A total consumption (up to 75 μL min-1) infrared-heated sample introduction system for inductively coupled plasma optical emission spectrometry

被引:10
作者
Al Hejami, Ahmed [1 ]
Burgener, Mirah J. [2 ]
Burgener, John [3 ]
Beauchemin, Diane [1 ]
机构
[1] Queens Univ, Dept Chem, Kingston, ON K7L 3N6, Canada
[2] Burgener Res, 9710 Second St,Unit 103, Sidney, BC V8L 3C4, Canada
[3] Burgener Res, 1680 Lakeshore Rd West,Unit 2, Mississauga, ON L5J 1J5, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
PHOTOCHEMICAL VAPOR GENERATION; TEMPERATURE SPRAY CHAMBER; HIGH-EFFICIENCY NEBULIZER; ATOMIC EMISSION; MASS-SPECTROMETRY; PNEUMATIC NEBULIZATION; ANALYTICAL PERFORMANCE; ICP-OES; DESOLVATION; RADIATION;
D O I
10.1039/d0ja00068j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study compares the analytical figures of merit of inductively coupled plasma optical emission spectrometry (ICPOES) with an infrared (IR)-heated modified cyclonic spray chamber (MC) system providing total sample consumption at up to 75 mu L min(1), Apex E desolvation system and IsoMist programmable temperature spray chamber. A baffled cyclonic spray chamber (BC) at room temperature was used as a reference sample introduction system. Sample uptake rates of 25, 50 and 75 mu L min(-1)were tested. Under operating conditions providing a compromise in sensitivity and robustness, and at a sample uptake rate of 50 mu L min(-1), the improvement in the sensitivity and detection limit of the MC and Apex was comparable and significantly higher than that for IsoMist. However, at a sample uptake rate of 75 mu L min(-1), the order of the improvement in the sensitivity and detection limit was Apex > MC > IsoMist. In general, at sample uptake rates of 50 and 75 mu L min(-1)and in contrast to IsoMist, the sensitivity and detection limit for the MC and Apex were not decreased against those obtained with the BC at a regular sample uptake rate of 1 mL min(-1). The MC system provided the highest plasma robustness. At a given sample uptake rate, the MC also provided a shorter washout time than Apex and IsoMist. Accurate ICPOES analysis of two certified reference materials of drinking and waste waters was performed with the MC using a simple external calibration without internal standardization and the results were similar to those obtained for the BC, IsoMist and Apex. This demonstrates the suitability of the MC as a sample introduction system for ICPOES.
引用
收藏
页码:1125 / 1130
页数:6
相关论文
共 34 条
[1]   Infrared heating of the top surface of a cyclonic spray chamber to improve the analytical performance of inductively coupled plasma optical emission spectrometry [J].
Al Hejami, Ahmed ;
Beauchemin, Diane .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2019, 34 (01) :232-238
[2]   Infrared heating of commercially available spray chambers to improve the analytical performance of inductively coupled plasma optical emission spectrometry [J].
Al Hejami, Ahmed ;
Beauchemin, Diane .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2018, 33 (11) :2008-2014
[3]  
Al Hejami A, 2018, J ANAL ATOM SPECTROM, V33, P738, DOI [10.1039/c8ja00044a, 10.1039/C8JA00044A]
[4]   Improving the analytical performances of ICP-AES by using a high-temperature single-pass spray chamber and segmented-injections micro-sample introduction for the analysis of environmental samples [J].
Ardini, Francisco ;
Grotti, Marco ;
Sanchez, Raquel ;
Luis Todoli, Jose .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2012, 27 (09) :1400-1404
[5]   Ultrasonic nebulization with an infrared heated pre-evaporation tube for sample introduction in ICP-OES: application to geological and environmental samples [J].
Asfaw, Alemayehu ;
MacFarlane, William R. ;
Beauchemin, Diane .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2012, 27 (08) :1254-1263
[6]   Liquid sample introduction in inductively coupled plasma atomic emission and mass spectrometry - Critical review [J].
Bings, N. H. ;
von Niessen, J. O. Oriandini ;
Schaper, J. N. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2014, 100 :14-37
[7]   Analysis of whole blood by ICP-MS equipped with a high temperature total sample consumption system [J].
Canabate, Agueda ;
Garcia-Ruiz, Esperanza ;
Resano, Martin ;
Todoli, Jose-Luis .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2017, 32 (01) :78-87
[8]  
Cerutti C, 2019, J ANAL ATOM SPECTROM, V34, P674, DOI [10.1039/C8JA00391B, 10.1039/c8ja00391b]
[9]   Recent advances in inductively coupled plasma optical emission spectrometry [J].
Donati, George L. ;
Amais, Renata S. ;
Williams, Charles B. .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2017, 32 (07) :1283-1296
[10]   Desolvating the sample aerosol with microwave radiation: further theoretical and experimental insight into the significance of such approach [J].
Douglass, KO ;
Fitzgerald, N ;
Ingebrethsen, BJ ;
Tyson, JE .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2004, 59 (02) :261-270