共 50 条
A comparison of polysomnography and the WatchPAT in the diagnosis of obstructive sleep apnea
被引:69
|作者:
Pang, Kenny P.
Gourin, Christine G.
Terris, David J.
机构:
[1] Pacific Sleep Ctr, Paragon Med Ctr, Dept Otolaryngol, Singapore, Singapore
[2] Pacific Sleep Ctr, Singapore, Singapore
[3] Med Coll Georgia, Dept Otolaryngol Head & Neck Surg, Augusta, GA 30912 USA
关键词:
D O I:
10.1016/j.otohns.2007.03.015
中图分类号:
R76 [耳鼻咽喉科学];
学科分类号:
100213 ;
摘要:
OBJECTIVE: Our goal was to validate the WatchPAT in the diagnosis of obstructive sleep apnea. STUDY DESIGN: We conducted a prospective, blinded, non-randomized clinical trial. METHODS: Patients with suspected obstructive sleep apnea scheduled for an overnight level I polysomnogram were offered enrollment in a study to compare the WatchPAT (Itamar Ltd, Israel) device with polysomnography. Patients wore the WatchPAT device simultaneously while undergoing polysomnography during evaluation in the sleep lab. RESULTS: Thirty-seven patients participated in the study. They had a mean age of 50.1 years (range, 31-73 years) and mean body mass index of 34.6 kg/m(2) (range, 21.2-46.8 kg/m(2)). There was high correlation between the polysomnogram and WatehPAT apnea-hypopnea index (r = 0.9288; 95% confidence interval = 0.8579-0.9650, P < 0.0001). The lowest oxygen saturation also showed high correlation (r = 0.989; 95% confidence interval = 0.9773-0.9947, P < 0.0001). The overall polysomnogram and WatchPAT sleep times revealed a correlation of r = 0.5815 (P = 0.005). CONCLUSION: The WatchPAT showed a high correlation with the polysomnogram in apnea-hypopnea index, lowest oxygen saturation, and sleep time. SIGNIFICANCE: It's use as a reliable tool in the diagnosis of Obstructive Sleep Apnea. (c) 2007 American Academy of Otolaryngology-Head and Neck Surgery Foundation. All rights reserved.
引用
收藏
页码:665 / 668
页数:4
相关论文