EANM dosimetry committee series on standard operational procedures: a unified methodology for 99mTc-MAA pre- and 90Y peri-therapy dosimetry in liver radioembolization with 90Y microspheres

被引:85
|
作者
Chiesa, Carlo [1 ]
Sjogreen-Gleisner, Katarina [2 ]
Walrand, Stephan [3 ]
Strigari, Lidia [4 ]
Flux, Glenn [5 ,6 ]
Gear, Jonathan [5 ,6 ]
Stokke, Caroline [7 ]
Gabina, Pablo Minguez [8 ]
Bernhardt, Peter [9 ,10 ]
Konijnenberg, Mark [11 ]
机构
[1] Fdn IRCCS Ist Nazl Tumori, Nucl Med Unit, Milan, Italy
[2] Lund Univ, Dept Med Radiat Phys, Lund, Sweden
[3] Catholic Univ Louvain, Nucl Med Mol Imaging Radiotherapy & Oncol Unit MI, IECR, Brussels, Belgium
[4] IRCCS Azienda Osped Univ Bologna, Med Phys Div, Bologna, Italy
[5] Royal Marsden Hosp, Joint Dept Phys, Sutton, Surrey, England
[6] Inst Canc Res, Sutton, Surrey, England
[7] Oslo Univ Hosp, Dept Diagnost Phys, Oslo, Norway
[8] Gurutzeta Cruces Univ Hosp, Dept Med Phys & Radiat Protect, Baracaldo, Spain
[9] Univ Gothenburg, Sahlgrenska Acad, Inst Clin Sci, Dept Radiat Phys, Gothenburg, Sweden
[10] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, Gothenburg, Sweden
[11] Erasmus MC, Dept Radiol & Nucl Med, Rotterdam, Netherlands
关键词
Radioembolization dosimetry; Tc-99m-MAA; Y-90; microspheres; Lung shunt; Liver dosimetry; Tumour dosimetry; PET; Prospective; retrospective dosimetry; SELECTIVE INTERNAL RADIOTHERAPY; VOXEL S-VALUES; HEPATOCELLULAR-CARCINOMA; GLASS MICROSPHERES; RADIATION-THERAPY; PARTITION MODEL; TC-99M-MACROAGGREGATED ALBUMIN; BREMSSTRAHLUNG SPECT/CT; IMAGE-RECONSTRUCTION; QUANTITATIVE SPECT;
D O I
10.1186/s40658-021-00394-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this standard operational procedure is to standardize the methodology employed for the evaluation of pre- and post-treatment absorbed dose calculations in Y-90 microsphere liver radioembolization. Basic assumptions include the permanent trapping of microspheres, the local energy deposition method for voxel dosimetry, and the patient-relative calibration method for activity quantification.The identity of Tc-99m albumin macro-aggregates (MAA) and Y-90 microsphere biodistribution is also assumed. The large observed discrepancies in some patients between Tc-99m-MAA predictions and actual Y-90 microsphere distributions for lesions is discussed. Absorbed dose predictions to whole non-tumoural liver are considered more reliable and the basic predictors of toxicity. Treatment planning based on mean absorbed dose delivered to the whole non-tumoural liver is advised, except in super-selective treatments. Given the potential mismatch between MAA simulation and actual therapy, absorbed doses should be calculated both pre- and post-therapy. Distinct evaluation between target tumours and non-tumoural tissue, including lungs in cases of lung shunt, are vital for proper optimization of therapy. Dosimetry should be performed first according to a mean absorbed dose approach, with an optional, but important, voxel level evaluation. Fully corrected Tc-99m-MAA Single Photon Emission Computed Tomography (SPECT)/computed tomography (CT) and Y-90 TOF PET/CT are regarded as optimal acquisition methodologies, but, for institutes where SPECT/CT is not available, non-attenuation corrected Tc-99m-MAA SPECT may be used. This offers better planning quality than non dosimetric methods such as Body Surface Area (BSA) or mono-compartmental dosimetry. Quantitative Y-90 bremsstrahlung SPECT can be used if dedicated correction methods are available. The proposed methodology is feasible with standard camera software and a spreadsheet. Available commercial or free software can help facilitate the process and improve calculation time.
引用
收藏
页数:44
相关论文
共 39 条
  • [1] EANM dosimetry committee series on standard operational procedures: a unified methodology for 99mTc-MAA pre- and 90Y peri-therapy dosimetry in liver radioembolization with 90Y microspheres
    Carlo Chiesa
    Katarina Sjogreen-Gleisner
    Stephan Walrand
    Lidia Strigari
    Glenn Flux
    Jonathan Gear
    Caroline Stokke
    Pablo Minguez Gabina
    Peter Bernhardt
    Mark Konijnenberg
    EJNMMI Physics, 8
  • [2] Factors modulating 99mTc-MAA planar lung dosimetry for 90Y radioembolization
    Lopez, Benjamin P.
    Mahvash, Armeen
    Long, James P.
    Lam, Marnix G. E. H.
    Kappadath, S. Cheenu
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12):
  • [3] SPECT/CT with 99mTc-MAA in Radioembolization with 90Y Microspheres in Patients with Hepatocellular Cancer
    Hamami, Monia E.
    Poeppel, Thorsten D.
    Mueller, Stephan
    Heusner, Till
    Bockisch, Andreas
    Hilgard, Philipp
    Antoch, Gerald
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (05) : 688 - 692
  • [4] An estimate of 90Y dosimetry for bremsstrahlung SPECT/CT imaging in liver therapy with 90Y microspheres
    Knesaurek, Karin
    EUROPEAN JOURNAL OF RADIOLOGY, 2021, 139
  • [5] Dosimetry and prescription in liver radioembolization with 90Y microspheres: 3D calculation of tumor-to-liver ratio from global 99mTc-MAA SPECT information
    Maneru, Fernando
    Abos, Dolores
    Bragado, Laura
    Fuentemilla, Naiara
    Caudepon, Fernando
    Pellejero, Santiago
    Miquelez, Santiago
    Rubio, Anastasio
    Goni, Elena
    Hernandez-Vitoria, Araceli
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (23) : 9099 - 9111
  • [6] PET-CT post therapy dosimetry in radioembolization with resin 90Y microspheres: Comparison with pre-treatment SPECT-CT 99mTc-MAA results
    Richetta, Elisa
    Pasquino, Massimo
    Poli, Matteo
    Cutaia, Claudia
    Valero, Chiara
    Tabone, Marco
    Paradisi, Benedetta Peiretti
    Pacilio, Massimiliano
    Pellerito, Riccardo Emanuele
    Stasi, Michele
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 64 : 16 - 23
  • [7] Comparative dosimetry between 99mTc-MAA SPECT/CT and 90Y PET/CT in primary and metastatic liver tumors
    Jadoul, Alexandre
    Bernard, Claire
    Lovinfosse, Pierre
    Gerard, Laurent
    Lilet, Henri
    Cornet, Olivier
    Hustinx, Roland
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2020, 47 (04) : 828 - 837
  • [8] Comparison of posttherapy 90Y positron emission tomography/computed tomography dosimetry methods in liver therapy with 90Y microspheres
    Knesaurek, Karin
    WORLD JOURNAL OF NUCLEAR MEDICINE, 2020, 19 (04) : 359 - 365
  • [9] A Workflow for Dosimetry of 90Y Radioembolization Based on Quantitative 99mTc-MAA SPECT/CT Imaging and a 3D-Printed Phantom
    Ungania, Sara
    D'Arienzo, Marco
    Mezzenga, Emilio
    Pizzi, Giuseppe
    Vallati, Giulio
    Ianiro, Anna
    Rea, Sandra
    Sciuto, Rosa
    Soriani, Antonella
    Strigari, Lidia
    APPLIED SCIENCES-BASEL, 2022, 12 (20):
  • [10] Comparative dosimetry between 99mTc-MAA SPECT/CT and 90Y PET/CT in primary and metastatic liver tumors
    Alexandre Jadoul
    Claire Bernard
    Pierre Lovinfosse
    Laurent Gérard
    Henri Lilet
    Olivier Cornet
    Roland Hustinx
    European Journal of Nuclear Medicine and Molecular Imaging, 2020, 47 : 828 - 837