Analytical evaluation of the epoc® point-of-care blood analysis system in cardiopulmonary bypass patients

被引:18
作者
Chen, Jianing [1 ,2 ]
Gorman, Monique [3 ]
O'Reilly, Bill [4 ]
Chen, Yu [1 ,5 ]
机构
[1] Dr Everett Chalmers Reg Hosp, Horizon Hlth Network, Dept Lab Med, Fredericton, NB E3B 5N5, Canada
[2] Natl Univ Ireland Univ Coll Cork, Fac Med, Cork, Ireland
[3] St Johns Hosp, Horizon Hlth Network, Dept Lab Med, St John, NB, Canada
[4] St Johns Hosp, Horizon Hlth Network, Div Cardiac Perfus, St John, NB, Canada
[5] Dalhousie Univ, Dept Pathol, Halifax, NS, Canada
关键词
Point-of-care testing; Blood gas analysis; epoc (R); Method evaluation; Comparison study; Cardiac surgery; RAPID SERUM TUBE; CARDIAC-SURGERY; HEMATOCRIT; CHEMISTRY; STAT; GAS; HEMOGLOBIN; MANAGEMENT; ANALYZER;
D O I
10.1016/j.clinbiochem.2015.12.015
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: The aim of this study was to evaluate the analytical performance of the new epoc (R) point-of-care blood analysis system in cardiopulmonary bypass patients. Design and methods: The precision study was conducted on 3 epoc (R) blood analysis systems using 5 levels of quality control materials twice per day for 5 days. The blood specimen was collected in blood gas syringes from 40 cardiac perfusion patients for the comparison study on epoc (R) (all 3 meters), Instrumentation Laboratory GEM4000, Abbott iSTAT, Nova CCX, and Roche Accu-Chek Inform II and Performa glucose meters. Results: The epoc (R) blood analysis systems demonstrated clinically acceptable precision for all analytes (from 0.07%, 0.07%, and 0.13% for pH 7.6, 7.4, and 7.0 levels; to 3.87%, 3.74%, and 7.56% for pO(2) 197, 103, and 56 mm Hg levels). Comparison studies yielded a correlation coefficient R from 0.9201 (sodium) to 0.9969 (pO2) with the GEM4000; from 0.9071 (sodium) to 0.9965 (potassium) with the iSTAT; from 0.8793 (sodium) to 0.9957 (pO(2)) with the CCX, and 0.9850 and 0.9904 with Roche Inform II and Performa meters respectively. Average biases for all analytes were within the total allowable error limits. Conclusion: The epoc (R) blood analysis system is acceptable for point-of-care testing in the cardiovascular surgery setting. (C) 2016 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:708 / 712
页数:5
相关论文
共 27 条
[1]  
[Anonymous], 2004, Evaluation of Precision Performance of Quantitative Measurement Methods
[2]  
Approved Guideline, VSecond
[3]  
[Anonymous], CLIN HEMATOLOGY THEO
[4]  
[Anonymous], 2003, EP06A CLSI
[5]  
[Anonymous], 1980, GRADWOHLS CLIN LAB M
[6]  
[Anonymous], FED REG
[7]  
[Anonymous], 2006, EP15A2 CLSI
[8]  
[Anonymous], EPOC SYST MAN
[9]  
[Anonymous], 2010, EP09A2IR CLSI
[10]  
Bonini P, 2002, CLIN CHEM, V48, P691