Can Dual Mobility Cups prevent Dislocation in All Situations After Revision Total Hip Arthroplasty?

被引:68
作者
Wegrzyn, Julien [1 ,2 ]
Tebaa, Eloise [1 ]
Jacque, Alexandre [1 ]
Carret, Jean-Paul [1 ]
Bejui-Hugues, Jacques [1 ]
Pibarot, Vincent [1 ]
机构
[1] Hop Edouard Herriot, Dept Orthoped Surg Pavillon T, F-69437 Lyon, France
[2] Univ Lyon, INSERM, UMR 1033, Lyon, France
关键词
revision; total hip arthroplasty; dual mobility cup; dislocation; intra-prosthetic dislocation; ACETABULAR COMPONENT; RISK-FACTORS; THA; IMPLANTS; CLASSIFICATION; INSTABILITY; MANAGEMENT; ALLOGRAFT; FAILURE; LINERS;
D O I
10.1016/j.arth.2014.10.034
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The outcome of a single design of dual mobility cup was prospectively evaluated in a continuous series of 994 revision THAs with respect to dislocation and intra-prosthetic dislocation (IPD). At a 7.3-year mean follow-up, the dislocation rate was 1.5% and the IPD rate was 0.2%. The 2 IPD occurred in acetabular-only revisions and were related to a poor head-to-neck ratio with early impingement and wear at the polyethylene mobile component chamfer. Dual mobility cups demonstrated a low dislocation rate in revision THA but did not compensate for potential perioperative technical errors. In addition, IPD did not appear to be a concern with respect to the benefit in term of instability prevention though caution is advised in acetabular-only revision associated with a poor head-to-neck ratio. Level of Evidence: Therapeutic study-Level IV (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:631 / 640
页数:10
相关论文
共 33 条
[1]   Dual mobility retentive acetabular liners and wear: Surface analysis of 40 retrieved polyethylene implants [J].
Adam, P. ;
Farizon, F. ;
Fessy, M. -H. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (01) :85-91
[2]   Revision Total Hip Arthroplasty with Retained Acetabular Component [J].
Adelani, Muyibat A. ;
Mall, Nathan A. ;
Nyazee, Humaa ;
Clohisy, John C. ;
Barrack, Robert L. ;
Nunley, Ryan M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2014, 96A (12) :1015-1020
[3]   Dislocationm after revision total hip arthroplasty - An analysis of risk factors and treatment options [J].
Alberton, GM ;
High, WA ;
Morrey, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (10) :1788-1792
[4]   Fractures of a Single Design of Highly Cross-Linked Polyethylene Acetabular Liners: An Analysis of Voluntary Reports to the United States Food and Drug Administration [J].
Ast, Michael P. ;
John, Thomas K. ;
Labbisiere, Anthony ;
Robador, Nicolas ;
Della Valle, Alejandro Gonzalez .
JOURNAL OF ARTHROPLASTY, 2014, 29 (06) :1231-1235
[5]   The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty - Examining the successes and failures [J].
Berend, KR ;
Lombardi, AV ;
Mallory, TH ;
Adams, JB ;
Russell, JH ;
Groseth, KL .
JOURNAL OF ARTHROPLASTY, 2005, 20 (07) :93-102
[6]   Surgical management of recurrent dislocation after total hip arthroplasty [J].
Charissoux, J. -L. ;
Asloum, Y. ;
Marcheix, P. -S. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (01) :S25-S34
[7]   A Dual-mobility Cup Reduces Risk of Dislocation in Isolated Acetabular Revisions [J].
Civinini, Roberto ;
Carulli, Christian ;
Matassi, Fabrizio ;
Nistri, Lorenzo ;
Innocenti, Massimo .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (12) :3542-3548
[8]   Low Rate of Dislocation of Dual-mobility Cups in Primary Total Hip Arthroplasty [J].
Combes, Antoine ;
Migaud, Henri ;
Girard, Julien ;
Duhamel, Alain ;
Fessy, Michel Henri .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (12) :3891-3900
[9]  
DANTONIO JA, 1989, CLIN ORTHOP RELAT R, P126
[10]   Dual mobility cups in total hip arthroplasty [J].
De Martino, Ivan ;
Triantafyllopoulos, Georgios Konstantinos ;
Sculco, Peter Keyes ;
Sculco, Thomas Peter .
WORLD JOURNAL OF ORTHOPEDICS, 2014, 5 (03) :180-187