBILITY OF INTRASURGICAL SPECTRAL-DOMAIN OPTICAL COHERENCE TOMOGRAPHY [J].
Binder, Susanne ;
Falkner-Radler, Christiane I. ;
Hauger, Christoph ;
Matz, Holger ;
Glittenberg, Carl .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2011, 31 (07) :1332-1336
[3]   INTRAOPERATIVE USE OF HANDHELD SPECTRAL DOMAIN OPTICAL COHERENCE TOMOGRAPHY IMAGING IN MACULAR SURGERY [J].
Dayani, Pouya N. ;
Maldonado, Ramiro ;
Farsiu, Sina ;
Toth, Cynthia A. .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2009, 29 (10) :1457-1468
[4]   INTRAOPERATIVE OPTICAL COHERENCE TOMOGRAPHY DURING VITREORETINAL SURGERY FOR DENSE VITREOUS HEMORRHAGE IN THE PIONEER STUDY [J].
Ehlers, Justis P. ;
Griffith, Joseph F. ;
Srivastava, Sunil K. .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2015, 35 (12) :2537-2542
[5]   Determination of Feasibility and Utility of Microscope-Integrated Optical Coherence Tomography During Ophthalmic Surgery The DISCOVER Study RESCAN Results [J].
Ehlers, Justis P. ;
Goshe, Jeff ;
Dupps, William J. ;
Kaiser, Peter K. ;
Singh, Rishi P. ;
Gans, Richard ;
Eisengart, Jonathan ;
Srivastava, Sunil K. .
JAMA OPHTHALMOLOGY, 2015, 133 (10) :1124-1132
[6]   Intrasurgical Assessment of Subretinal tPA Injection for Submacular Hemorrhage in the PIONEER Study Utilizing Intraoperative OCT [J].
Ehlers, Justis P. ;
Petkovsek, Daniel S. ;
Yuan, Alex ;
Singh, Rishi P. ;
Srivastava, Sunil K. .
OPHTHALMIC SURGERY LASERS & IMAGING RETINA, 2015, 46 (03) :327-332
[7]   Factors Associated With Persistent Subfoveal Fluid and Complete Macular Hole Closure in the PIONEER Study [J].
Ehlers, Justis P. ;
Itoh, Yuji ;
Xu, Lucy T. ;
Kaiser, Peter K. ;
Singh, Rishi P. ;
Srivastava, Sunil K. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (02) :1141-1146
[8]   The Prospective Intraoperative and Perioperative Ophthalmic ImagiNg With Optical CoherEncE TomogRaphy (PIONEER) Study: 2-Year Results [J].
Ehlers, Justis P. ;
Dupps, William J. ;
Kaiser, Peter K. ;
Goshe, Jeff ;
Singh, Rishi P. ;
Petkovsek, Daniel ;
Srivastava, Sunil K. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2014, 158 (05) :999-1007
[9]   Intraoperative optical coherence tomography using the RESCAN 700: preliminary results from the DISCOVER study [J].
Ehlers, Justis P. ;
Kaiser, Peter K. ;
Srivastava, Sunil K. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2014, 98 (10) :1329-1332
[10]   UTILITY OF INTRAOPERATIVE OPTICAL COHERENCE TOMOGRAPHY DURING VITRECTOMY SURGERY FOR VITREOMACULAR TRACTION SYNDROME [J].
Ehlers, Justis P. ;
Tam, Tiffany ;
Kaiser, Peter K. ;
Martin, Daniel F. ;
Smith, Gina M. ;
Srivastava, Sunil K. .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2014, 34 (07) :1341-1346